Literature DB >> 28469106

In vitro Dosimetric Study of Biliary Stent Loaded with Radioactive 125I Seeds.

Li-Hong Yao1, Jun-Jie Wang1, Charles Shang2, Ping Jiang1, Lei Lin1, Hai-Tao Sun1, Lu Liu1, Hao Liu3, Di He4, Rui-Jie Yang1.   

Abstract

BACKGROUND: A novel radioactive 125I seed-loaded biliary stent has been used for patients with malignant biliary obstruction. However, the dosimetric characteristics of the stents remain unclear. Therefore, we aimed to describe the dosimetry of the stents of different lengths - with different number as well as activities of 125I seeds.
METHODS: The radiation dosimetry of three representative radioactive stent models was evaluated using a treatment planning system (TPS), thermoluminescent dosimeter (TLD) measurements, and Monte Carlo (MC) simulations. In the process of TPS calculation and TLD measurement, two different water-equivalent phantoms were designed to obtain cumulative radial dose distribution. Calibration procedures using TLD in the designed phantom were also conducted. MC simulations were performed using the Monte Carlo N-Particle eXtended version 2.5 general purpose code to calculate the radioactive stent's three-dimensional dose rate distribution in liquid water. Analysis of covariance was used to examine the factors influencing radial dose distribution of the radioactive stent.
RESULTS: The maximum reduction in cumulative radial dose was 26% when the seed activity changed from 0.5 mCi to 0.4 mCi for the same length of radioactive stents. The TLD's dose response in the range of 0-10 mGy irradiation by 137Cs γ-ray was linear: y = 182225x - 6651.9 (R2=0.99152; y is the irradiation dose in mGy, x is the TLDs' reading in nC). When TLDs were irradiated by different energy radiation sources to a dose of 1 mGy, reading of TLDs was different. Doses at a distance of 0.1 cm from the three stents' surface simulated by MC were 79, 93, and 97 Gy.
CONCLUSIONS: TPS calculation, TLD measurement, and MC simulation were performed and were found to be in good agreement. Although the whole experiment was conducted in water-equivalent phantom, data in our evaluation may provide a theoretical basis for dosimetry for the clinical application.

Entities:  

Mesh:

Year:  2017        PMID: 28469106      PMCID: PMC5421181          DOI: 10.4103/0366-6999.204936

Source DB:  PubMed          Journal:  Chin Med J (Engl)        ISSN: 0366-6999            Impact factor:   2.628


Introduction

Clinical management of malignant biliary obstruction remains challenging. For these patients, minimally invasive biliary stenting is often preferred. However, restenosis due to tumor ingrowth or compression is a problem.[1] External beam radiotherapy (EBRT) has been used to prolong stent patency, but almost inevitably results in normal tissue toxicity because of the proximity of vital organs. More recently, good results have been reported with the use of a combination of intraluminal 192Ir brachytherapy and stenting.[2345] Recently, low-dose rate 125I seed-loaded intraluminal stents have been developed. Zhu et al.[6] were the first to report encouraging results with its clinical application. However, seed activity, reference point of prescription dose, and the irradiation dose of target were different in the related studies.[67] Therefore, comparison of clinical efficacy among different studies becomes difficult. The American Association of Physicists in Medicine (AAPM) Task Group No. 60 and 149 reports recommend that, for each type of radioactive stent (i.e., of various lengths, diameters, and activities), the three-dimensional (3D) dose distributions should be carefully determined before clinical application.[89] However, there are limited reports that describe the dosimetry of the stents. In this study, we aimed to use the treatment planning system (TPS) calculation, thermoluminescent dosimeter (TLD) measurement, and Monte Carlo (MC) simulation to evaluate the characteristics of dosimetric distribution of the novel radioactive stents of different lengths — with different number as well as activities of 125I seeds –– and to provide a theoretical basis for dosimetry for the clinical application.

