Margin estimation and disturbances of irradiation field in layer-stacking carbon-ion beams for respiratory moving targets.

Shinya Tajiri1,2, Mutsumi Tashiro3,4, Tomohiro Mizukami2, Chihiro Tsukishima2, Masami Torikoshi1,4, Tatsuaki Kanai1,4.   

Abstract

Carbon-ion therapy by layer-stacking irradiation for static targets has been practised in clinical treatments. In order to apply this technique to a moving target, disturbances of carbon-ion dose distributions due to respiratory motion have been studied based on the measurement using a respiratory motion phantom, and the margin estimation given by the square root of the summation Internal margin2+Setup margin2 has been assessed. We assessed the volume in which the variation in the ratio of the dose for a target moving due to respiration relative to the dose for a static target was within 5%. The margins were insufficient for use with layer-stacking irradiation of a moving target, and an additional margin was required. The lateral movement of a target converts to the range variation, as the thickness of the range compensator changes with the movement of the target. Although the additional margin changes according to the shape of the ridge filter, dose uniformity of 5% can be achieved for a spherical target 93 mm in diameter when the upward range variation is limited to 5 mm and the additional margin of 2.5 mm is applied in case of our ridge filter. Dose uniformity in a clinical target largely depends on the shape of the mini-peak as well as on the bolus shape. We have shown the relationship between range variation and dose uniformity. In actual therapy, the upper limit of target movement should be considered by assessing the bolus shape.
© The Author 2017. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Radiation Oncology.

Entities:  

Keywords:  carbon-ion; gated irradiation; layer-stacking; margin; respiratory motion

Mesh:

Year:  2017        PMID: 28339740      PMCID: PMC5710598          DOI: 10.1093/jrr/rrx001

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


INTRODUCTION

The success of carbon-ion therapy with passive beam irradiation that utilizes wobbler magnets, a scatterer, a ridge filter and a multileaf collimator (MLC) has previously been demonstrated at the Heavy Ion Medical Accelerator in Chiba (HIMAC) [1], Hyogo Ion Beam Medical Center (HIBMC), Gunma University Heavy Ion Medical Center (GHMC) [2, 3] and the SAGA Heavy Ion Medical Accelerator in Tosu (SAGA HIMAT). To improve the dose localization in a target, the layer-stacking irradiation technique has been used at the HIMAC and GHMC [4, 5]. Layer-stacking irradiation can reduce unnecessary irradiation of normal tissues on the proximal side of the spread-out Bragg peak (SOBP). However, the dose uniformity may be disturbed when the target is moving. Thus, the method has not been clinically used for tissues that undergo respiratory motion, such as the lung or liver regions. It was our aim to investigate the use of this technique for a moving target. Mori et al. have reported on ensuring dose uniformity in a moving target with the layer-stacking irradiation method through the use of calculation [6, 7]. They reported that better dose uniformity was achieved with passive irradiation than with layer-stacking irradiation. Additionally, dose uniformity with gated layer-stacking irradiation was better than that achieved with free breathing [6]. Dose assessments were later conducted with carbon-ion layer-stacking irradiation, using a numeric lung phantom and changing several parameters. It was reported that shortening of motion displacement by means of a gating strategy and increasing the irradiation time was important, based on simulations [7]. Tashiro et al. have reported a method for incorporating image acquisition into the margins and smearing value of a moving target [8]. In their report, the margins and smearing value were determined by measuring the amount of movement of the target in each direction, based on 4D computed tomography (4D CT) images. In addition, the validity of the margins and the smearing value for the moving target has been verified by dose measurements in the case of broad-beam irradiation (S. Tajiri et al., submitted for publication). For a moving target, two factors need to be considered. One is the dose uniformity inside the clinical target volume (CTV), and the other is appropriate margin estimation. In this paper, we experimentally verify the dose uniformity inside the CTV and the appropriate margins for layer-stacking irradiation of a moving target.

