Effect of Pupil Size on Flicker ERGs Recorded With RETeval System: New Mydriasis-Free Full-Field ERG System.

Kumiko Kato, Mineo Kondo, Masahiko Sugimoto, Kengo Ikesugi, Hisashi Matsubara.   

Abstract

PURPOSE: We studied whether pupil size affects the flicker electroretinograms (ERGs) recorded by RETeval, a new mydriasis-free full-field flicker ERG system.
METHODS: We studied 10 healthy subjects. The RETeval manufacturer claims that the system delivers a constant flash retinal illuminance by adjusting the flash luminance to compensate for changes in the pupil size. Two experiments were performed. First, the flicker ERG was recorded every 3 minutes after the instillation of mydriatics. Second, the flicker ERG was recorded while the subjects wore soft contact lenses with two different artificial pupil sizes.
RESULTS: The first experiment showed that as pupil size increased, the amplitudes of the fundamental component of the flicker ERG did not change significantly, but the implicit times of the fundamental component were significantly prolonged for larger pupil sizes. There was a significant positive correlation between the pupillary area and implicit time of the fundamental component (r = 0.93, P < 0.001). The second experiment showed that the implicit times of the fundamental component in the flicker ERG were significantly longer with larger artificial pupil.
CONCLUSIONS: The results suggest that the effective retinal illuminance of the stimulus delivered by the RETeval system decreases for large pupil sizes. However, in most clinical testing situations, patients' undilated pupils will likely be sufficiently small to fall within the range for which the system delivers a stimulus of constant retinal illuminance.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26047169     DOI: 10.1167/iovs.14-16349

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  30 in total

1.  Retinal vasculature-function correlation in non-proliferative diabetic retinopathy.

Authors:  Yunkao Zeng; Dan Cao; Dawei Yang; Xuenan Zhuang; Yunyan Hu; Miao He; Honghua Yu; Jun Wang; Cheng Yang; Liang Zhang
Journal:  Doc Ophthalmol       Date:  2019-09-24       Impact factor: 2.379

2.  Flicker electroretinogram recorded with portable ERG device in prematurely born schoolchildren with and without ROP.

Authors:  Manca Tekavčič Pompe; Maja Šuštar
Journal:  Doc Ophthalmol       Date:  2019-04-11       Impact factor: 2.379

3.  Flicker electroretinograms recorded with mydriasis-free RETeval system before and after cataract surgery.

Authors:  G Miura; E Sato; S Yamamoto
Journal:  Eye (Lond)       Date:  2017-06-16       Impact factor: 3.775

4.  Hand-held, dilation-free, electroretinography in children under 3 years of age treated with vigabatrin.

Authors:  Xiang Ji; Michelle McFarlane; Henry Liu; Annie Dupuis; Carol A Westall
Journal:  Doc Ophthalmol       Date:  2019-03-02       Impact factor: 2.379

5.  Pupil assessment with a new handheld pupillometer in healthy subjects.

Authors:  Ken Asakawa; Mei Imai; Mizuki Ohta; Naomi Kawata; Nanako Kawatsu; Hitoshi Ishikawa
Journal:  Int Ophthalmol       Date:  2022-07-12       Impact factor: 2.029

6.  Flicker electroretinogram in newborn infants.

Authors:  James V M Hanson; Caroline Weber; Oliver A Pfäffli; Dirk Bassler; Daphne L McCulloch; Christina Gerth-Kahlert
Journal:  Doc Ophthalmol       Date:  2022-10-06       Impact factor: 1.854

7.  Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data.

Authors:  Henry Liu; Xiang Ji; Sabrina Dhaliwal; Syeda Naima Rahman; Michelle McFarlane; Anupreet Tumber; Jeff Locke; Tom Wright; Ajoy Vincent; Carol Westall
Journal:  Doc Ophthalmol       Date:  2018-10-24       Impact factor: 2.379

8.  The influence of mild cataract on ISCEV standard electroretinogram recorded from mydriatic eyes.

Authors:  Atsuhiro Tanikawa; Keita Suzuki; Ryoko Nomura; Hidenori Tanaka; Tadashi Mizuguchi; Yoshiaki Shimada; Masayuki Horiguchi
Journal:  Doc Ophthalmol       Date:  2020-09-12       Impact factor: 2.379

9.  A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy.

Authors:  April Y Maa; William J Feuer; C Quentin Davis; Ensa K Pillow; Tara D Brown; Rachel M Caywood; Joel E Chasan; Stephen R Fransen
Journal:  J Diabetes Complications       Date:  2015-12-09       Impact factor: 2.852

10.  Role of a mydriasis-free, full-field flicker ERG device in the detection of diabetic retinopathy.

Authors:  Mehmet Fatih Kağan Değirmenci; Sibel Demirel; Figen Batıoğlu; Emin Özmert
Journal:  Doc Ophthalmol       Date:  2018-10-17       Impact factor: 2.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.