Methods

Radioactive 125I source

The geometry of model 6711 125I seed used in this study was provided by the manufacturer (HTA Co., Ltd, Beijing, China). The source capsule consisted of a 0.05-mm-thick titanium cylinder (ρ = 4.54 g/cm3), 4.5 mm long × 0.8 mm in diameter, and with end-weld thickness of 0.4 mm, containing a silver core (ρ = 10.5 g/cm3) of 3.0 mm length and 0.5 mm diameter, onto which a 1-μm layer of 125I had been uniformly absorbed. Source activity was in the range of 0.1–6.0 mCi. The dosimetric characteristics of a single model 6711 125I source in a homogeneous water medium has been previously investigated by several groups.[1011]

Radioactive stent model

The radioactive stent was designed as two separate parts: an inner 8-mm diameter conventional self-expanding biliary nitinol stent to provide support and an outer 10-mm diameter radioactive self-expandable stent loaded with 125I seeds [Nanjing Microinvasive Medical Inc., Nanjing, China; Figure 1a]. Each seed casing, 10 mm long and 0.8 mm in diameter, tightly holds the radioactive seed. The minimal distance between two adjacent radioactive seeds is approximately 7.1 mm in the cross-sectional view [center to center; Figure 1b] and 10.0 mm in the longitudinal direction [end to end; Figure 1c]. In this investigation, three representative models of 4 cm, 6 cm, and 8 cm length loaded with 8, 16, and 24 125I seeds, respectively, were studied. In addition, various radioactive strengths, ranging from 0.4 to 1.0 mCi per seed (in 0.1-mCi increments), were evaluated independently.
Figure 1

Location map of radioactive biliary stents in a rectangular coordinate system. (a) Photograph of the radioactive biliary stent. (b) A cross-sectional view of the biliary stent. (c) Seeds’ location in a longitudinal view of a stent of 4-cm length.

Location map of radioactive biliary stents in a rectangular coordinate system. (a) Photograph of the radioactive biliary stent. (b) A cross-sectional view of the biliary stent. (c) Seeds’ location in a longitudinal view of a stent of 4-cm length.

Treatment planning system calculation

This process and the equivalent model have been described in detail in a previous published paper.[12] In brief, solid paraffin (ρ = 0.880–0.915 g/cm3, Leica, Germany), welded to a polymethylmethacrylate (PMMA) cylindrical barrel of 15-cm height and 20-cm diameter, was used as the water-equivalent phantom. The stent loaded with dummy 125I sources (model 6711; China Isotope Corporation, Beijing, China) was vertically mounted at the bottom of the barrel. For CT scan, the cylindrical PMMA barrel was mounted on a head immobilization. Scans were performed with 5-mm slice thickness and spacing to cover the entire length of the stent. DICOM images were sent to the TPS (Prowess - 3D, SSGI, USA) for contouring and planning. Calculation points were selected from polar angles ranging from 0° to 360° in 45° increments and at radial distances ranging from 1 cm to 8 cm in 0.5-cm increments and points of 1.25 and 1.75 cm. Thus, each data point represented the average of the eight calculations.

