Literature DB >> 28437349

A New Navigational Tool for Pedicle Screw Placement in Patients With Severe Scoliosis: A Pilot Study to Prove Feasibility, Accuracy, and Identify Operative Challenges.

Michael Putzier1, Patrick Strube, Riccardo Cecchinato, Claudio Lamartina, Eike K Hoff.   

Abstract

STUDY
DESIGN: Pilot study.
OBJECTIVE: To develop and validate the feasibility and accuracy of a newly designed navigational spinal instrument guiding tool for patient-matched thoracic and lumbar pedicle screw placement in patients with severe scoliosis, and to identify intraoperative challenges that may be relevant. SUMMARY OF BACKGROUND DATA: Pedicle screw placement is challenging in severely rotated scoliotic spine with small diameters and asymmetrical shape of the pedicles and vertebrae. Patient-specific drill/positioning guides with preplanned trajectory have been developed as a promising solution in spinal surgery for precise screw insertion.
METHODS: In 4 patients with severe scoliosis, navigational templates and models of all vertebrae to be instrumented were manufactured using a computed tomography (CT)-based 3-dimensional model of the thoracic and lumbar spine. The guides were designed differently for thoracic and lumbar segments according to the individual anatomy to achieve an optimal coupling to the surface of the patient's spine, to maximize the stability of the device itself, and to increase user friendliness for the complete screw positioning process. Intraoperative challenges and opportunities for device and process improvements regarding the handling of the guides during the surgery were recorded. Postoperatively, the intrapedicular screw positions were evaluated versus the preoperative plan and evaluated for cortical violation based on CT scans.
RESULTS: A total of 76 pedicle screws were implanted (56 thoracic, 20 lumbar). Two screws (2.6%) were assessed to be misplaced intraoperatively and repositioned. Eighty-four percent of the pedicle screws were completely intrapedicular, 96.1% within <2 mm cortical breech. CT scans did not demonstrate medial pedicle violation, or misplaced screw contact to neurovascular structures. No screw-related clinical complaints were reported postoperatively.
CONCLUSIONS: The new custom-made positioning guide is a feasible navigational tool that permits a safe and accurate implantation of pedicle screws in patients with severe scoliosis.

Entities:  

Mesh:

Year:  2017        PMID: 28437349     DOI: 10.1097/BSD.0000000000000220

Source DB:  PubMed          Journal:  Clin Spine Surg        ISSN: 2380-0186            Impact factor:   1.876


  20 in total

1.  Are computer numerical control (CNC)-manufactured patient-specific metal templates available for posterior thoracic pedicle screw insertion? Feasibility and accuracy evaluation.

Authors:  Xiangxue Kong; Lei Tang; Qiang Ye; Wenhua Huang; Jianyi Li
Journal:  Eur Spine J       Date:  2017-07-17       Impact factor: 3.134

2.  CORR Insights®: Does Three-dimensional Printing Plus Pedicle Guider Technology in Severe Congenital Scoliosis Facilitate Accurate and Efficient Pedicle Screw Placement?

Authors:  George H Thompson
Journal:  Clin Orthop Relat Res       Date:  2019-08       Impact factor: 4.176

3.  Design of mulitlevel OLF approach ("V"-shaped decompressive laminoplasty) based on 3D printing technology.

Authors:  Qinjie Ling; Erxing He; Hanbin Ouyang; Jing Guo; Zhixun Yin; Wenhua Huang
Journal:  Eur Spine J       Date:  2017-07-27       Impact factor: 3.134

4.  Application of a novel 3D drill template for cervical pedicle screw tunnel design: a cadaveric study.

Authors:  Zhengxi Yu; Guodong Zhang; Xuanhuang Chen; Xu Chen; Changfu Wu; Yijun Lin; Wenhua Huang; Haibin Lin
Journal:  Eur Spine J       Date:  2017-06-10       Impact factor: 3.134

5.  Robotic-guided sacro-pelvic fixation using S2 alar-iliac screws: feasibility and accuracy.

Authors:  Xiaobang Hu; Isador H Lieberman
Journal:  Eur Spine J       Date:  2016-06-07       Impact factor: 3.134

6.  Accuracy of cannulated pedicle screw versus conventional pedicle screw for extra-pedicular screw placement in dysplastic pedicles without cancellous channel in adolescent idiopathic scoliosis: a computerized tomography (CT) analysis.

Authors:  Chee Kean Lee; Chris Yin Wei Chan; Siti Mariam Abd Gani; Mun Keong Kwan
Journal:  Eur Spine J       Date:  2017-08-17       Impact factor: 3.134

7.  Accuracy of pedicle screw insertion by AIRO® intraoperative CT in complex spinal deformity assessed by a new classification based on technical complexity of screw insertion.

Authors:  S Rajasekaran; Manindra Bhushan; Siddharth Aiyer; Rishi Kanna; Ajoy Prasad Shetty
Journal:  Eur Spine J       Date:  2018-01-09       Impact factor: 3.134

8.  Accurate placement of cervical pedicle screws using 3D-printed navigational templates : An improved technique with continuous image registration.

Authors:  Guodong Zhang; Zhengxi Yu; Xuanhuang Chen; Xu Chen; Changfu Wu; Yijun Lin; Wenhua Huang; Haibin Lin
Journal:  Orthopade       Date:  2018-05       Impact factor: 1.087

9.  Accuracy of thoracic pedicle screw placement in adolescent patients with severe spinal deformities: a retrospective study comparing drill guide template with free-hand technique.

Authors:  Yue Pan; G H Lü; Lei Kuang; Bing Wang
Journal:  Eur Spine J       Date:  2017-12-12       Impact factor: 3.134

10.  Lumbar and sacral pedicle screw placement using a template does not improve the midterm pain and disability outcome in comparison with free-hand method.

Authors:  Matjaz Merc; Gregor Recnik; Zmago Krajnc
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-01-19
View more