Substernal implantation of a subcutaneous implantable cardioverter-defibrillator in a patient with preexisting Hemodialysis Reliable Outflow graft.

Thomas A Boyle1, Joshua Cohen1, Roger Carrillo1.   

Abstract

Entities:  

Keywords:  HeRO graft; Hemodialysis; ICD; Lead placement; S-ICD; Substernal

Year:  2016        PMID: 28491723      PMCID: PMC5419941          DOI: 10.1016/j.hrcr.2016.05.006

Source DB:  PubMed          Journal:  HeartRhythm Case Rep        ISSN: 2214-0271


× No keyword cloud information.

Introduction

KEY TEACHING POINTS Alternative placement of subcutaneous implantable cardioverter-defibrillator (S-ICD) leads may help avoid damage to existing devices. Fluoroscopy can be a helpful addition to the S-ICD implantation procedure in difficult cases. For patients undergoing hemodialysis, S-ICDs may offer an effective strategy for preserving venous real estate. Central vein stenosis is a well-documented consequence of cardiovascular implantable electronic devices. In hemodialysis patients, where preserving venous real estate is crucial, use of subcutaneous implantable cardioverter-defibrillators (S-ICDs) remains controversial. We present a novel approach for implanting an S-ICD in a patient with end-stage renal disease, central venous stenosis, and precordial hemodialysis graft obstructing the normal subcutaneous implant site.

Case report

History

A 34-year-old woman with end-stage renal disease secondary to lupus nephritis, nonischemic cardiomyopathy, central venous stenosis, and a Hemodialysis Reliable Outflow (HeRO) graft (Merit Medical Systems, South Jordan, UT) was referred to our center for implantation of an automatic implantable cardioverter-defibrillator (AICD). The patient had a documented episode of sustained ventricular tachycardia 6 months prior to referral, which took place in the setting of methicillin-resistant Staphylococcus aureus bacteremia and required external defibrillation. At that time, transthoracic echocardiography was performed and revealed a left ventricular ejection fraction of 25%, whereas previous studies had been normal. Upon resolution of the infection, and placement of a HeRO graft, the patient received optimal medical therapy for new-onset heart failure with reduced ejection fraction and was advised to wear a LifeVest (ZOLL Medical Corporation, Pittsburgh, PA). Subsequent echocardiography showed no change in ventricular function and the patient was scheduled to undergo implantation of an AICD for primary prevention of sudden cardiac death. Venograms performed prior to referral reported bilateral subclavian vein occlusion and stenosis of the superior vena cava. Owing to this documented stenosis and the presence of a HeRO graft extending from right brachial artery to left internal jugular vein and then into the right atrium (Figure 1), traditional AICD implantation techniques were of limited value. The decision was made to place a Boston Scientific Emblem S-ICD in the substernal position.
Figure 1

Preoperative 3-dimensional computed tomography reconstruction showing prior Hemodialysis Reliable Outflow graft placement. Anterior and left-lateral views are shown. The arterial graft component originates out of frame from the right brachial artery and crosses the chest subcutaneously (a). The metallic venous outflow portion can be seen traversing superficially to the body of the sternum (b). The graft courses above the left clavicle and enters the left internal jugular vein (c). The graft terminates in the right atrium after entering the heart through the superior vena cava (d).

The placement resulted in excellent sensing from all 3 vectors, showing neither T waves nor artificial potentials (Figure 2). Defibrillation thresholds were tested by inducing ventricular fibrillation, and the first 70-joule shock effectively terminated the arrhythmia 2 times. The shock impedance was recorded as 68 ohms. There were no complications from the case and the patient was doing well at 2-month follow-up.
Figure 2

Postimplant sensing from the subcutaneous implantable cardioverter-defibrillator showing excellent readings in all 3 vectors: primary, secondary, and alternate (25 mm/sec, 2.5 mm/mV).

