Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer.

Daisuke Takagi1, Akira Sawada, Tetsuya Yamamoto.   

Abstract

PURPOSE: To compare measurements from an Icare HOME rebound self-tonometer compared with those from a Goldmann applanation tonometer (GAT).
MATERIALS AND METHODS: A total of 130 subjects with suspected or confirmed glaucoma in our outpatient clinic were examined. Intraocular pressure (IOP) was measured using both an Icare HOME unit, by an ophthalmologist and by each participant, and again using a GAT, by an ophthalmologist. The Icare HOME self-measurement [Icare HOME performed by the patients (HOMEp)] was compared with those taken by the specialist using that unit and a GAT.
RESULTS: In total, 128 participants (98%, 128/130) were able to correctly conduct self-measurement. Of the 128 participants, the mean IOP was 12.2±2.8 mm Hg (range, 7 to 20 mm Hg) via GAT, 12.8±3.7 mm Hg (range, 6 to 24 mm Hg) with HOMEp and 13.1±3.8 mm Hg (range, 6 to 25 mm Hg) by Icare HOME specialist measurement [Icare HOME performed by the ophthalmologist (HOMEo)]. The mean difference between HOMEp and HOMEo was 0.21 mm Hg (P=0.068; paired t test). The mean difference between the HOMEp and GAT measurements was 0.70 mm Hg (P<0.001; paired t test), and between the HOMEo and GAT measurements it was 1.00 mm Hg (P<0.001; paired t test). The IOP difference between the HOMEp and GAT measurements was >3 mm Hg in 9.4% of cases (12/128), and >5 mm Hg in 2.3% of cases (3/128). A positive correlation was found between the central corneal thickness (CCT) and each of the 3 IOP values (HOMEp, HOMEo, and GAT: r=0.405, P<0.001; r=0.424, P<0.001; and r=0.400, P<0.001, respectively; Spearman correlation coefficient). The difference between HOMEp and GAT values was significantly increased with increasing CCT (P=0.024; Spearman correlation coefficient), indicating that a 10% increase in CCT predicted a 1.2% increase in the difference.
CONCLUSIONS: The Icare HOME tonometer is feasible for use in self-monitoring of IOP. Icare HOME tonometry measurements tend to overestimate IOP relative to GAT measurements.

Entities:  

Mesh:

Year:  2017        PMID: 28369004     DOI: 10.1097/IJG.0000000000000674

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  11 in total

Review 1.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

2.  Association Between 24-Hour Intraocular Pressure Monitored With Contact Lens Sensor and Visual Field Progression in Older Adults With Glaucoma.

Authors:  Carlos Gustavo De Moraes; Kaweh Mansouri; Jeffrey M Liebmann; Robert Ritch
Journal:  JAMA Ophthalmol       Date:  2018-07-01       Impact factor: 7.389

3.  Analysis of pattern electroretinogram signals of early primary open-angle glaucoma in discrete wavelet transform coefficients domain.

Authors:  Homa Hassankarimi; Seyed Mohammad Reza Noori; Ebrahim Jafarzadehpour; Shahin Yazdani; Fatemeh Radinmehr
Journal:  Int Ophthalmol       Date:  2019-02-06       Impact factor: 2.031

4.  Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.

Authors:  Alicja Strzalkowska; Nina Pirlich; Julia V Stingl; Alexander K Schuster; Jasmin Rezapour; Felix M Wagner; Justus Buse; Esther M Hoffmann
Journal:  J Clin Med       Date:  2022-05-18       Impact factor: 4.964

5.  Home Self-tonometry Trials Compared with Clinic Tonometry in Patients with Glaucoma.

Authors:  Elyse J McGlumphy; Aleksandra Mihailovic; Pradeep Y Ramulu; Thomas V Johnson
Journal:  Ophthalmol Glaucoma       Date:  2021-04-09

Review 6.  Diurnal Variation of IOP in Angle Closure Disease: Are We Doing Enough?

Authors:  Shibal Bhartiya; Meenakshi Wadhwani; Oshin Rai; Mariah Patuel; Syril Dorairaj; Kumar Namagiri Sirish
Journal:  Rom J Ophthalmol       Date:  2019 Jul-Sep

Review 7.  Icare Home Tonometer: A Review of Characteristics and Clinical Utility.

Authors:  John Liu; Ticiana De Francesco; Matthew Schlenker; Iqbal Ike Ahmed
Journal:  Clin Ophthalmol       Date:  2020-11-23

8.  Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma.

Authors:  Shunsuke Nakakura; Ryo Asaoka; Etsuko Terao; Yuki Nagata; Yasuko Fukuma; Satomi Oogi; Miku Shiraishi; Yoshiaki Kiuchi
Journal:  Eye Vis (Lond)       Date:  2021-07-01

Review 9.  Optical Coherence Tomography and Glaucoma.

Authors:  Alexi Geevarghese; Gadi Wollstein; Hiroshi Ishikawa; Joel S Schuman
Journal:  Annu Rev Vis Sci       Date:  2021-07-09       Impact factor: 7.745

Review 10.  Icare® rebound tonometers: review of their characteristics and ease of use.

Authors:  Shunsuke Nakakura
Journal:  Clin Ophthalmol       Date:  2018-07-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.