Prevention of Hypoglycemia With Predictive Low Glucose Insulin Suspension in Children With Type 1 Diabetes: A Randomized Controlled Trial.

Tadej Battelino1,2, Revital Nimri3, Klemen Dovc4, Moshe Phillip3,5, Natasa Bratina4.   

Abstract

OBJECTIVE: To investigate whether predictive low glucose management (PLGM) of the MiniMed 640G system significantly reduces the rate of hypoglycemia compared with the sensor-augmented insulin pump in children with type 1 diabetes. RESEARCH DESIGN AND METHODS: This randomized, two-arm, parallel, controlled, two-center open-label study included 100 children and adolescents with type 1 diabetes and glycated hemoglobin A1c ≤10% (≤86 mmol/mol) and using continuous subcutaneous insulin infusion. Patients were randomly assigned to either an intervention group with PLGM features enabled (PLGM ON) or a control group (PLGM OFF), in a 1:1 ratio, all using the same type of sensor-augmented insulin pump. The primary end point was the number of hypoglycemic events below 65 mg/dL (3.6 mmol/L), based on sensor glucose readings, during a 14-day study treatment. The analysis was performed by intention to treat for all randomized patients.
RESULTS: The number of hypoglycemic events below 65 mg/dL (3.6 mmol/L) was significantly smaller in the PLGM ON compared with the PLGM OFF group (mean ± SD 4.4 ± 4.5 and 7.4 ± 6.3, respectively; P = 0.008). This was also true when calculated separately for night (P = 0.025) and day (P = 0.022). No severe hypoglycemic events occurred; however, there was a significant increase in time spent above 140 mg/dL (7.8 mmol/L) in the PLGM ON group (P = 0.0165).
CONCLUSIONS: The PLGM insulin suspension was associated with a significantly reduced number of hypoglycemic events. Although this was achieved at the expense of increased time in moderate hyperglycemia, there were no serious adverse effects in young patients with type 1 diabetes.
© 2017 by the American Diabetes Association.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28351897     DOI: 10.2337/dc16-2584

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  37 in total

1.  Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes.

Authors:  Sue A Brown; Boris P Kovatchev; Dan Raghinaru; John W Lum; Bruce A Buckingham; Yogish C Kudva; Lori M Laffel; Carol J Levy; Jordan E Pinsker; R Paul Wadwa; Eyal Dassau; Francis J Doyle; Stacey M Anderson; Mei Mei Church; Vikash Dadlani; Laya Ekhlaspour; Gregory P Forlenza; Elvira Isganaitis; David W Lam; Craig Kollman; Roy W Beck
Journal:  N Engl J Med       Date:  2019-10-16       Impact factor: 91.245

2.  The Incremental Benefits of the Predictive Low-Glucose Suspend Function Compared to the Low-Glucose Suspend Function as Automation Against Hypoglycemia in Sensor-Augmented Pump Therapy.

Authors:  Pilar Isabel Beato-Víbora; Estela Gil-Poch; Laura Galán-Bueno; Lucía Lázaro-Martín; Francisco Javier Arroyo-Díez
Journal:  J Diabetes Sci Technol       Date:  2018-07-30

Review 3.  Insulin Pumps.

Authors:  Rayhan Lal; Lalantha Leelarathna
Journal:  Diabetes Technol Ther       Date:  2020-02       Impact factor: 6.118

4.  Glycemic Outcomes of Use of CLC Versus PLGS in Type 1 Diabetes: A Randomized Controlled Trial.

Authors:  Sue A Brown; Roy W Beck; Dan Raghinaru; Bruce A Buckingham; Lori M Laffel; R Paul Wadwa; Yogish C Kudva; Carol J Levy; Jordan E Pinsker; Eyal Dassau; Francis J Doyle; Louise Ambler-Osborn; Stacey M Anderson; Mei Mei Church; Laya Ekhlaspour; Gregory P Forlenza; Camilla Levister; Vinaya Simha; Marc D Breton; Craig Kollman; John W Lum; Boris P Kovatchev
Journal:  Diabetes Care       Date:  2020-05-29       Impact factor: 19.112

Review 5.  Diabetes in Childhood and Adolescence.

Authors:  Ralph Ziegler; Andreas Neu
Journal:  Dtsch Arztebl Int       Date:  2018-03-02       Impact factor: 5.594

6.  Realizing a Closed-Loop (Artificial Pancreas) System for the Treatment of Type 1 Diabetes.

Authors:  Rayhan A Lal; Laya Ekhlaspour; Korey Hood; Bruce Buckingham
Journal:  Endocr Rev       Date:  2019-12-01       Impact factor: 19.871

7.  REAL-WORLD IMPLICATIONS OF HYBRID CLOSE LOOP (HCL) INSULIN DELIVERY SYSTEM.

Authors:  Eileen R Faulds; Janet Zappe; Kathleen M Dungan
Journal:  Endocr Pract       Date:  2019-03-13       Impact factor: 3.443

8.  Modern diabetes devices in the school setting: Perspectives from school nurses.

Authors:  Christine A March; Michelle Nanni; Traci M Kazmerski; Linda M Siminerio; Elizabeth Miller; Ingrid M Libman
Journal:  Pediatr Diabetes       Date:  2020-04-22       Impact factor: 4.866

9.  Predicting Nocturnal Hypoglycemia from Continuous Glucose Monitoring Data with Extended Prediction Horizon.

Authors:  Long Vu; Sarah Kefayati; Tsuyoshi Idé; Venkata Pavuluri; Gretchen Jackson; Lisa Latts; Yuxiang Zhong; Pratik Agrawal; Yuan-Chi Chang
Journal:  AMIA Annu Symp Proc       Date:  2020-03-04

Review 10.  Type 1 diabetes mellitus management in young children: implementation of current technologies.

Authors:  Michal Nevo-Shenker; Moshe Phillip; Revital Nimri; Shlomit Shalitin
Journal:  Pediatr Res       Date:  2019-11-12       Impact factor: 3.756

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.