Literature DB >> 28470392

Performance of a New Blunt-Tip Coaxial Needle for Percutaneous Biopsy and Drainage of "Hard-To-Reach" Targets.

Roberto Luigi Cazzato1, Julien Garnon2, Behnam Shaygi3, Jean Caudrelier2, Salem Bauones2, Georgia Tsoumakidou2, Guillaume Koch2, Afshin Gangi2.   

Abstract

AIM: To present a new blunt-tip coaxial needle (SoftGuard) applied to access "hard-to-reach" targets undergoing percutaneous image-guided biopsy or drainage.
MATERIALS AND METHODS: All consecutive patients presenting between August and December 2016 with "hard-to-reach" (<10 mm from a critical nearby structure such as vessels, nerves, bowel or adjacent parenchymal organs) solid lesions requiring biopsy (group A) or abscesses requiring drainage for sepsis (group B) were prospectively included. The individual features of each patient and lesion as well as technical and clinical data were collected and analysed.
RESULTS: Twenty-six patients (18 males, 8 females, mean age 59.81 ± 17.53 years) were enrolled in group A and nine (6 males, 3 females, mean age 58.33 ± 13.8 years) in group B. Technical success was achieved in 92.3% of cases from group A and 100% of cases from group B. Five (19.2%) minor complications were noted in group A (four small self-limiting pneumothoraces and one small self-limiting peri-pancreatic haematoma). There were no complications in group B. Histological results in group A accounted for 95% sensitivity, 100% specificity and 95.2% diagnostic accuracy. In group B, mean post-operative C-reactive protein was 41 ± 48.3 mg/L in comparison with 155 ± 117.5 mg/L at baseline (P = 0.004).
CONCLUSIONS: The SoftGuard blunt-tip needle is a safe and effective tool when applied as a coaxial working cannula for percutaneous biopsy or drainage of "hard-to-reach" targets.

Entities:  

Keywords:  Biopsy; Blunt tip; Coaxial needle; Drainage

Mesh:

Substances:

Year:  2017        PMID: 28470392     DOI: 10.1007/s00270-017-1663-8

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  2 in total

1.  Technical safety and efficacy of a blunt-tip microwave ablation electrode for CT-guided ablation of pulmonary ground-glass opacity nodules.

Authors:  Jiachang Chi; Zhi Wang; Min Ding; Hao Hu; Bo Zhai
Journal:  Eur Radiol       Date:  2021-04-02       Impact factor: 5.315

2.  Safety and efficacy of ultrasound-guided percutaneous coaxial core biopsy of pancreatic lesions: a retrospective study.

Authors:  Yujing Xin; Yi Yang; Yi Chen; Yanan Wang; Xiao-Jing Cao; Xiang Zhou
Journal:  J Ultrasound       Date:  2020-07-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.