Methods

After obtaining informed consent, a 4-cm subxiphoid incision was made to access the retrosternal space. Blunt dissection was accomplished using a 10-mm cherry dissector. Under fluoroscopic guidance, an S-ICD was placed through an 11 French ultra-sheath. The lead was positioned deep to the left sternal border and then secured in the subxiphoid fascia using the sleeve provided by the manufacturer (Boston Scientific, Marlborough, MA). The lead was then tunneled to the axillary position and connected to the generator, which was then secured to the fascia (Figure 3).1, 2
Figure 3

Postoperative chest radiograph showing final placement of subcutaneous implantable cardioverter-defibrillator.

Discussion

Although the literature is sparse, alternative lead positions for S-ICDs have been supported by image-based modeling and a limited number of case reports.3, 4, 5 Our experience reinforces the notion that substernal lead placement, in particular, is readily achievable and beneficial in complex scenarios. Entirely subcutaneous ICDs are especially useful in patients with advanced kidney disease because of their susceptibility to central venous stenosis and relatively high risk of sudden cardiac death. However, implanting an S-ICD in this patient as normally indicated was presumed to carry a high risk of damage to either the lead or the preexisting graft. Standard lead placement is in the subcutaneous tissue along the left sternal border. The metallic portion of the dialysis graft occupied that space in this patient, posing a great risk for damage to either device during implantation or during hemodialysis, when the conduit may pulsate. If the lead and the graft were spared from damage during traditional implant, it was theorized that oversensing or “chatter” might occur owing to the metallic skeleton of the HeRO graft. As demonstrated by a limited number of cases where abandoned electrodes cause interference with new devices, nearby metal need not be carrying current to interfere with new ICD leads.7, 8 As a result, placement in the substernal location was warranted.

Conclusion

In patients with central vein stenosis, an entirely subcutaneous ICD is an appealing and efficacious treatment option.1, 9 This case demonstrates how the unique capabilities of the technology can be leveraged to avoid damage to existing implanted devices. In addition, it highlights some of the reasons why S-ICDs are of particular value in the hemodialysis patient population. Evidence is accumulating for the safety and efficacy of S-ICDs in patients with renal failure who are undergoing hemodialysis, but more is still needed.10, 11

KEY TEACHING POINTS

Alternative placement of subcutaneous implantable cardioverter-defibrillator (S-ICD) leads may help avoid damage to existing devices.

Fluoroscopy can be a helpful addition to the S-ICD implantation procedure in difficult cases.

For patients undergoing hemodialysis, S-ICDs may offer an effective strategy for preserving venous real estate.

  11 in total

1.  An entirely subcutaneous implantable cardioverter-defibrillator.

Authors:  Gust H Bardy; Warren M Smith; Margaret A Hood; Ian G Crozier; Iain C Melton; Luc Jordaens; Dominic Theuns; Robert E Park; David J Wright; Derek T Connelly; Simon P Fynn; Francis D Murgatroyd; Johannes Sperzel; Jörg Neuzner; Stefan G Spitzer; Andrey V Ardashev; Amo Oduro; Lucas Boersma; Alexander H Maass; Isabelle C Van Gelder; Arthur A Wilde; Pascal F van Dessel; Reinoud E Knops; Craig S Barr; Pierpaolo Lupo; Riccardo Cappato; Andrew A Grace
Journal:  N Engl J Med       Date:  2010-05-12       Impact factor: 91.245

2.  Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator.

