| Literature DB >> 28383157 |
Hajime Monzen1, Kazuki Kubo1, Mikoto Tamura1, Masaru Hayakawa2, Yasumasa Nishimura3.
Abstract
We developed a novel low-radiation-absorbent lok-bar (HM-bar) that is used to secure the immobilizers to the couch. The aim of this study was to investigate the X-ray scattering and absorption properties of the HM-bar in computed tomography (CT) simulation and radiotherapy dose delivery using the Varian Exact™ lok-bar (VL-bar) as a benchmark. CT images were obtained with or without lok-bar, and then each image was visually evaluated for artifacts. The attenuation rates for each lok-bar were measured using a farmer-type ionization chamber (PTW30013) and the I'mRT phantom (IBA Dosimetry GmbH). Measurement points were between gantry angles of 110 and 180°. The treatment apparatus was a NovalisTx (Brainlab AG); X-ray energies were set at 6 MV and 10 MV. In the presence of each lok-bar, the radiation dose was measured in accordance with 10 volumetric modulated arc therapy-stereotactic body radiation therapy (VMAT-SBRT) plans for lung cancer. Artifacts were seldom observed in the CT scans of the HM-bar. The attenuation rate of each lok-bar was higher when the X-ray energy was set at 6 MV than at 10 MV. The highest attenuation rate in the VL-bar was observed at a gantry angle of 112°; the rates were 22.4% at 6 MV and 19.3% at 10 MV. Similarly, the highest attenuation rate for the HM-bar was also observed at a gantry angle of 112°; the rates were 12.2% and 10.1% at 6 MV and 10 MV, respectively. When the VL-bar was evaluated, the isocenter dose of the VMAT-SBRT plans was attenuated by 2.6% as a maximum case. In the case of the HM-bar, the maximum attenuation was 1.4%. In the measurements of each VMAT-SBRT plan, the difference of the dose attenuation rate between the VL-bar and HM-bar was approximately 1%. The HM-bar could be used to minimize the occurrence of artifacts and provide good images in CT scans regarding radiotherapy planning and dose calculation. It can be used for patient therapy at hospitals to provide accurate dose delivery because of its low X-ray scattering and absorption characteristics.Entities:
Keywords: artifacts; immobilizer; lok-bar; stereotactic body radiation therapy; volumetric modulated arc therapy
Mesh:
Year: 2017 PMID: 28383157 PMCID: PMC5689867 DOI: 10.1002/acm2.12066
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Photographs showing the HM‐bar and the VL‐bar. (a) HM‐bar (upper) and VL‐bar (lower). (b) The component of the VL‐bar that fixes to the couch. (c) The component of the HM‐bar that fixes to the couch. None of the components protrude.
Figure 2Schematic view of the calculation and measurement geometry.
Summary of the 10 VMAT‐SBRT plans including PTV volume, field size, and MU per daily prescription dose
| Plan number | PTV volume (cm3) | Field size (cm) | MU/prescription dose |
|---|---|---|---|
| 1 | 9.4 | 3.7 × 3.5 | 1489 |
| 2 | 9.3 | 3.2 × 3.4 | 2518 |
| 3 | 39.7 | 5.6 × 5.0 | 1973 |
| 4 | 39.2 | 5.0 × 5.1 | 1953 |
| 5 | 21.8 | 4.2 × 5.2 | 1446 |
| 6 | 44.3 | 6.7 × 6.8 | 1926 |
| 7 | 29.3 | 4.4 × 4.7 | 1559 |
| 8 | 29.3 | 4.5 × 5.0 | 1770 |
| 9 | 139.4 | 10.1 × 9.9 | 1643 |
| 10 | 23.5 | 6.9 × 5.2 | 1723 |
Figure 3Artifacts produced by the lok‐bars. (a) VL‐bar. (b) HM‐bar. (c) VL‐bar; difference in CT image as a result of the presence of the lok‐bar. (d) HM‐bar; difference in the CT image as a result of the presence of the lok‐bar. (e) Profile curve for the CT values regarding the subtracted image of the VL‐bar (black dotted line) shown in (c). (f) Profile curve for the CT values regarding the subtracted image of the HM‐bar on the X‐axis (black dotted line in (d)).
Computed radiation dose attenuation rate data for each gantry angle using the treatment planning software
| Attenuation rate (%) | ||||||||
|---|---|---|---|---|---|---|---|---|
| Gantry angle (degree) | VL‐bar | HM‐bar | ||||||
| AAA | Acuros XB | AAA | Acuros XB | |||||
| 6 MV | 10 MV | 6 MV | 10 MV | 6 MV | 10 MV | 6 MV | 10 MV | |
| 100 | 2.1 | 1.7 | 2.4 | 1.8 | 0.0 | 0.0 | 0.0 | 0.0 |
| 110 | 16.3 | 13.0 | 23.7 | 19.2 | 0.7 | 0.6 | 0.5 | 0.8 |
| 112 | 10.3 | 8.4 | 15.1 | 12.5 | 3.0 | 2.5 | 4.1 | 3.5 |
| 120 | 2.8 | 2.2 | 2.7 | 2.3 | 2.3 | 1.6 | 2.5 | 2.1 |
| 130 | 7.5 | 5.9 | 11.4 | 8.8 | 3.9 | 3.1 | 5.3 | 4.0 |
| 140 | 1.2 | 1.0 | 1.2 | 1.0 | 1.6 | 1.3 | 2.1 | 1.6 |
| 150 | 1.4 | 1.0 | 1.0 | 0.8 | 1.2 | 1.0 | 1.8 | 1.4 |
| 160 | 1.1 | 0.8 | 0.9 | 0.9 | 0.9 | 0.8 | 1.5 | 1.3 |
| 170 | 1.2 | 0.7 | 0.9 | 0.8 | 1.0 | 0.6 | 1.3 | 1.2 |
| 180 | 0.9 | 0.8 | 0.9 | 0.7 | 0.8 | 0.6 | 1.4 | 1.2 |
Figure 4Radiation dose attenuation rate for each lok‐bar. (a) Cases involving 6 MV X‐rays. (b) Cases involving 10 MV X‐rays.
Figure 5Comparison between the radiation dose attenuation rates calculated using the treatment planning software and the actual measurements using the dosimeter. (a) 6 MV X‐rays; VL‐bar. (b) 10 MV X‐rays; VL‐bar. (c) 6 MV X‐rays; HM‐bar. (d) 10 MV X‐rays; HM‐bar.
Influence of two lok‐bars on radiation dose attenuation using the VMAT‐SBRT plans for the treatment of lung cancer
| Plan number | Attenuation rate (%) | ||
|---|---|---|---|
| HM‐bar | VL‐bar | Difference | |
| 1 | 1.3 | 2.4 | −1.1 |
| 2 | 1.4 | 2.6 | −1.2 |
| 3 | 0.9 | 1.8 | −0.9 |
| 4 | 1.0 | 1.9 | −0.9 |
| 5 | 0.9 | 1.8 | −0.9 |
| 6 | 0.4 | 1.0 | −0.6 |
| 7 | 1.1 | 2.3 | −1.2 |
| 8 | 1.1 | 2.0 | −0.9 |
| 9 | 1.2 | 2.1 | −0.9 |
| 10 | 0.6 | 1.3 | −0.7 |
| Average | 1.0 | 1.9 | −0.9 |