| Literature DB >> 28469106 |
Li-Hong Yao1, Jun-Jie Wang1, Charles Shang2, Ping Jiang1, Lei Lin1, Hai-Tao Sun1, Lu Liu1, Hao Liu3, Di He4, Rui-Jie Yang1.
Abstract
BACKGROUND: A novel radioactive 125I seed-loaded biliary stent has been used for patients with malignant biliary obstruction. However, the dosimetric characteristics of the stents remain unclear. Therefore, we aimed to describe the dosimetry of the stents of different lengths - with different number as well as activities of 125I seeds.Entities:
Mesh:
Year: 2017 PMID: 28469106 PMCID: PMC5421181 DOI: 10.4103/0366-6999.204936
Source DB: PubMed Journal: Chin Med J (Engl) ISSN: 0366-6999 Impact factor: 2.628
Figure 1Location map of radioactive biliary stents in a rectangular coordinate system. (a) Photograph of the radioactive biliary stent. (b) A cross-sectional view of the biliary stent. (c) Seeds’ location in a longitudinal view of a stent of 4-cm length.
Energy response of TLD to irradiation by standard energies of different radiation sources
| Radiation sources | Relative value |
|---|---|
| 33 keV (X-ray) | 1.23 |
| 241Am (59.5 keV) | 1.22 |
| 65 keV (X-ray) | 1.14 |
| 83 keV (X-ray) | 1.02 |
| 137Cs (662 keV) | 1.00 |
| 6 MV (X-ray) | 1.28 |
Energy response values were normalized to TLD’s reading for 137Cs irradiation. TLD: Thermoluminescent dosimeter.
Figure 2Comparison of radial dose rates for radioactive stent. The dose rates were normalized to the dose rate at a radial distance of 10 mm. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively.
Figure 3Monte Carlo simulated dose rates along the axial axis of the stent. The different lines represent doses at different radial distances from the surface of the stent. (a-c) Stents loaded with 8, 16, and 24 125I seeds, respectively.
Figure 4Dose maps were calculated for distances ranging from 0.1 cm to 10 cm radially from the stent surface and axial distances of up to 10 cm from the center of the stent. (a-c) 3D dose rate distribution for stents loaded with 8, 16, and 24 125I seeds, respectively. (d-f) The uncertainty of the measurement points of the three stent models corresponding to the Monte Carlo simulation.
Cumulative radial dose distribution of different length radioactive biliary stents in different activity of 125I seeds (Gy)
| Radial distance (cm) | 0.4 mCi | 0.5 mCi | 0.6 mCi | 0.7 mCi | 0.8 mCi | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 4 cm | 6 cm | 8 cm | 4 cm | 6 cm | 8 cm | 4 cm | 6 cm | 8 cm | 4 cm | 6 cm | 8 cm | 4 cm | 6 cm | 8 cm | |
| 0.1 | 79 | 93 | 97 | 99 | 117 | 122 | 119 | 140 | 146 | 139 | 163 | 170 | 158 | 187 | 195 |
| 0.5 | 50 | 63 | 67 | 63 | 79 | 84 | 76 | 94 | 110 | 88 | 110 | 118 | 101 | 126 | 134 |
| 1.0 | 23 | 32 | 36 | 29 | 41 | 45 | 34 | 49 | 57 | 40 | 57 | 63 | 46 | 65 | 72 |
| 1.5 | 12 | 19 | 22 | 15 | 23 | 27 | 18 | 28 | 33 | 21 | 33 | 38 | 24 | 37 | 44 |
| 2.0 | 7 | 12 | 14 | 9 | 14 | 18 | 10 | 17 | 20 | 12 | 20 | 25 | 14 | 23 | 28 |
Radial distance was from stent surface. From the formula (2), when t→∞, after seed decay completely, the total cumulative dose of the interest point is equal to D0 (r, θ)/λ, λ is equal to ln2/T1/2 (T1/2= 59.4 days). D0 in this table was simulated by MCNPX. MCNPX: Monte Carlo N-Particle eXtended.