Thermoluminescent dosimeter measurement

Radial dose distribution around the stents was measured in a PMMA phantom using TLD-GR 200A (Beijing Guangtong Yirun radiation monitoring equipment Co., Ltd.) circular chips of 0.8-mm thickness and 4.5-mm diameter. Before exposure, all the TLDs were annealed at 240°C for 10 min, and then fast cooled. Annealing was repeated as necessary to complete all measurements. The irradiated TLDs were read 24–48 h postirradiation using an automated TLD reader (CTLD-350; Institute of Radiation and Radiation Medicine, Military Medical Science Academy of the PLA). An initial TLD calibration reading was made to study the TLD's linearity of dose response and energy response scaled for exposure duration. To study the dose response of the TLDs, irradiation was performed with a standard 137Cs (662 keV γ-ray) irradiator at the Standard Dosimetry Laboratory (China Institute of Atomic Energy Metering Station, Beijing). In the process, 10 TLDs were used for each configuration of 0.5 mGy, 1 mGy, 2 mGy, 5 mGy, and 10 mGy. As the energy of emitted photons of 125I is in the range of 7.2–35.5 keV,[1011] 33 keV, 65keV, and 83 keV narrow-spectrum X-ray standard radiation sources, 241Am (59.5 keV γ-ray) and 6 MV X-ray (Axesse™ Linear Accelerator, Elekta Medical Systems, Sweden) were selected to irradiate each group of 10 TLDs to a dose of 1 mGy to measure the TLD's energy response. A water-equivalent PMMA phantom was constructed to provide full scattering conditions with thirty pieces of 30 cm × 30 cm × 1 cm slabs. For the central phantom slab to accommodate the circular TLD chips, there was a hole (1.2 cm in diameter) in the center where the radioactive stents could be placed. The square holes (5.0 mm long, 2.0 mm wide, and 4.5 mm deep) were drilled such that the TLD surface would be perpendicular to the slab plane and with the centers of the dosimeters being parallel to the long axis of the source. The whole pattern of the spiral configuration was chosen to minimize inter-dosimeter attenuation and perturbation effects.[1314] The dosimeters were positioned at radial distances of 1 cm to 8 cm in 0.5-cm increments, points of 1.25 and 1.75 cm, and polar angles ranging from 0° to 90° in 10°, with respect to the stent's long axis. Each dosimetric data point represented the average of the four TLD measurements. All measurements were performed within 24 h after in-phantom exposures[15] and repeated thrice to increase the confidence level. According to the AAPM TG43,[10] the dose rate in water surrounding a 125I source can be calculated as follows: where, Sk is the air kerma strength given by Sk = k · A0 · e−λ, k is the air kerma conversion factor, A0 is the initial activity, and λis the decay constant of 125I. Therefore, the cumulative absorbed dose in interest points over a period of time can be calculated as follows: where, Ḋ(r, θ) is the initial dose rate of the interest point. The measured dose rate from the TLD responses from each point irradiated in the phantom was calculated as follows: where, R is the reading of the TLDs (nC), Cf is the calibration factor for the TLD response (mGy/nC), is the relative absorbed dose sensitivity of the TLD to establish the equivalent dose to water per unit reading in the 125I field, Pphant is the phantom correction factor to derive the corresponding dose to water in water from the 125I seed, and t is in-phantom exposure time equaling to 24 h. and Pphant were taken from literature[16] to be 1.51 and 0.898, respectively. All the irradiations were in the linear response region of the TLDs, so no correction was necessary for the supralinearity effect.

Monte Carlo simulation

The Monte Carlo N-Particle eXtended (MCNPX) (version 2.5) code, which transports many particle types at almost all energies, was used to simulate the movement of photons with random positions, directions, and energy in the given region to calculate the stent's 3D dose rate distribution in liquid water. The MC simulations were performed for comparison with the TPS calculations and TLD measurements at 0.5–10.0 cm from stent surface. In this investigation, the 125I photon spectra were adopted from the AAPM TG43U1 report.[11] The recommended photons per disintegration of 125I are 0.406 (27.202 keV), 0.757 (27.472 keV), 30.98 (0.202 keV), 31.71 (0.0439 keV), and 0.0668 (35.492 keV).[11] The radioactive stent model was centered in a 30 cm water-equivalent (ρ = 0.998 g/cm3) cylinder adequate for all the scattering effects from the surrounding medium. For scoring the 3D dose rate distribution of the radioactive stent, the cylinder was divided into a set of concentric rings with different widths and thicknesses. For scoring the radial dose rate distribution at different distances from the radioactive stent surface, the width and thickness of the ring were 0.1 mm for distances to the surface of the stent <1 cm, 0.2 mm for distances of 1–2 cm, 0.3 mm for distances of 2–3 cm, and 0.5 mm for distances of 3–10 cm. For scoring the axial dose rate distribution in the yz plane [Figure 1c] from z = 0 cm to y = z = 10 cm, the width and thickness of the ring were 0.1 mm for distances to the center of the stent of <2 cm, 0.2 mm for distances of 2–3 cm, and 0.5 mm for distances of 3–10 cm. Due to the cylindrical symmetry, the dose rate distribution at z and −z (at equal to y) is the same. 125I seeds, water-equivalent cylinder, and tally cells were determined in the same coordinates [Figure 1c]. Every 125I seed was simulated as a line source by the AAPM TG43 approximation, and using the MCNP F4 tally (particles/cm2), particle tracks and relative properties were recorded in each cylindrical annulus. Energy fluency was converted into dose rate using the dose energy (DE), dose function (DF), and tally multiplier (Fm) cards. The photon interaction cross-section file used in this study was the DLC-200 library, distributed by the Radiation Shielding Information Computing Center (Oak Ridge, TN, USA). Evaluation of radial and axial dose rates was performed with 108 photon histories in water to comply with good MC practice recommendations[11] regarding statistical uncertainty. The photon cutoff energy was set to 1 keV.