MATERIALS AND METHODS

Layer-stacking irradiation method

Figure 1 shows a schematic illustration of the layer-stacking irradiation technique. In the layer-stacking method [5], the target is divided into several layers in the beam direction, and is sequentially irradiated, layer by layer, with a mini-peak that is produced by the ridge filter. The opening of the MLC is synchronized with the layer-stacking irradiation in accordance with the shape of the proximal part of the target.
Fig. 1.

Schematic illustration of layer-stacking irradiation.

Schematic illustration of layer-stacking irradiation. In the layer-stacking method used at our institution, the dose at each slice is assessed by measuring the peak dose before the irradiation. Then, we have to position the flat part in the depth dose distribution of the mini-peak. Figure 2 shows the physical and biological dose distributions achieved with a ridge filter. The corresponding width of the mini-peak was 5 mm. The mini-peak was swept longitudinally in steps of 2.5 mm by inserting a range shifter. The mini-peak was designed so that it overlaps at each step, because it may be resistant to movement of the target. The mini peak and the step sizes, which are clinically used in static target irradiation, were used in the measurements.
Fig. 2.

An example of physical and biological dose distribution achieved by the ridge filter. Blue and red lines are physical dose and biological dose, respectively.

An example of physical and biological dose distribution achieved by the ridge filter. Blue and red lines are physical dose and biological dose, respectively.

Margin estimation

We defined the planning target volume (PTV), a virtual CTV and a treated volume as described in a previous paper (S. Tajiri et al., submitted for publication). We measured dose distributions, changing the amplitude of the target movement, while a range compensator (RC) was placed at a fixed position. Internal margins for each direction differ with the amplitude. In order to assess the method for determining margins (see Fig. 3), we first determined the PTV for a moving target. We took the PTV to be the same as the treated volume for a static target. Then, we inversely defined a virtual CTV, using margins determined from the movement amplitude. The virtual CTV is defined by the contours resulting from subtraction of the total margin (distal: deeper region of beam axis; proximal: shallower region of beam axis; lateral: perpendicular to the beam axis) from the PTV. The total margin in the lateral direction therefore corresponds to the smearing values.
Fig. 3.

Schematic illustration of determination of the IM. The blue arrows show motion in the superior–inferior (SI), left–right (LR) and anterior–posterior (AP) directions. The red arrows show the IM in each direction.

Schematic illustration of determination of the IM. The blue arrows show motion in the superior–inferior (SI), left–right (LR) and anterior–posterior (AP) directions. The red arrows show the IM in each direction. For heavy ion therapies, the clinical dose distribution, which is defined as the physical dose multiplied by the relative biological effectiveness (RBE), should be considered. In order to consider a change in the clinical dose due to movement of the target, the beam quality or linear energy transfer should be measured. We did not consider change in the beam quality, and we assumed that the variation in the physical dose would approximate that of the clinical dose. In our experiment, the center of a spherical target 93 mm in diameter was placed at a depth of 120.5 mm in water. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target [9-11]. The RC was designed for static irradiation. We then measured dose distributions in a respiratory motion–phantom with the RC. Based on the work of Tashiro et al. [8], determination of the internal margin (IM) is schematically illustrated in Fig. 3 (S. Tajiri et al., submitted for publication). The blue arrows show motion in the superior–inferior (SI), left–right (LR) and anterior–posterior (AP) directions, and the red arrows show the IM in each direction. Following our treatment planning policy, the IM in each direction was set by adding one-third of the motion during the gating window in each direction. The total margin (TM) in the superior, inferior, left, right, anterior and posterior directions was calculated as follows: where the setup margin (SM) was taken as ~3 mm in each direction for treatment planning. This value was taken as the square root of the summation such as the positional error, displacement of the beam axis, the positions of the RC and MLC, the machining accuracy of the RC, the uniformity of the RC material, and beam range uncertainties [8]. In the measurements, a multilayer ionization chamber (MLIC) [12] was positioned by a laser pointer instead of by an X-ray positioning system. The SM was assumed to be 2 mm in the lateral direction and 3 mm in the beam direction. This is because the MLIC had a few percentage errors in water-equivalent length values of electrodes, an uncertainty in position, etc. Then, we defined the virtual CTV as the PTV minus the total margins [8].

