Literature DB >> 28492599

Allergic reaction to vanadium causes a diffuse eczematous eruption and titanium alloy orthopedic implant failure.

Sally Engelhart1, Robert J Segal2.   

Abstract

Allergy as a cause of adverse outcomes in patients with implanted orthopedic hardware is controversial. Allergy to titanium-based implants has not been well researched, as titanium is traditionally thought to be inert. We highlight the case of a patient who developed systemic dermatitis and implant failure after surgical placement of a titanium alloy (Ti6Al4V) plate in the left foot. The hardware was removed and the eruption cleared in the following weeks. The plate and screws were submitted for metal analysis. The elemental composition of both the plate and screws included 3 major elements-titanium, aluminum, and vanadium-as well as trace elements. Metal analysis revealed that the plate and screws had different microstructures, and electrochemical studies demonstrated that galvanic corrosion could have occurred between the plate and screws due to their different microstructures, contributing to the release of vanadium in vivo. The patient was patch tested with several metals including components of the implant and had a positive patch test reaction only to vanadium trichloride. These findings support a diagnosis of vanadium allergy and suggests that clinicians should consider including vanadium when patch testing patients with a suspected allergic reaction to vanadium-containing implants.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28492599

Source DB:  PubMed          Journal:  Cutis        ISSN: 0011-4162


  5 in total

Review 1.  Metallic Implants Used in Lumbar Interbody Fusion.

Authors:  Jakub Litak; Michał Szymoniuk; Wojciech Czyżewski; Zofia Hoffman; Joanna Litak; Leon Sakwa; Piotr Kamieniak
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

2.  Heat Treatments of Metastable β Titanium Alloy Ti-24Nb-4Zr-8Sn Processed by Laser Powder Bed Fusion.

Authors:  Maxwell Hein; Nelson Filipe Lopes Dias; Sudipta Pramanik; Dominic Stangier; Kay-Peter Hoyer; Wolfgang Tillmann; Mirko Schaper
Journal:  Materials (Basel)       Date:  2022-05-25       Impact factor: 3.748

Review 3.  Mixed oxide nanotubes in nanomedicine: A dead-end or a bridge to the future?

Authors:  Masoud Sarraf; Bahman Nasiri-Tabrizi; Chai Hong Yeong; Hamid Reza Madaah Hosseini; Saeed Saber-Samandari; Wan Jefrey Basirun; Takuya Tsuzuki
Journal:  Ceram Int       Date:  2020-09-24       Impact factor: 4.527

Review 4.  Protective Effects of Dietary Antioxidants against Vanadium-Induced Toxicity: A Review.

Authors:  Iwona Zwolak
Journal:  Oxid Med Cell Longev       Date:  2020-01-07       Impact factor: 6.543

5.  Effects of new beta-type Ti-40Nb implant materials, brain-derived neurotrophic factor, acetylcholine and nicotine on human mesenchymal stem cells of osteoporotic and non osteoporotic donors.

Authors:  Vivien Kauschke; Annett Gebert; Mariana Calin; Jürgen Eckert; Sebastian Scheich; Christian Heiss; Katrin Susanne Lips
Journal:  PLoS One       Date:  2018-02-28       Impact factor: 3.240

  5 in total