Humoud Al-Otaibi1, Mohammed D Al-Otaibi1, Rajiv Khandekar2, Ches Souru2, Abdulelah A Al-Abdullah2, Hassan Al-Dhibi2, Donald U Stone2,3, Igor Kozak2,4.
Abstract
PURPOSE: We studied the validity, usefulness, and relative cost to detect diabetic retinopathy (DR) and sight-threatening DR (STDR) by using a hand-held electrophysiologic tool compared to digital fundus photography.Entities:
Keywords: RETeval; diabetic retinopathy; electroretinogram; screening
Year: 2017 PMID: 28516001 PMCID: PMC5433802 DOI: 10.1167/tvst.6.3.3
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1Picture of electrophysiologic tool RETeval for DR screening.
Demographic Profile of Diabetic Patients in the Study
Status of DR Based on Digital Photography Evaluation by Retina Specialist
Figure 2Amplitude of ERG spikes as displayed by RETeval while screening of eyes with different stages of DR. In x axis different stages of DR are given while y axis shows amplitude of electrophysiologic spike (mV) as measured by RETeval.
Validity of RETeval-DR in Screening for STDR Compared to Digital Fundus Photographs Evaluated by Retina Specialist
Validity of RETeval-DR in Screening for DR of All Grades Compared to Digital Fundus Photographs Evaluated by Retina Specialist
Figure 3The ROC curve for eyes with STDR (proliferative DR and/or diabetic macular edema) as diagnosed by RETeval screening and confirmed by digital fundus photography. The x axis shows fraction of 1-specificity while y axis shows fraction of sensitivity. The area under the red line depicts AUC.
Figure 4The ROC for eyes with nonproliferative and proliferative DR as diagnosed by RETeval screening and confirmed by digital fundus photography. The x axis shows fraction of 1-specificity while y axis shows fraction of sensitivity. The area under the red line depicts AUC.
Figure 5The ROC for eyes with any stage of DR as diagnosed by RETeval screening and confirmed by digital fundus photography. The x axis shows fraction of 1-specificity while y axis shows fraction of sensitivity. The area under the red line depicts AUC.