Conversion of lateral movement to range variation

The thickness of the RC used in this experiment increased with the off-center distance. Then, lateral movement of the target was converted to range variation due to movement of the target, assuming the straight line approximation of the beam path. The central part of the target has smaller changes of the beam direction, but the peripheral region has larger changes. Figure 4 illustrates the schematic of the beam range variation due to target motion. For a moving target, the two arrows show the same irradiated position in the target; however, the beam range in the target becomes shorter when the target moves from inferior to superior. This is because the path lengths in the RC become thicker.
Fig. 4.

Schematic illustration of the beam range variation due to respiratory motion.

Schematic illustration of the beam range variation due to respiratory motion.

Experimental conditions

We measured dose distributions with and without respiratory motion by using the MLIC [12] that was mounted on a laterally moving stage. The MLIC consists of 64 small ionization chambers stacked with ~4-mm water-equivalent intervals in the beam direction. A beam energy of 400 MeV/n was used. A prescribed dose of 5 Gy (the RBE) was administered to the target from a horizontal irradiation port. The dose measurement setup is shown in Fig. 5. The static target was located on the beam line passing through the iso center (IC). The target moved towards the right (the superior direction) with respect to the beam direction; also, the movement of the beam direction was simulated by changing the thickness of a wedge-shaped phantom. We measured dose distributions at various movement amplitudes (1, 4 and 20 mm), for beam intensities of 100% and 50%, and for free breathing and gated irradiation, as summarized in Table 1. In order to compare free breathing with gated irradiation under similar conditions, the gating windows of the lateral and beam directions were set to 4 mm and 0.8 mm, respectively, when the total amplitudes of lateral and beam direction were 10 mm and 1 mm, respectively. The beam was gated only for lateral movement, using a respiratory gating system (Anzai Medical AZ-733V with laser respiration sensor, Tokyo, Japan) equipped with a beam delivery system (Mitsubishi Electric, Tokyo, Japan). The dose rate of 100% beam intensity at the IC position was 6.18 Gy/min. We measured dose distributions using the respiratory gating system for gated irradiation. The values of margins are also summarized in Table 1.
Fig. 5.

Schematic illustration of a respiratory motion phantom

Table 1.

Breathing amplitude and margin

Case no.Phantom motionParameterBeam directionLateral direction
ProximalDistalSame lateral direction of respiratory motionOpposite lateral direction of respiratory motion
1Static0000
2Free breathing (100% beam intensity)Breathing amplitude [mm]0110
Margin3.03.32.42.0
3Free breathing (100% beam intensity)Breathing amplitude [mm]0140
Margin3.03.35.72.4
4Free breathing (100% beam intensity)Breathing amplitude [mm]01200
Margin3.03.326.77.0
5Free breathing (50% beam intensity)Breathing amplitude [mm]0140
Margin3.03.35.72.4
6Gated irradiation (50% beam intensity)Breathing amplitude [mm]00.8 (Amplitude is 1)4 (Amplitude is 10)0
Margin3.03.35.72.4
Breathing amplitude and margin Schematic illustration of a respiratory motion phantom The lateral movement of the MLIC was synchronized with the wedge-shaped phantom. The respiratory cycle was set to an arbitrary value of 3.5 s, which was not the same as the synchrotron operation cycle. Respiratory motion was assumed to be sinusoidal. The operation cycle of the synchrotron was 2.9 s, and the spill cycle length was 0.97 s. In our measurements, beam direction and lateral direction were defined by the LR direction and the SI direction.

RESULTS

Margin assessment

For comparison of dose distributions between the static target and the targets moving due to respiration, the relative dose ratio was evaluated as shown below: where was the measured physical dose for a static target, and was the measured dose for a moving target. Figure 6 shows the dose ratio between the static target (Case no. 1 in Table 1) and the moving target (Case nos 2–4 in Table 1). The black region represents the area in which the variation of the dose ratio exceeds 5%. The colored region depicts the area in which the variation of the dose ratio falls within 5%. The green dotted line is the PTV that is defined by the treated volume for a static target, while the red-brown dotted line is the virtual CTV calculated from the PTV and moving distances. The treated volume for a moving target in which the variation of the dose ratio is more than ±5% (the black region) was observed at the edges of the virtual CTV. For this reason, the dose weight among all layers was markedly higher in the most distal layer, which resulted in severe interplay effect at this layer. Coverage of the treated volumes for the moving targets relative to the defined virtual CTVs were 97.6%, 95.0% and 88.0% in cases of lateral motion of 1 mm, 4 mm and 20 mm, respectively.
Fig. 6.

