Xiaoyu Xu1, Xinxing Guo1, Hui Xiao1, Lan Mi1, Xiangxi Chen1, Xing Liu2.
Abstract
BACKGROUND: To investigate the reproducibility of macular ganglion cell-inner plexiform layer (GCIPL) thickness measurement in normal eyes determined by different operators and two different raster scanning protocols of Cirrus high-definition optical coherence tomography (HD-OCT).Entities:
Keywords: Ganglion cell-inner plexiform layer; Optical coherence tomography; Protocol; Reproducibility
Mesh:
Year: 2017 PMID: 28376748 PMCID: PMC5379644 DOI: 10.1186/s12886-017-0434-2
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
GCIPL thickness (Mean ± SD, μm) measured by two operators using macular cube 512 × 128 scanning protocol
| Parameter | Operator | 1st Measurement | 2nd Measurement | 3rd Measurement |
|
|
|---|---|---|---|---|---|---|
| Average | A | 84.33 ± 5.76 | 84.27 ± 5.54 | 84.23 ± 5.63 | 0.991 | 0.846 |
| B | 84.38 ± 5.58 | 84.42 ± 5.59 | 84.39 ± 5.64 | 0.999 | ||
| Minimum | A | 81.44 ± 5.61 | 81.65 ± 5.30 | 81.23 ± 5.76 | 0.859 | 0.850 |
| B | 81.79 ± 5.72 | 81.71 ± 5.69 | 81.16 ± 5.97 | 0.689 | ||
| Superotemporal | A | 82.88 ± 5.29 | 82.82 ± 5.20 | 82.80 ± 5.42 | 0.993 | 0.938 |
| B | 82.74 ± 5.41 | 82.73 ± 5.36 | 82.71 ± 5.35 | 0.999 | ||
| Superior | A | 85.48 ± 6.15 | 85.33 ± 5.90 | 85.38 ± 5.99 | 0.983 | 0.935 |
| B | 85.31 ± 5.88 | 85.31 ± 5.89 | 85.41 ± 5.93 | 0.992 | ||
| Superonasal | A | 87.06 ± 6.57 | 86.86 ± 6.40 | 86.77 ± 6.56 | 0.943 | 0.966 |
| B | 86.76 ± 6.37 | 86.88 ± 6.30 | 86.82 ± 6.43 | 0.990 | ||
| Inferonasal | A | 84.94 ± 6.46 | 84.96 ± 6.24 | 84.91 ± 6.06 | 0.998 | 0.847 |
| B | 85.29 ± 6.13 | 85.17 ± 6.11 | 85.34 ± 6.39 | 0.978 | ||
| Inferior | A | 82.42 ± 6.25 | 82.44 ± 6.03 | 82.30 ± 6.07 | 0.982 | 0.792 |
| B | 82.72 ± 6.20 | 82.73 ± 6.29 | 82.79 ± 6.38 | 0.996 | ||
| Inferotemporal | A | 83.27 ± 5.81 | 83.38 ± 5.67 | 83.34 ± 5.86 | 0.990 | 0.916 |
| B | 83.46 ± 5.68 | 83.56 ± 5.82 | 83.53 ± 5.91 | 0.991 |
acomparison between the three repeated measurements of the GCIPL thickness measured by each operator
bcomparison of the GCIPL thickness measured by the two operators
Intra- and inter- operator reproducibility of GCIPL thickness measurements
| Operator A | Operator B | Inter-operator | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Parameter | ICC | 95% CI | CoV (%) | ICC | 95% CI | CoV (%) | ICC | 95% CI | CoV (%) |
| Average | 0.986 | 0.981–0.990 | 0.55 | 0.992 | 0.989–0.994 | 0.44 | 0.991 | 0.987–0.994 | 0.47 |
| Minimum | 0.887 | 0.849–0.918 | 1.28 | 0.875 | 0.833–0.909 | 1.37 | 0.882 | 0.827–0.919 | 1.24 |
| Superotemporal | 0.963 | 0.949–0.973 | 0.94 | 0.977 | 0.969–0.984 | 0.79 | 0.967 | 0.952–0.977 | 0.73 |
| Superior | 0.956 | 0.940–0.968 | 1.08 | 0.983 | 0.977–0.988 | 0.78 | 0.978 | 0.968–0.985 | 0.88 |
| Superonasal | 0.953 | 0.936–0.966 | 0.95 | 0.983 | 0.976–0.988 | 0.65 | 0.948 | 0.925–0.964 | 0.71 |
| Inferonasal | 0.956 | 0.940–0.968 | 0.93 | 0.973 | 0.964–0.981 | 0.86 | 0.973 | 0.961–0.982 | 0.81 |
| Inferior | 0.960 | 0.946–0.971 | 1.15 | 0.959 | 0.944–0.971 | 1.16 | 0.961 | 0.944–0.973 | 1.01 |
| Inferotemporal | 0.981 | 0.973–0.986 | 0.81 | 0.979 | 0.971–0.985 | 0.80 | 0.977 | 0.967–0.984 | 0.81 |
ICC intraclass correlation coefficient, CI confidence interval, CoV coefficients of variation
GCIPL thickness (Mean ± SD, μm) determined by two macular cube scanning protocols
| Parameter | Protocol | 1st measurement | 2nd measurement | 3rd measurement |
