Abstract

In this study, biomechanical behaviors of three different screw materials (stainless steel, titanium and cobalt-chromium) have analyzed to fix with triangle fixation under axial loading in femoral neck fracture and which material is best has been investigated. Point cloud obtained after scanning the human femoral model with the three dimensional (3D) scanner and this point cloud has been converted to 3D femoral model by Geomagic Studio software. Femoral neck fracture was modeled by SolidWorks software for only triangle configuration and computer-aided numerical analyses of three different materials have been carried out by AnsysWorkbench finite element analysis (FEA) software. The loading, boundary conditions and material properties have prepared for FEA and Von-Misses stress values on upper and lower proximity of the femur and screws have been calculated. At the end of numerical analyses, the best advantageous screw material has calculated as titanium because it creates minimum stress at the upper and lower proximity of the fracture line.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28397164     DOI: 10.1007/s10856-017-5890-y

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  6 in total

Review 1.  Implant materials, designs, and surface topographies: their effect on osseointegration. A literature review.

Authors:  N Sykaras; A M Iacopino; V A Marker; R G Triplett; R D Woody
Journal:  Int J Oral Maxillofac Implants       Date:  2000 Sep-Oct       Impact factor: 2.804

2.  Modelling the fibrous tissue layer in cemented hip replacements: experimental and finite element methods.

Authors:  V Waide; L Cristofolini; J Stolk; N Verdonschot; G J Boogaard; A Toni
Journal:  J Biomech       Date:  2004-01       Impact factor: 2.712

3.  The effects of cement-stem debonding in THA on the long-term failure probability of cement.

Authors:  N Verdonschot; R Huiskes
Journal:  J Biomech       Date:  1997-08       Impact factor: 2.712

Review 4.  Magnesium and its alloys as orthopedic biomaterials: a review.

Authors:  Mark P Staiger; Alexis M Pietak; Jerawala Huadmai; George Dias
Journal:  Biomaterials       Date:  2005-10-24       Impact factor: 12.479

5.  Clinical and finite element analysis of a modular femoral prosthesis consisting of a head and stem component in the treatment of pertrochanteric fractures.

Authors:  H J Andress; S Kahl; C Kranz; P Gierer; M Schürmann; G Lob
Journal:  J Orthop Trauma       Date:  2000-11       Impact factor: 2.512

6.  The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study.

Authors:  Jérôme M Goffin; Pankaj Pankaj; A Hamish Simpson
Journal:  J Orthop Res       Date:  2012-11-08       Impact factor: 3.494

  6 in total
  3 in total

1.  Finite element modeling of proximal femur with quantifiable weight-bearing area in standing position.

Authors:  Peng Yang; Tian-Ye Lin; Jing-Li Xu; Hui-Yu Zeng; Da Chen; Bing-Lang Xiong; Feng-Xiang Pang; Zhen-Qiu Chen; Wei He; Qiu-Shi Wei; Qing-Wen Zhang
Journal:  J Orthop Surg Res       Date:  2020-09-04       Impact factor: 2.359

2.  Biomechanical characteristics of tibio-femoral joint after partial medial meniscectomy in different flexion angles: a finite element analysis.

Authors:  Xiaohui Zhang; Shuo Yuan; Jun Wang; Bagen Liao
Journal:  BMC Musculoskelet Disord       Date:  2021-04-02       Impact factor: 2.362

3.  A systematic review of the use of titanium versus stainless steel implants for fracture fixation.

Authors:  Collin C Barber; Matthew Burnham; Ogaga Ojameruaye; Michael D McKee
Journal:  OTA Int       Date:  2021-08-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.