Relative dose ratio between the static target (Case no. 1 in Table 1) and the moving target in case of (a) 1 mm (Case no. 2 in Table 1), (b) 4 mm (Case no. 3 in Table 1) and (c) 20 mm (Case no. 4 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target.

Relative dose ratio between the static target (Case no. 1 in Table 1) and the moving target in case of (a) 1 mm (Case no. 2 in Table 1), (b) 4 mm (Case no. 3 in Table 1) and (c) 20 mm (Case no. 4 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target. Figure 7 shows the average dose ratio and standard deviation in the above virtual CTV at each lateral position. Red, blue and green dotted lines indicate the average dose ratio for 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1) of the moving target, respectively. The vertical bars of Fig. 7 indicate the dispersion of deviations (2σ) inside the virtual CTV at each lateral position. Figure 7 also shows that the slope of the average dose ratio increased as respiratory motion increased.
Fig. 7.

Average relative dose ratio in the virtual CTV at each lateral position. Red, blue and green lines are 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1), respectively.

Average relative dose ratio in the virtual CTV at each lateral position. Red, blue and green lines are 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1), respectively.

Assessment of range variation

Figure 8 shows (a) the standard deviation (2σ) of the dose ratio and (b) the average dose ratio in the virtual CTV, where the abscissa is the range variation converted from a change in the lateral position. These results were obtained for a lateral movement of 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1). The red and blue circles in this figure indicate the central region and peripheral region in the virtual CTV, respectively. Here, the peripheral regions were defined by regions 10 mm inside the edge of the virtual CTV.
Fig. 8.

(a) Standard deviation and (b) average dose ratio in the virtual CTV, where the abscissa shows the range variation. The red and blue dotted circles indicate the central region and peripheral region in the virtual CTV, respectively. In Fig. 8a, the solid black line is the average standard deviation. In Fig. 8b, the solid black line is the average dose ratio, and the black dashed lines are the acceptable dose ratio.

(a) Standard deviation and (b) average dose ratio in the virtual CTV, where the abscissa shows the range variation. The red and blue dotted circles indicate the central region and peripheral region in the virtual CTV, respectively. In Fig. 8a, the solid black line is the average standard deviation. In Fig. 8b, the solid black line is the average dose ratio, and the black dashed lines are the acceptable dose ratio. In Fig. 8a, the solid black line represents the linear approximation of the standard deviation (average standard deviation) from all cases. Although the standard deviations seem to be dispersed, the red circles as a whole are not. In Fig. 8b, the solid black line is also a linear approximation of the dose ratio (average dose ratio) for all cases, to evaluate the trend of the dose ratio per range variation. The black dashed lines indicate the maximum and minimum dose ratios that were obtained by the standard deviation of the dose ratio. From this figure, dose uniformity became worse as range variation increased. The dose ratios were within 5% in the region of a 3-mm range variation.

Effect of beam intensity and gated irradiation

Figure 9 shows the dose ratio between the static target and the moving target (Case nos 5 and 6 of Table 1) for 50% beam intensity with ungated and gated irradiation. The black region represents the area in which the variation in the dose ratio exceeds 5%. The colored region shows the area in which the variation in the dose ratio falls within 5%. The green dotted line is the PTV obtained by a 95% dose area of the prescribed dose for a static target, and the red-brown dotted line is the virtual CTV calculated from the PTV of the static irradiation and margins calculated from the target movements.
Fig. 9.

Relative dose ratio between the static target (Case no. 1 in Table 1) and the target moving as a result of respiration in the case of (a) 50% beam intensity of free breathing (Case no. 5 in Table 1) and (b) 50% beam intensity of gated irradiation (Case no. 6 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target.

