| Literature DB >> 28405939 |
Joakim Romson1,2, Johan Jacksén1,3, Åsa Emmer4.
Abstract
Here we present a method to manufacture peptide-concentrating MALDI-plates with alkyl ketene dimer (AKD) as a new superhydrophobic coating. The fabrication of the hydrophobic plates included application of AKD by airbrush, and negative contact printing to generate the concentration sites. Deposited sample droplets were contained within the prestructured sites, and self-adjusted onto the site if slightly misplaced. No AKD contamination was observed, and the plates could easily be cleaned and regenerated. The S/N values for four model peptides was about twice as high compared with a standard steel plate and a commercial concentration plate. Graphical Abstract ᅟ.Entities:
Keywords: Alkyl ketene dimer; Concentration plates; MALDI
Year: 2017 PMID: 28405939 PMCID: PMC5507968 DOI: 10.1007/s13361-017-1657-4
Source DB: PubMed Journal: J Am Soc Mass Spectrom ISSN: 1044-0305 Impact factor: 3.109
Figure 1Photographs of drying 0.5 μL sample droplets (100 nM of each peptide and 0.35 mg/mL HCCA) on the three MALDI plates. GS plate at the top, AC in the middle, and AKD at the bottom
Comparison of S/N Value Ratios for the Four Peptides Obtained Using the AKD Plates in Relation to the Two Other Plates (GS and AC). P is the Probability of Falsely Rejecting the Null Hypothesis
| Ratio | 0.1 nM | 0.05 nM | Overall | |||||||
| AKD/GS | P % | AKD/AC | P % | AKD/GS | P % | AKD/AC | P % | AKD/GS | AKD/AC | |
| Manual | ||||||||||
| AngII | 0.943 | >10 | 1.61 | < 5 | 1.07 | > 10 | 1.54 | < 5 | ||
| AngI | 1.32 | < 5 | 4.37 | < 5 | 1.96 | < 5 | 2.88 | < 5 | ||
| GFpB | 1.28 | > 10 | 0.775 | > 10 | 1.25 | < 5 | 1.30 | < 5 | ||
| NT | 1.66 | < 5 | 2.89 | < 5 | 2.87 | < 5 | 3.08 | < 5 | ||
| Average ratio | 1.30 | 2.41 | 1.78 | 2.20 | 1.54 | 2.31 | ||||
| SD | 0.255 | 1.36 | 0.708 | 0.788 | ||||||
| RSD % | 19.6 | 56.4 | 39.7 | 35.8 | ||||||
| Auto | ||||||||||
| AngII | 1.90 | < 5 | 2.04 | < 5 | 1.44 | < 5 | 3.03 | < 5 | ||
| AngI | 2.53 | < 5 | 1.23 | < 10 | 1.30 | > 10 | 1.18 | > 10 | ||
| GFpB | 2.38 | < 5 | 1.47 | > 10 | 1.71 | < 5 | 2.65 | < 5 | ||
| NT | 3.46 | < 5 | 1.42 | > 10 | 3.09 | < 5 | 2.04 | < 5 | ||
| Average ratio | 2.57 | 1.54 | 1.88 | 2.23 | 2.22 | 1.88 | ||||
| SD | 0.564 | 0.304 | 0.711 | 0.698 | ||||||
| RSD % | 22.0 | 19.7 | 37.7 | 31.4 | ||||||
Figure 2Spectra showing highest S/N values obtained using manual acquisition (a)–(c), and using automated acquisition (d–(f); 0.05 nM each of AngI, AngII, GFpB. and NT. (a and d) GS plate, (b and e) AC plate, (c and f) AKD plate