Statistical analysis

Statistical analysis was conducted using IBM SPSS for Windows, version 19.0 (IBM Corp., Armonk, NY, USA). Analysis of covariance was used to examine the factors influencing radial dose distribution of the radioactive stent. Two-tailed P < 0.05 was considered statistically significant.

Results

Factors influencing cumulative radial dose

For TPS results, the relative errors of the three time measurements were <3%. When the activity of the 125I seeds remained the same, the length of the stent or the number of radioactive seeds affected cumulative radial dose significantly (F = 14.704, P < 0.001). When the stent length changed from 8 cm to 6 cm, the cumulative radial dose decreased by 6–44%, and when the length changed from 6 cm to 4 cm, the decrease was 34–97%. For the same length of radioactive stent, the cumulative radial dose changed significantly (F = 35.510, P < 0.001) when the seed activity was altered by 0.1 mCi or more. When the source strength changed from 0.5 mCi to 0.4 mCi, the percentage of the cumulative radial dose reduction was the largest for the same length of radioactive stents, and the maximum dose reduction for the three radioactive stents in this study was 26%.

Comparison of the three methods

Since the dose rate is linearly related to activity of the radioactive seeds, in the TLD measurement and MC simulation, only 0.4 mCi 125I seeds were employed in this study. Our data show that the TLD's dose response in the range of 0–10 mGy irradiation by 137Cs γ-ray was linear: y = 182225x − 6651.9 (R2= 0.99152; y is the irradiation dose in mGy, x is the TLDs’ reading in nC). For the TLDs energy response, when TLDs were irradiated by different energy radiation sources to a dose of 1 mGy, reading of TLDs was different [Table 1]. For TLD results with the measured data, the calibration factor Cf for the TLD response in our work was 0.0012 mGy/nC; the relative error of the TLD measurements of radial dose rate was <6%.
Table 1

Energy response of TLD to irradiation by standard energies of different radiation sources

Radiation sourcesRelative value
33 keV (X-ray)1.23
241Am (59.5 keV)1.22
65 keV (X-ray)1.14
83 keV (X-ray)1.02
137Cs (662 keV)1.00
6 MV (X-ray)1.28

Energy response values were normalized to TLD’s reading for 137Cs irradiation. TLD: Thermoluminescent dosimeter.

Energy response of TLD to irradiation by standard energies of different radiation sources Energy response values were normalized to TLD’s reading for 137Cs irradiation. TLD: Thermoluminescent dosimeter. In this experiment, when the 4-cm stents loaded with 8 125I seeds were simulated (single seed activity, 0.4 mCi), the MCNPX-derived values of radial dose rate agreed to be within 5.7% of the TPS results and to be within 7.0% of the TLD measurements for most of the data points. For the 6-cm and 8-cm stents loaded with 16 and 24 125I seeds, respectively, the corresponding values were 6.9% and 7.0% and 7.3% and 6.7%, respectively. As shown in Figure 2, the simulations by MCNPX agreed well with the TPS calculations and TLD measurements.
Figure 2

Comparison of radial dose rates for radioactive stent. The dose rates were normalized to the dose rate at a radial distance of 10 mm. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively.

