10-Year Follow-Up Wear Analysis of Marathon Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty.

Jared S Bookman1, Ian D Kaye1, Kevin K Chen1, Fredrick F Jaffe1, Ran Schwarzkopf1.   

Abstract

BACKGROUND: Short-term and intermediate-term wear rates for highly cross-linked polyethylene (HCLPE) liners in total hip arthroplasty (THA) are significantly lower than published rates for traditional polyethylene liners. The aim of this study was to report the longest-to-date follow-up of a specific HCLPE liner.
METHODS: A series of 35 THAs using a specific HCLPE liner were reviewed. Anteroposterior radiographs were reviewed for femoral head penetration, the presence of femoral and/or acetabular osteolysis, long-term survival, total wear, and wear rates in all patients.
RESULTS: The average patient age at time of surgery was 70 years with an average follow-up of 10 years (118 months; range, 7.2-13.4 years). The mean wear rate in our cohort was 0.07 mm/y. Total wear was 0.71 mm over the study period. No hips showed evidence of osteolysis in any zones. Survivorship at latest follow-up was 100% with all-cause revision as an end point.
CONCLUSION: The wear rate of HCLPE liners continues to be lower than published wear rates for traditional polyethylene and continues to reaffirm the acceptably low wear rates using HCLPE acetabular liner in primary THA.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  10-year follow-up; Marathon liner; highly cross-linked polyethylene; total hip arthroplasty; wear analysis

Mesh:

Substances:

Year:  2017        PMID: 28438454     DOI: 10.1016/j.arth.2017.03.016

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

Review 1.  A prospective osseointegration retrieval analysis of second generation cementless shells.

Authors:  M Curtin; E Pomeroy; M Grigoras; T Murphy; Fiachra E Rowan
Journal:  J Clin Orthop Trauma       Date:  2022-06-03

2.  Less Midterm Damage and Oxidation Are Seen in Retrieved Highly Crosslinked Ultrahigh-Molecular-Weight Polyethylene Tibial Inserts than in Direct Compression Molded Polyethylene Inserts.

Authors:  Alexandra Stavrakis; Lydia Weitzler; Timothy Wright; Douglas E Padgett
Journal:  HSS J       Date:  2018-03-26

3.  Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty.

Authors:  Omid Jalali; Travis Scudday; Marie-Claire Fickenscher; Steven Barnett; Robert Gorab
Journal:  Arthroplast Today       Date:  2020-06-02
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.