Relative dose ratio between the static target (Case no. 1 in Table 1) and the target moving as a result of respiration in the case of (a) 50% beam intensity of free breathing (Case no. 5 in Table 1) and (b) 50% beam intensity of gated irradiation (Case no. 6 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target. When comparing free breathing (Fig. 9a) with gated irradiation (Fig. 9b) conditions, it can be seen that the green regions, in which the dose ratios were almost 100%, were increased for gated irradiation. Figure 10 shows the average dose ratio and the standard deviation in the virtual CTV at each lateral position. A comparison of a 100% (Case no. 3 in Table 1) with a 50% (Case no. 5 in Table 1) beam intensity shows similar results, despite the 50% reduction in beam intensity. The slope of the average dose ratio of gated irradiation (Case no. 6 in Table 1) was the most gradual. The virtual CTV could not cover the treated volume for the moving target with a variation in dose ratio within 5%. The coverage of the treated volumes for the moving targets relative to the virtual CTVs were 95.0%, 94.8% and 97.2% in Case nos 3, 5 and 6 in Table 1, respectively.
Fig. 10.

Average relative dose ratio in the virtual CTV at each lateral position. Red, blue, and green lines are Case nos 3, 5 and 6 in Table 1, respectively.

Average relative dose ratio in the virtual CTV at each lateral position. Red, blue, and green lines are Case nos 3, 5 and 6 in Table 1, respectively.

DISCUSSION

We experimentally verified the dose uniformity inside the CTV with respect to margin, range variation, and the effect of beam intensity, as well as the appropriate margins for a moving target, and gated irradiation. We noted that the dose uniformity deteriorated as respiratory motion increased. Moreover, the peripheral region in the virtual CTV showed markedly disturbed dose uniformity as compared with the central region in the CTV. Additionally, dose uniformity was not affected by a reduction in the beam intensity, even by half. On the other hand, the dose uniformity was improved by gated irradiation under a similar amplitude. This is because the positional variation in the beam direction in the case of the gated irradiation was lower than that in the case of free breathing, even for the same lateral positional variation, and the fraction of time in which the target exists at the greatest exhalation phase for the gated irradiation was larger than for the free breathing case [8]. Figure 11 shows the range variation in the virtual CTV, where the abscissa indicates the lateral positions. Red, blue and green lines indicate the lateral amplitudes of 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1), respectively. Black dashed lines indicate 3.3 mm from the distal margin. Colored vertical lines represent the region of the virtual CTV in each lateral direction. The range variations partly exceeded the distal margin at the peripheral region for the 1 mm and 4 mm illustrations. Therefore, additional margins may be necessary in layer-stacking irradiation of a moving target to improve dose uniformity and coverage.
Fig. 11.

Range variation in the virtual CTV, where the abscissa is the each lateral position. Red, blue and green lines are 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1), respectively. Colored vertical lines indicate the region of the virtual CTV in each lateral direction.

Range variation in the virtual CTV, where the abscissa is the each lateral position. Red, blue and green lines are 1 mm (Case no. 2 in Table 1), 4 mm (Case no. 3 in Table 1) and 20 mm (Case no. 4 in Table 1), respectively. Colored vertical lines indicate the region of the virtual CTV in each lateral direction. The edge of the target (distal layer) was affected by respiration, as the most distal layer, which was the highest dose weight, was most affected by motion of the target motion. Therefore, layer-stacking irradiation of a moving target requires use of additional margins in the distal and lateral direction in order to obtain the required dose uniformity. The TM for layer-stacking irradiation should thus be better calculated as follows: where additional margin is set to the slice thickness of 2.5 mm. The solid black lines in Fig. 11 indicate the distal margin, including the additional margin. Compared with the distal margin and range variations, 1 mm and 4 mm were roughly adequate as the range variation when adding the additional margin. For 20 mm, however, the range variations were far from the distal margin, even when the additional margin was added. Figure 12 shows (a) the standard deviation and (b) the average dose ratio in the virtual CTV, when including the additional margin, as a function of range variation. Similar to Fig. 8, red and blue dotted circles indicate the central region and peripheral region in the virtual CTV, respectively. As shown in Fig. 12a, the standard deviations were decreased as compared with those in Fig. 8a, where the additional margin was not included. In Fig. 12b, the variation in the dose ratio fell within 5% for a range variation of 5 mm. When the additional margin of 5.0 mm was included, the variation in the dose ratio fell within the 5% for a range variation of 6 mm. When comparing the data shown in Fig. 8 with those in Fig. 12, it is clear that the additional margin improved the acceptable range variation, in which the variation of the dose ratio was within 5%. The additional margin for the moving target should be determined based on the range variation. However, unlimited large margins may not be acceptable in clinical situations. Therefore, gated layer-stacking irradiation should be applied with consideration of the available amount of additional margin and the patient's clinical condition.
Fig. 12.