Comparison of radial dose rates for radioactive stent. The dose rates were normalized to the dose rate at a radial distance of 10 mm. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively.

3D dose rate simulated by Monte Carlo N-Particle eXtended

For the three stent models, the 3D dose rates along the axial axis as estimated by MCNPX simulations are shown in Figure 3. Using these dosimetry data, dose maps in the axial plane around a linear array of multiple 125I seeds were simulated [Figure 4]. The MC-simulated uncertainties of these three stent models were all within the limits recommended by the AAPM TG43U1. Since prescription doses were often defined at 0.5, 1.0, and 2.0 cm,[6717] depth dose calculations of stents of different lengths and with different activities of 125I seeds were performed at 0.1, 0.5, 1.0, 1.5, and 2.0 cm from the surface of the stent [Table 2].
Figure 3

Monte Carlo simulated dose rates along the axial axis of the stent. The different lines represent doses at different radial distances from the surface of the stent. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively.

Figure 4

Dose maps were calculated for distances ranging from 0.1 cm to 10 cm radially from the stent surface and axial distances of up to 10 cm from the center of the stent. (a-c) 3D dose rate distribution for stents loaded with 8, 16, and 24 125I seeds, respectively. (d-f) The uncertainty of the measurement points of the three stent models corresponding to the Monte Carlo simulation.

Table 2

Cumulative radial dose distribution of different length radioactive biliary stents in different activity of 125I seeds (Gy)

Radial distance (cm)0.4 mCi0.5 mCi0.6 mCi0.7 mCi0.8 mCi





4 cm6 cm8 cm4 cm6 cm8 cm4 cm6 cm8 cm4 cm6 cm8 cm4 cm6 cm8 cm
0.179939799117122119140146139163170158187195
0.5506367637984769411088110118101126134
1.0233236294145344957405763466572
1.5121922152327182833213338243744
2.07121491418101720122025142328

Radial distance was from stent surface. From the formula (2), when t→∞, after seed decay completely, the total cumulative dose of the interest point is equal to D0 (r, θ)/λ, λ is equal to ln2/T1/2 (T1/2= 59.4 days). D0 in this table was simulated by MCNPX. MCNPX: Monte Carlo N-Particle eXtended.

Monte Carlo simulated dose rates along the axial axis of the stent. The different lines represent doses at different radial distances from the surface of the stent. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively. Dose maps were calculated for distances ranging from 0.1 cm to 10 cm radially from the stent surface and axial distances of up to 10 cm from the center of the stent. (a-c) 3D dose rate distribution for stents loaded with 8, 16, and 24 125I seeds, respectively. (d-f) The uncertainty of the measurement points of the three stent models corresponding to the Monte Carlo simulation. Cumulative radial dose distribution of different length radioactive biliary stents in different activity of 125I seeds (Gy) Radial distance was from stent surface. From the formula (2), when t→∞, after seed decay completely, the total cumulative dose of the interest point is equal to D0 (r, θ)/λ, λ is equal to ln2/T1/2 (T1/2= 59.4 days). D0 in this table was simulated by MCNPX. MCNPX: Monte Carlo N-Particle eXtended.

