Jiaqi Wang, Hui Xie, Tsing Chung, Leanne Lai Hang Chan, Stella W Pang.
Abstract
Gold (Au) resistive temperature sensors were integrated on flexible polyimide-based neural probes to monitor temperature changes during neural probe implantation and stimulation. Temperature changes were measured as neural probes were implanted to infer the positions of the neural probes, and as the retina or the deep brain region was stimulated electrically. The temperature sensor consisted of a serpentine Au resistor and surrounded by four Au electrodes with 200 and [Formula: see text] diameter (dia.). The Au temperature sensors had temperature coefficient of 0.32%, and they were biocompatible and small in size. In vivo measurements of temperature changes during implantation and stimulation were carried out in the retina and deep brain region in rats. The desired implantation position was reached when temperature measured by the sensor increased to the calibrated level and became stable. There was no temperature increase when low level stimulation current of 8 and [Formula: see text] each for the two 200- and 400- [Formula: see text]-dia. electrodes, respectively, were applied. When higher level stimulation current of 100 and [Formula: see text] each were applied to the two 200- and 400- [Formula: see text]-dia. electrodes, respectively, maximum temperature increases of 1.2 °C in retina and 1 °C in deep brain region were found.Entities:
Mesh:
Year: 2017 PMID: 28362612 DOI: 10.1109/TNSRE.2016.2634584
Source DB: PubMed Journal: IEEE Trans Neural Syst Rehabil Eng ISSN: 1534-4320 Impact factor: 3.802