(a) The standard deviation and (b) the average dose ratio in the virtual CTV using Eq. (3), where the abscissa is the range variation. The red and blue dotted circles indicate the central region and peripheral region in the virtual CTV, respectively.

(a) The standard deviation and (b) the average dose ratio in the virtual CTV using Eq. (3), where the abscissa is the range variation. The red and blue dotted circles indicate the central region and peripheral region in the virtual CTV, respectively. Figure 13 shows the dose ratio between the static target and the moving target in case of (a) 4 mm of free breathing (Case no. 5 in Table 1), (b) 4 mm of gated irradiation (Case no. 6 in Table 1) and (c) 20 mm of free breathing (Case no. 4 in Table 1) with an additional margin of 2.5 mm. The virtual CTV (red-brown dotted line) was obtained from the PTV using the margins determined by Eq. (3). As shown in Fig. 13a and b, the treated volume for the moving target in which the variation of the dose ratio is within ±5% covers the virtual CTV. On the other hand, the black region, in which the variation of the dose ratio exceeded ±5% is shown in Fig. 13c. These results indicate that layer-stacking irradiation of a moving target requires larger margins or an upper-limit of movement.
Fig. 13.

Relative dose ratio between the static target and the moving target in the case of (a) 4 mm of free breathing (Case no. 5 in Table 1), (b) 4 mm of gated irradiation (Case no. 6 in Table 1) and (c) 20 mm of free breathing (Case no. 4 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target.

Relative dose ratio between the static target and the moving target in the case of (a) 4 mm of free breathing (Case no. 5 in Table 1), (b) 4 mm of gated irradiation (Case no. 6 in Table 1) and (c) 20 mm of free breathing (Case no. 4 in Table 1). Green and red-brown dotted lines are the PTV and the virtual CTV, respectively. The PTV is given by the treated volume for the static target that covers 95% of the prescribed dose for a static target. The coverages of the treated volumes for the moving targets relative to the virtual CTVs are summarized in Fig. 14. For the case of an additional margin of 2.5 mm, the coverages of the treated volumes for the moving targets relative to the virtual CTVs were improved to 98.2%, 99.2% and 92.3% in cases of (a) 4 mm of free breathing, (b) 4 mm of gated irradiation and (c) 20 mm of free breathing, respectively. When adding a further margin of 5.0 mm, the coverage was 99.8%, 99.9% and 95.6% in cases of (a) 4 mm of free breathing, (b) 4 mm of gated irradiation and (c) 20 mm of free breathing, respectively. This coverage may be acceptable, even with 20 mm of motion. However, the value of the coverage depends on the shape of the target. Therefore, the relationship between range variation and dose uniformity should be considered.
Fig. 14.

The coverages [%] of the treated volumes for the moving targets relative to the virtual CTVs.