Discussion

Over the past 10 years, several animal experiments[18192021] and Phase I–II clinical studies[67222324] have demonstrated the short-term efficacy and safety of intraluminal brachytherapy with 125I seed-loaded stents. The favorable results of a multicenter, single-blind, randomized, Phase III trial fueled interest in the use of intraluminal brachytherapy for intraluminal malignancies.[24] However, there are few dosimetric studies on radioactive stents, and the range of target irradiation dose reported was very broad. Thus, to allow comparisons between studies, standardization of the clinical application of 125I seed-loaded radioactive stents is urgently needed. At present, experimental TLD measurements and/or MC simulations is the standard method to accurately measure the dosimetric characteristics of radioactive brachytherapy sources. In this investigation, TG43-based brachytherapy 3D-TPS calculation was also performed for the purpose of data intercomparison. The TPS employs a point source approximation to calculate dose distributions. When using the TPS for calculation of dose, the stent and the medium around it are simulated as water, ignoring the structure inhomogeneities with only limited dose uncertainties.[25] The TPS in this study calculated the dose using point source simulations, which in fact agreed to be within about 2% with the full geometric simulations at all distances between 1 cm and 10 cm;[26] thus, within this practical range, TPS agrees very well with the TLD measurements and MCNPX. Therefore, the TPS utilized by us can be reliably used for clinical applications involving biliary stents. It has been reported that, in combination therapy of esophageal cancers with stenting followed by EBRT or brachytherapy (mainly 192Ir), dose perturbations would inevitably be caused by the esophageal stents, resulting in hot and cold spots in the esophageal mucosa.[2728] In our study, however, the dose measured by the TLD-GR 200A circular chips did not suggest significant deviation from that simulated by MC, which suggested that the metal stent did not alter the brachytherapy source dose distribution in homogeneous water medium. The symmetrically arranged three radioactive sources might overcome the dose perturbations due to the metal stent. Uncertainties associated with TLD dose rate distribution measurements are due to variation between repeated measurements; relative intrinsic energy dependence; phantom material attenuation correction factor; relative absorbed DE dependence; and TLD position relative to the sources. Due to the extremely large dose gradients and other technological limitations, doses at distances of the order of a millimeter from the radioactive stents are poorly known by TPS calculation or TLD measurement.[8] In our study, the 3D dose rate distributions around the radioactive stents were determined by MC simulation. To obtain the most accurate data, a full seed simulation was performed by MCNPX, though a point source simulation is a fast and efficient way to check a full seed simulation.[26] In DE and DF input cards, we entered a point-wise response function (the American National Standards Institute standard flux-to-dose conversion factors) to modify the regular tally card. In addition, a constant was needed for use with the Fm input card to transform the result into the dose rate (Gy/s) we required. For radial dose rate distribution, the good agreement with acceptable experimental error between the calculated and measured values in water and PMMA indicated that the correct source geometry was used in the MC simulation. Based on this agreement, the simulations were performed in water in the range of 0.1–10 cm away from and along the surface of the stents. Representing the stent surface doses, cumulative radial doses at distances of 0.1 cm from stent surface were shown in Table 2, which may be used to predict the risk of bile duct wall perforation in clinical application. In summary, our current study was purely radiophysics; TPS calculation, TLD measurement, and MC simulation were performed and were found to be in good agreement. Although the whole experiment was conducted in water-equivalent phantom, data in our evaluation may provide a theoretical basis for dosimetry for the clinical application. More instructive data may be obtained through further in vivo experiments.

Financial support and sponsorship

This study was supported by a grant of the Capital Featured Clinical Application Research Project (No. Z151100004015171).

Conflicts of interest

There are no conflicts of interest.
  28 in total

1.  Intravascular brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group no. 60. American Association of Physicists in Medicine.

Authors:  R Nath; H Amols; C Coffey; D Duggan; S Jani; Z Li; M Schell; C Soares; J Whiting; P E Cole; I Crocker; R Schwartz
Journal:  Med Phys       Date:  1999-02       Impact factor: 4.071

2.  Dosimetry of the I-Plant Model 3500 iodine-125 brachytherapy source.

Authors:  D M Duggan; B L Johnson
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

3.  The use of ¹²⁵I seed strands for intraluminal brachytherapy of malignant obstructive jaundice.

Authors:  Yi Chen; Xiao-Lin Wang; Zhi-Ping Yan; Jian-Hua Wang; Jie-Min Cheng; Gao-Quan Gong; Jian-Jun Luo
Journal:  Cancer Biother Radiopharm       Date:  2011-09-08       Impact factor: 3.099

4.  Effect of improved TLD dosimetry on the determination of dose rate constants for (125)I and (103)Pd brachytherapy seeds.