Authors:  Raul Weiss; Bradley P Knight; Michael R Gold; Angel R Leon; John M Herre; Margaret Hood; Mayer Rashtian; Mark Kremers; Ian Crozier; Kerry L Lee; Warren Smith; Martin C Burke
Journal:  Circulation       Date:  2013-08-27       Impact factor: 29.690

3.  A multicenter study of shock pathways for subcutaneous implantable defibrillators.

Authors:  Jürgen Kuschyk; Goran Milasinovic; Volker Kühlkamp; Paul R Roberts; Markus Zabel; Franck Molin; Stephen Shorofsky; Kurt D Stromberg; Paul J DeGroot; Francis D Murgatroyd
Journal:  J Cardiovasc Electrophysiol       Date:  2013-10-09

4.  Ventricular oversensing after ICD lead replacement: what is the mechanism?

Authors:  Samuel Sossalla; Joachim Seegers; Lars Lüthje; Christian Sohns; Markus Zabel; Dirk Vollmann
Journal:  Pacing Clin Electrophysiol       Date:  2013-11-29       Impact factor: 1.976

5.  Substernal lead implantation: a novel option to manage DFT failure in S-ICD patients.

Authors:  M Guenther; S Kolschmann; M Knaut
Journal:  Clin Res Cardiol       Date:  2014-10-02       Impact factor: 5.460

6.  Lead interaction: rare cause of oversensing during implantation procedure of implantable cardioverter-defibrillator system.

Authors:  Rafal Gardas; Rafal Mlynarski; Katarzyna Staszak; Anna Drzewiecka; Eugeniusz Pilat; Tadeusz Zajac; Wlodzimierz Kargul
Journal:  Pacing Clin Electrophysiol       Date:  2006-10       Impact factor: 1.976

7.  Outcome of Subcutaneous Implantable Cardioverter Defibrillator Implantation in Patients with End-Stage Renal Disease on Dialysis.

Authors:  Mikhael F El-Chami; Mathew Levy; Heval M Kelli; Mary Casey; Michael H Hoskins; Abhinav Goyal; Jonathan J Langberg; Anshul Patel; David Delurgio; Michael S Lloyd; Angel R Leon; Faisal M Merchant
Journal:  J Cardiovasc Electrophysiol       Date:  2015-06-16

8.  Two-incision technique for implantation of the subcutaneous implantable cardioverter-defibrillator.

Authors:  Reinoud E Knops; Louise R A Olde Nordkamp; Joris R de Groot; Arthur A M Wilde
Journal:  Heart Rhythm       Date:  2013-05-21       Impact factor: 6.343

9.  Outcomes of subcutaneous implantable cardioverter-defibrillator implantation in patients on hemodialysis.

Authors:  Eduard Koman; Ashwani Gupta; Faiz Subzposh; Heath Saltzman; Steven P Kutalek
Journal:  J Interv Card Electrophysiol       Date:  2016-01-14       Impact factor: 1.900

Review 10.  [Cardiac arrest in dialysis patients: Risk factors, preventive measures and management in 2015].

Authors:  Yosu Luque; Aurélien Bataille; Guillaume Taldir; Éric Rondeau; Christophe Ridel
Journal:  Nephrol Ther       Date:  2015-11-04       Impact factor: 0.722

View more
  2 in total

Review 1.  The development of the extravascular defibrillator with substernal lead placement: A new Frontier for device-based treatment of sudden cardiac arrest.

Authors:  Amy E Thompson; Brett Atwater; Lucas Boersma; Ian Crozier; Gregory Engel; Paul Friedman; J Rod Gimbel; Bradley P Knight; Jaimie Manlucu; Francis Murgatroyd; David O'Donnell; Juergen Kuschyk; Paul DeGroot
Journal:  J Cardiovasc Electrophysiol       Date:  2022-05-08       Impact factor: 2.942

2.  The extravascular implantable cardioverter-defibrillator: The pivotal study plan.

Authors:  Ian Crozier; David O'Donnell; Lucas Boersma; Francis Murgatroyd; Jaimie Manlucu; Bradley P Knight; Ulrika Maria Birgersdotter-Green; Christophe Leclercq; Amy Thompson; Robert Sawchuk; Sarah Willey; Christopher Wiggenhorn; Paul Friedman
Journal:  J Cardiovasc Electrophysiol       Date:  2021-08-05       Impact factor: 2.942

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.