OBJECTIVE: This retrospective study sought to research the adequacy of the follow-up and optimization of cardiac implantable electronic devices (CIEDs) performed by industry representatives.Entities:
Mesh:
Year: 2017 PMID: 28430113 PMCID: PMC5512194 DOI: 10.14744/AnatolJCardiol.2017.7374
Source DB: PubMed Journal: Anatol J Cardiol ISSN: 2149-2263 Impact factor: 1.596
Baseline characteristics of 403 patients enrolled in the SUPER FOLLOW-UP study
| n | % | |
|---|---|---|
| 262 | 65.0 | |
| 67 (18–97) | ||
| Hypertension | 226 | 56.1 |
| Diabetes mellitus | 106 | 26.3 |
| Smoking | 85 | 21.1 |
| Hyperlipidemia | 202 | 50.1 |
| Coronary artery disease | 195 | 48.3 |
| Congestive heart failure | 157 | 38.9 |
| Cerebrovascular accident | 31 | 7.6 |
| 62 | 15.4 | |
| Pacemakers (n=246) | 61.0 | |
| Single chamber | 24 | 9.8 |
| AAIR | 3 | 1.2 |
| VVIR | 8.5 | |
| Dual chamber | 222 | 90.2 |
| DDD | 155 | 63.0 |
| DDDR | 67 | 27.2 |
| ICDs (n=117) | 29.0 | |
| Primary prevention | 33 | 28.2 |
| Secondary prevention | 84 | 71.8 |
| CRT-D (n=40) | 9.9 | |
| Sick sinus syndrome, tachybradycardias | 65 | 26.4 |
| Atrioventricular conduction disturbances | 155 | 63.0 |
| Atrial fibrillation with slow ventricular response | 21 | 8.5 |
| AVCD + SSS | 5 | 2.0 |
AVCD - atrioventricular conduction disturbances; CRT-D - cardiac resynchronization therapy with defibrillator; ICD - implantable cardioverter defibrillator; SSS - sick sinus syndrome
Types of erroneous CIED programming
| n | % | |
|---|---|---|
| Inconvenient active R mode | 21 | 8.5 |
| Inconvenient inactive R mode | 11 | 4.5 |
| Improper UTR/USR setting | 29 | 11.8 |
| Improper AVI setting | 20 | 8.1 |
| Inappropriate PVARP setting | 14 | 5.7 |
| Inappropriate RV pacing amplitude | 5 | 2 |
| Improper manual mode switching | 17 | 6.9 |
| | 117 | 47.6 |
| Improper base rate setting | 23 | 19.7 |
| Improper VT-VF zone determination | 7 | 6 |
| | 30 | 25.6 |
| Improper AVI setting leading to biventricular underpacing | 3 | 7.5 |
| Improper LV pacing amplitude | 2 | 5 |
| Inappropriate single RV pacing | 1 | 2.5 |
| | 6 | 15 |
| | 153/403 | 38 |
AVI - atrioventricular interval; CIED - cardiac implantable electronic device; CRT-D - cardiac resynchronization therapy with defibrillator; ICD - implantable cardioverter defibrillator; LV - left ventricle; PVARP - postventricular atrial refractory period; R - rate response; RV - right ventricle; USR - upper sensor rate; UTR - upper tracking rate; VF - ventricular fibrillation; VT - ventricular tachycardia.
leading to unnecessary RV pacing
Figure 1Percentages of erroneous programming of CIEDs
AVI - atrioventricular interval; CIED - cardiac implantable electronic device; CRT-D - cardiac resynchronization therapy with defibrillator; ICD - implantable cardioverter defibrillator; PVARP - postventricular atrial refractory period; R - rate response; RV - right ventricle; UTR/USR - upper tracking rate/upper sensor rate; VT-VF - ventricular tachycardia-ventricular fibrillation