David Anthony Steel1, Surajit Basu2.
Abstract
BACKGROUND: Deep brain stimulation (DBS) of the subthalamic nucleus (STN) is now a key treatment choice for advanced Parkinson's disease. The optimum target area within the STN is well established. However, no emphasis on the impact of trajectory exists. The ellipsoid shape of the STN and the off-centre traditional target point mean that variation in the electrode inclination should affect STN engagement. Understanding of this relationship could inform trajectory selection during planning by improving STN engagements and margins for error.Entities:
Keywords: Deep brain stimulation; Parkinson’s disease; Subthalamic nucleus; Targeting; Trajectory
Mesh:
Year: 2017 PMID: 28361249 PMCID: PMC5486602 DOI: 10.1007/s00701-017-3151-y
Source DB: PubMed Journal: Acta Neurochir (Wien) ISSN: 0001-6268 Impact factor: 2.216
Imaging specifications
| Modality | Details |
|---|---|
| CT | Slice thickness, 1.5 mm; 120kV/300 mA; cycle time, 2 s; rotation time, 0.75 s; standard resolution |
| MRI (T2) | Slice thickness, 2 mm; voxel size, 0.78 mm × 0.78 m; TSE factor, 15; TE, 80 ms; TR, 3,000 ms; NSA, 3 |
| MRI (IR) | Slice thickness, 2 mm; voxel size, 0.72 mm A-P, 0.91 mm right-to-left; TSE factor, 7; TE, 15 ms; TR, 3,198 ms; NSA, 2; IR-delay, 400 ms |
| MRI (SWI) | Slice thickness, 0.5 mm; voxel size, 1 mm A-P, 0.99 mm right-to-left; TE, 22 ms; TR, 15 ms; NSA, 1 |
| MRI (MP-RAGE) | Voxel size, 1 mm × 1 mm |
A-P antero-posterior, IR inversion recovery, MP-RAGE magnetisation-prepared rapid gradient echo, NSA number of signal averages, SWI susceptibility weighted imaging, TE echo time, TR repetition time
Fig. 1a ‘Probe’s eye view’ used for measurement of error margin. Screenshot from a StealthStation S7 Surgical Navigation System with Framelink software (Medtronic, Minneapolis, MN, USA). b In the blown-up schematic, the centre of the dashed outline represents the probe and the solid line is the distance recorded for this trajectory
Fig. 2Obliquely cut MRI slices whose depths increase consecutively by 1 mm, transparently overlaid for illustration, such that the STN is eventually obscured and a minimum length of engagement can be determined. In this example, the recorded minimum engagement would be 6 mm. (Screenshot from a StealthStation S7 Surgical Navigation System)
Mean angular margin for error in degrees for each electrode trajectory with greyscale formatting. Trajectories that approximately overlie the coronal suture are in bold
Trajectories that approximately overlie the coronal suture are in bold
Mean engagement length in millimetres for each trajectory within the subset
In brackets are the minimum recorded engagement lengths in millimetres for each trajectory within the subset. Trajectories that approximately overlie the coronal suture are in bold
Side-effect mapping data, in terms of voltage causing capsular side effects, for eight patients whose procedures were planned using the methodology posed by this study
| Patient | ||||||||
|---|---|---|---|---|---|---|---|---|
| Lead | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
| 0 | 4.0 | 4.0 | 3.0 | - | 2.5 | - | 2.0 | - |
| 1 | 4.0 | - | 3.0 | 4.0 | 3.0 | - | 2.0 | - |
| 2 | - | 4.0 | - | 4.0 | 3.5 | - | 2.0 | - |
| 3 | - | 3.0 | - | - | - | - | 2.0 | - |
| 8 | 4.0 | - | - | - | 3.5 | - | - | - |
| 9 | - | - | - | - | 3.5 | - | - | - |
| 10 | - | - | - | - | - | - | - | - |
| 11 | 4.0 | - | - | - | - | - | - | - |
A hyphen indicates that no internal capsule stimulation was elicited by the highest voltage tested in that patient (generally 4.0 V but 3.5 V in patient 6 at contacts 1 and 11, and patient 8 at contact 2)
Programming settings for eight patients whose procedures were planned using the methodology posed by this study
| Patient | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | ||
| Left | Lead | 3 | 2 | 2 | 0 | 3 | 1 | 3 | 1 |
| Voltage (V) | 2.0 | 3.0 | 2.5 | 2.9 | 2.4 | 2.6 | 1.9 | 1.0 | |
| Pulse Width (ms) | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | |
| Frequency (Hz) | 135 | 135 | 135 | 135 | 135 | 130 | 135 | 130 | |
| Right | Lead | 10 | 10 | 9 | 8 | 10 | 9 | 10 | 9 |
| Voltage (V) | 2.7 | 3.0 | 2.5 | 2.8 | 2.1 | 2.6 | 1.9 | 1.0 | |
| Pulse Width (ms) | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | |
| Frequency (Hz) | 135 | 135 | 135 | 135 | 135 | 130 | 135 | 130 | |
Fig. 3Combined margin for error and length of engagement data, where each circle represents one trajectory. The circle diameter is proportional to mean margin for error. Within the demarcated subset the colour gradient correlates with mean engagement length. (Screenshot from a StealthStation S7 Surgical Navigation System)