|
|
|---|---|---|---|---|---|---|
| Average | 512 × 128 | 84.65 ± 5.79 | 84.50 ± 5.63 | 84.46 ± 5.65 | 0.984 | 0.892 |
| 200 × 200 | 84.61 ± 5.71 | 84.70 ± 5.69 | 84.72 ± 5.61 | 0.994 | ||
| Minimum | 512 × 128 | 81.80 ± 5.33 | 81.72 ± 5.13 | 81.43 ± 5.25 | 0.927 | 0.632 |
| 200 × 200 | 81.72 ± 5.87 | 81.70 ± 5.73 | 80.91 ± 6.28 | 0.722 | ||
| Superotemporal | 512 × 128 | 83.04 ± 5.06 | 82.76 ± 5.11 | 82.85 ± 5.21 | 0.96 | 0.909 |
| 200 × 200 | 82.67 ± 5.40 | 82.87 ± 5.37 | 82.70 ± 5.05 | 0.977 | ||
| Superior | 512 × 128 | 85.83 ± 6.10 | 85.67 ± 6.22 | 86.00 ± 6.07 | 0.961 | 0.974 |
| 200 × 200 | 85.72 ± 6.03 | 85.80 ± 6.03 | 85.79 ± 5.98 | 0.997 | ||
| Superonasal | 512 × 128 | 87.96 ± 6.56 | 87.72 ± 6.24 | 87.83 ± 6.15 | 0.981 | 0.963 |
| 200 × 200 | 87.81 ± 6.39 | 87.83 ± 6.35 | 87.92 ± 6.27 | 0.995 | ||
| Inferonasal | 512 × 128 | 85.22 ± 6.33 | 85.11 ± 6.24 | 84.94 ± 6.12 | 0.973 | 0.772 |
| 200 × 200 | 85.52 ± 6.22 | 85.35 ± 6.17 | 85.70 ± 6.30 | 0.96 | ||
| Inferior | 512 × 128 | 82.50 ± 6.32 | 82.48 ± 6.07 | 82.20 ± 6.36 | 0.963 | 0.726 |
| 200 × 200 | 82.54 ± 6.51 | 82.61 ± 6.28 | 83.00 ± 6.52 | 0.923 | ||
| Inferotemporal | 512 × 128 | 83.46 ± 5.77 | 83.39 ± 5.58 | 83.28 ± 5.72 | 0.986 | 0.973 |
| 200 × 200 | 83.35 ± 5.62 | 83.52 ± 5.69 | 83.42 ± 5.64 | 0.988 |
aComparison between the three repeated measurements of the GCIPL thickness determined by each raster scanning protocol
bComparison of the GCIPL thickness determined by the two raster scanning protocols
Intra- and inter- protocol reproducibility of GCIPL thickness measurements
| Macular Cube 512 × 128 Protocol | Macular Cube 200 × 200 Protocol | Inter-protocol | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Parameter | ICC | 95% CI | CoV (%) | ICC | 95% CI | CoV (%) | ICC | 95% CI | CoV (%) |
| Average | 0.987 | 0.980–0.992 | 0.58 | 0.991 | 0.986–0.995 | 0.39 | 0.991 | 0.984–0.995 | 0.44 |
| Minimum | 0.960 | 0.938–0.975 | 1.47 | 0.951 | 0.924–0.970 | 1.04 | 0.876 | 0.786–0.928 | 1.45 |
| Superotemporal | 0.955 | 0.930–0.972 | 0.89 | 0.971 | 0.955–0.982 | 0.67 | 0.942 | 0.902–0.966 | 0.94 |
| Superior | 0.957 | 0.934–0.974 | 1.14 | 0.985 | 0.976–0.991 | 0.81 | 0.983 | 0.970–0.990 | 0.61 |
| Superonasal | 0.970 | 0.953–0.981 | 1.06 | 0.990 | 0.984–0.994 | 0.74 | 0.987 | 0.978–0.993 | 0.48 |
| Inferonasal | 0.979 | 0.968–0.987 | 1.02 | 0.980 | 0.968–0.988 | 0.88 | 0.983 | 0.971–0.990 | 0.75 |
| Inferior | 0.953 | 0.928–0.971 | 1.12 | 0.953 | 0.928–0.971 | 1.07 | 0.960 | 0.932–0.977 | 1.10 |
| Inferotemporal | 0.978 | 0.965–0.986 | 0.77 | 0.974 | 0.960–0.984 | 0.73 | 0.973 | 0.953–0.984 | 0.92 |
ICC intraclass correlation coefficient, CI confidence interval, CoV coefficients of variation
The correlation of mean signal strength and coefficient of variation (CoV) of repeated measurements of GCIPL thickness obtained by two macular cube scanning protocols
| Macula Cube 512 × 128 Protocol | Macula Cube 200 × 200 Protocol | |||
|---|---|---|---|---|
| Parameter |
|
|
|
|
| Average | −0.261 | 0.008 | −0.210 | 0.034 |
| Minimum | −0.243 | 0.014 | −0.372 | <0.001 |
| Superotemporal | −0.179 | 0.072 | −0.208 | 0.036 |
| Superior | −0.064 | 0.525 | 0.029 | 0.774 |
| Superonasal | −0.135 | 0.176 | −0.043 | 0.669 |
| Inferonasal | −0.280 | 0.004 | −0.186 | 0.062 |
| Inferior | 0.025 | 0.805 | −0.293 | 0.003 |
| Inferotemporal | −0.049 | 0.627 | −0.288 | 0.003 |
aPearson’s correlation coefficient