Evaluation of kinetic programs in various automated perimeters.

Shigeki Hashimoto1, Chota Matsumoto2, Mariko Eura2, Sachiko Okuyama2, Yoshikazu Shimomura2.   

Abstract

PURPOSE: Kinetic programs in four automated perimeters were evaluated and compared for their clinical usefulness using four simulated visual field (VF) patterns.
METHODS: Using the results of conventional Goldmann manual kinetic perimetry (MKP), simulated fields with concentric contraction, a temporal residual island only, a small central island with a temporal island, and a ring scotoma were created. Four kinetic programs, Humphrey 750i Kinetic Test (Humphrey), OCULUS Twinfield 2 Kinetic Perimetry (OCULUS), OCTOPUS 900 Goldmann Kinetic Perimetry (OCTOPUS GKP), and Kowa AP-7000 Isopter (Kowa) were tested by the 4 simulated defect patterns using stimuli of V/4e, I/4e, I/3e, I/2e, and I/1e at speeds of 3 and 5°/s.
RESULTS: Except Humphrey, OCULUS, OCTOPUS GKP, and Kowa could obtain isopters nearly comparable to those of Goldmann MKP. However, their results were considerably influenced by the examiner's skill. Besides, Humphrey had restrictions on target presentation, and OCULUS and Kowa had problems in isopter drawing and in filling in the scotoma. OCTOPUS GKP was the only method that could correctly detect and depict all four defect patterns. It also had relatively shorter test durations among the three methods excluding Humphrey, which did not have a built-in function for test duration measurement. The perimeters' test ranges for the periphery were 90° for Humphrey, OCULUS, and OCTOPUS GKP, and 80° for Kowa.
CONCLUSION: To assess kinetic fields with various defect patterns, OCTOPUS GKP seems to be the most useful method.

Entities:  

Keywords:  Automated kinetic perimetry; Humphrey 750i Kinetic Test; Kowa AP-7000 Isopter; OCTOPUS 900 Goldmann Kinetic Perimetry; OCULUS Twinfield 2 Kinetic Perimetry

Mesh:

Year:  2017        PMID: 28444485     DOI: 10.1007/s10384-017-0516-y

Source DB:  PubMed          Journal:  Jpn J Ophthalmol        ISSN: 0021-5155            Impact factor:   2.447


  18 in total

1.  Reaction time in automated kinetic perimetry: effects of stimulus luminance, eccentricity, and movement direction.

Authors:  U Schiefer; H Strasburger; S T Becker; R Vonthein; J Schiller; T J Dietrich; W Hart
Journal:  Vision Res       Date:  2001-07       Impact factor: 1.886

2.  Peripheral visual field testing in glaucoma by automated kinetic perimetry with the Humphrey Field Analyzer.

Authors:  B J Ballon; D A Echelman; M B Shields; A R Ollie
Journal:  Arch Ophthalmol       Date:  1992-12

3.  K-Train--a computer-based, interactive training program with an incorporated certification system for practicing kinetic perimetry: evaluation of acceptance and success rate.

Authors:  U Schiefer; K Nowomiejska; E Krapp; J Pätzold; C A Johnson
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-03-21       Impact factor: 3.117

4.  Development of a new strategy of visual field testing for macular dysfunction in patients with open angle glaucoma.

Authors:  Kazuko Omodaka; Shiho Kunimatsu-Sanuki; Ryu Morin; Satoru Tsuda; Yu Yokoyama; Hidetoshi Takahashi; Kazuichi Maruyama; Hiroshi Kunikata; Toru Nakazawa
Journal:  Jpn J Ophthalmol       Date:  2013-06-29       Impact factor: 2.447

5.  Peripheral visual field testing by automated kinetic perimetry in glaucoma.

Authors:  W C Stewart; M B Shields; A R Ollie
Journal:  Arch Ophthalmol       Date:  1988-02

6.  Comparison between semiautomated kinetic perimetry and conventional Goldmann manual kinetic perimetry in advanced visual field loss.

Authors:  Katarzyna Nowomiejska; Reinhard Vonthein; Jens Paetzold; Zbigniew Zagorski; Randy Kardon; Ulrich Schiefer
Journal:  Ophthalmology       Date:  2005-08       Impact factor: 12.079

7.  Automated perimetry: background, instruments and methods.

Authors:  G L Portney; M A Krohn
Journal:  Surv Ophthalmol       Date:  1978 Jan-Feb       Impact factor: 6.048

8.  The use of semi-automated kinetic perimetry (SKP) to monitor advanced glaucomatous visual field loss.

Authors:  J Nevalainen; J Paetzold; E Krapp; R Vonthein; C A Johnson; U Schiefer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-06-18       Impact factor: 3.117

9.  Patterns of visual field progression in patients with retinitis pigmentosa.

Authors:  S Grover; G A Fishman; J Brown
Journal:  Ophthalmology       Date:  1998-06       Impact factor: 12.079

10.  The relationship between intraocular pressure and visual field progression in glaucoma.

Authors:  B C Chauhan; S M Drance
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1992       Impact factor: 3.117

View more
  1 in total

1.  Accuracy of kinetic perimetry assessment with the Humphrey 850; an exploratory comparative study.

Authors:  Fiona J Rowe; Lauren R Hepworth; Kerry L Hanna; Meera Mistry; Carmel P Noonan
Journal:  Eye (Lond)       Date:  2019-07-22       Impact factor: 3.775

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.