| Literature DB >> 28427428 |
Steven A Harvey1, Sandra Incardona2, Nina Martin3, Cristina Lussiana4, Elizabeth Streat5, Stephanie Dolan4, Nora Champouillon2, Daniel J Kyabayinze2, Robert Mugerwa5, Grace Nakanwagi5, Nancy Njoki4, Ratsimandisa Rova4, Jane Cunningham6.
Abstract
BACKGROUND: Use of antigen-detecting malaria rapid diagnostic tests (RDTs) has increased exponentially over the last decade. WHO's Global Malaria Programme, FIND, and other collaborators have established a quality assurance scheme to guide product selection, lot verification, transport, storage, and training procedures. Recent concerns over the quality of buffer packaging and test accessories suggest a need to include these items in product assessments. This paper describes quality problems with buffer and accessories encountered in a project promoting private sector RDT use in five African countries and suggests steps to avoid or more rapidly identify and resolve such problems.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28427428 PMCID: PMC5397682 DOI: 10.1186/s12936-017-1820-1
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 2.979
Fig. 1Leakage from buffer vials evidenced by stained packaging (Kenya)
Fig. 2Plastic pipettes designed to collect blood by squeezing a bulb at the top (Madagascar)
Fig. 3Buffer vials showing different volumes of buffer (Madagascar)
Fig. 4Buffer vials nearly empty with buffer stuck at tip
Fig. 5Inverted cup blood collection device (FIND Geneva)
Volume of blood collected using pre-packaged blood-collection devices (inverted cup—10 devices per lot tested)
| Test lot # | Mean | SD | Min | Max | % variation |
|---|---|---|---|---|---|
| 1 (Uganda) | 2.71 | 0.56 | 1.75 | 3.50 | −46 |
| 1 (Nigeria) | 3.28 | 0.63 | 2.27 | 4.20 | −34 |
| 2 | 3.98 | 0.61 | 3.07 | 5.11 | −20 |
| 3 (Uganda) | 3.93 | 0.49 | 3.33 | 4.70 | −21 |
| 3 (Nigeria) | 4.76 | 0.33 | 4.39 | 5.31 | −5 |
| 4 | 5.33 | 0.49 | 4.29 | 6.02 | 7 |
| 5 | 5.46 | 0.53 | 4.69 | 6.12 | 9 |
| 6 | 5.43 | 0.62 | 4.80 | 6.43 | 9 |
All manufacturers specify a 5 μl fingerstick blood sample. % variation is based on the difference between this 5 μl standard and the mean volume in column 2. Lots 1 and 3 were sampled from both Uganda and Nigeria; all other lots were sampled from Nigeria only
Volume of buffer (μl) contained in pre-packaged vials (10 vials per lot tested)
| Test lot # | Specified volume | Acceptable deviation (μl) | Mean volume | SD | Min–max | Mean volume falls within acceptable deviation? | % variation |
|---|---|---|---|---|---|---|---|
| 1 (Uganda) | 70 | ±10 | 53.17 | 18.07 | 17.78–77.44 | No | −24.0% |
| 1 (Nigeria | 70 | ±10 | 43.32 | 15.77 | 21.15–66.88 | No | −38.1% |
| 2 | 130 | ±39 | 110.02 | 9.48 | 93.48–120.44 | Yes | −15.4% |
| 3 (Uganda) | 130 | ±39 | 109.08 | 5.48 | 101.18–118.35 | Yes | −16.1 |
| 3 (Nigeria) | 130 | ±39 | 113.06 | 4.25 | 107.91–121.07 | Yes | −13.0% |
| 4 | 130 | ±39 | 112.12 | 8.05 | 95.91–123.11 | Yes | −13.8% |
| 5 | 130 | ±39 | 111.34 | 5.55 | 103.67–121.53 | Yes | −14.4% |
| 6 | 130 | ±39 | 116.13 | 4.65 | 108.91–123.98 | Yes | −10.7% |
Lots 1 and 3 sampled from Uganda and Nigeria; all other lots sampled from Nigeria only. % variation is based on the difference between the manufacturer’s specified volume in column 2 and the mean volume in column 4