Authors:  M Rodriguez; D W O Rogers
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

5.  A novel biliary stent loaded with (125)I seeds in patients with malignant biliary obstruction: preliminary results versus a conventional biliary stent.

Authors:  Hai-Dong Zhu; Jin-He Guo; Guang-Yu Zhu; Shi-Cheng He; Wen Fang; Gang Deng; Yong-Lin Qin; Guo-Zhao Li; Douglas M Coldwell; Gao-Jun Teng
Journal:  J Hepatol       Date:  2012-01-17       Impact factor: 25.083

6.  A Phase I study of high-dose-rate intraluminal brachytherapy as palliative treatment in extrahepatic biliary tract cancer.

Authors:  Gian Carlo Mattiucci; Rosa Autorino; Andrea Tringali; Vincenzo Perri; Mario Balducci; Francesco Deodato; Maria Antonietta Gambacorta; Giovanna Mantini; Luca Tagliaferri; Massimiliano Mutignani; Alessio Giuseppe Morganti
Journal:  Brachytherapy       Date:  2015-01-13       Impact factor: 2.362

7.  A novel approach for treatment of unresectable extrahepatic bile duct carcinoma: design of radioactive stents and an experimental trial in healthy pigs.

Authors:  Yan Liu; Jun-Lou Liu; Zhen-Zhai Cai; Zheng Lu; Yan-Fang Gong; Hong-Yu Wu; Xiao-Hua Man; Zhen-Dong Jin; Zhao-Shen Li
Journal:  Gastrointest Endosc       Date:  2009-03       Impact factor: 9.427

8.  Impact of intraluminal brachytherapy on survival outcome for radiation therapy for unresectable biliary tract cancer: a propensity-score matched-pair analysis.

Authors:  Yasuo Yoshioka; Kazuhiko Ogawa; Hirobumi Oikawa; Hiroshi Onishi; Naoto Kanesaka; Tetsuro Tamamoto; Takashi Kosugi; Kazuo Hatano; Masao Kobayashi; Yoshinori Ito; Makoto Takayama; Mitsuhiro Takemoto; Katsuyuki Karasawa; Hisayasu Nagakura; Michiko Imai; Yasuhiro Kosaka; Hideya Yamazaki; Fumiaki Isohashi; Kenji Nemoto; Yasumasa Nishimura
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-07-15       Impact factor: 7.038

9.  Dose calculation formalisms and consensus dosimetry parameters for intravascular brachytherapy dosimetry: recommendations of the AAPM Therapy Physics Committee Task Group No. 149.

Authors:  Sou-Tung Chiu-Tsao; Dennis R Schaart; Christopher G Soares; Ravinder Nath
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

10.  Malignant obstructive jaundice - brachytherapy as a tool for palliation.

Authors:  Sandeep Jain; Tejinder Kataria; Shyam Singh Bisht; Deepak Gupta; Subramani Vikraman; Sanjay Baijal; Randhir Sud
Journal:  J Contemp Brachytherapy       Date:  2013-06-28
View more
  3 in total

Review 1.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

2.  Stenting of the Portal Vein Combined with Different Numbers of Iodine-125 Seed Strands: Dosimetric Analyses.

Authors:  Li-Hong Yao; Liang Su; Lu Liu; Hai-Tao Sun; Jun-Jie Wang
Journal:  Chin Med J (Engl)       Date:  2017-09-20       Impact factor: 2.628

3.  Brachytherapy Drainage Catheter and Chemotherapy for Unresectable Pancreatic Carcinoma Combined with Obstructive Jaundice.

Authors:  Dechao Jiao; Kaihao Xu; Gauri Mukhiya; Yiming Liu; Kunpeng Wu; Zongming Li; Jianzhuang Ren; Xinwei Han
Journal:  Front Oncol       Date:  2022-07-14       Impact factor: 5.738

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.