The coverages [%] of the treated volumes for the moving targets relative to the virtual CTVs. As seen in Fig. 11, use of a constant additional margin may be too large in a central region because the range variation in this area is small. Smearing can be applied to cover the range to the target for lateral positional errors [13, 14]. Thus, the change in the range due to motion may also be compensated for by setting smearing instead of using an additional margin, if the smearing values in the lateral direction of respiratory motion are set to the TMs that represent the maximum range variation inside the virtual CTV. The treated volumes for the moving targets could cover the virtual CTVs to some degree without use of an additional margin. In order to improve dose uniformity, an additional margin is needed for layer-stacking irradiation. The suggested additional margin of 2.5 mm is not a rigid value, but it is an appropriate value in some range variations for our particular ridge filter. For large lateral movements (20 mm), an additional margin of 2.5 mm would be inadequate, as compared with small lateral movements (4 mm). In clinical practice, there should be an upper limit of movement amplitude. Lateral movement of a target converts to the range variation accompanied by the movement of the target. Target uniformity of 5% is achieved when the upward range variation is limited and when using an additional margin. Although better dose uniformity could be obtained by an additional margin, it may be better to simulate the dose distributions for moving targets, as some high doses appear outside the target, where organs at risk may be present. In order to decrease high doses in such organs at risk, the moving distance should be limited. Dose uniformity is achieved by superposition of slices. In this case, dose uniformity in a clinical target would largely depend on the shape of the mini-peak as well as on the bolus shape. In this report, we have expounded the relationship between range variation and dose uniformity when using a mini-peak with a fixed shape. In an actual therapy case, we would consider using an upper limit of target movement based on the shape of the bolus.
  11 in total

1.  Treatment planning for the layer-stacking irradiation system for three-dimensional conformal heavy-ion radiotherapy.

Authors:  Nobuyuki Kanematsu; Masahiro Endo; Yasuyuki Futami; Tatsuaki Kanai; Hiroshi Asakura; Hiroyoshi Oka; Ken Yusa
Journal:  Med Phys       Date:  2002-12       Impact factor: 4.071

2.  Four-dimensional proton treatment planning for lung tumors.

Authors:  Martijn Engelsman; Eike Rietzel; Hanne M Kooy
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-04-01       Impact factor: 7.038

3.  Four-dimensional lung treatment planning in layer-stacking carbon ion beam treatment: comparison of layer-stacking and conventional ungated/gated irradiation.

Authors:  Shinichiro Mori; Nobuyuki Kanematsu; Hiroshi Asakura; Gregory C Sharp; Motoki Kumagai; Suguru Dobashi; Mio Nakajima; Naoyoshi Yamamoto; Susumu Kandatsu; Masayuki Baba
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-23       Impact factor: 7.038

4.  New accelerator facility for carbon-ion cancer-therapy.

Authors:  Koji Noda; Takuji Furukawa; Takashi Fujisawa; Yoshiyuki Iwata; Tatsuaki Kanai; Mitsutaka Kanazawa; Atsushi Kitagawa; Masataka Komori; Shinichi Minohara; Takeshi Murakami; Masayuki Muramatsu; Shinji Sato; Yuka Takei; Mutsumi Tashiro; Masami Torikoshi; Satoru Yamada; Ken Yusa
Journal:  J Radiat Res       Date:  2007       Impact factor: 2.724

5.  Four-dimensional layer-stacking carbon-ion beam dose distribution by use of a lung numeric phantom.

Authors:  Shinichiro Mori; Motoki Kumagai; Kentaro Miki
Journal:  Radiol Phys Technol       Date:  2015-04-02

6.  Biophysical characteristics of HIMAC clinical irradiation system for heavy-ion radiation therapy.

Authors:  T Kanai; M Endo; S Minohara; N Miyahara; H Koyama-ito; H Tomura; N Matsufuji; Y Futami; A Fukumura; T Hiraoka; Y Furusawa; K Ando; M Suzuki; F Soga; K Kawachi
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-04-01       Impact factor: 7.038

7.  Automated range compensation for proton therapy.

Authors:  M S Wagner
Journal:  Med Phys       Date:  1982 Sep-Oct       Impact factor: 4.071

8.  Development of a multi-layer ionization chamber for heavy-ion radiotherapy.

Authors:  Kaori Yajima; Tatsuaki Kanai; Yohsuke Kusano; Takuya Shimojyu
Journal:  Phys Med Biol       Date:  2009-03-17       Impact factor: 3.609

9.  Technical approach to individualized respiratory-gated carbon-ion therapy for mobile organs.

Authors:  Mutsumi Tashiro; Takayoshi Ishii; Jun-ichi Koya; Ryosuke Okada; Yuji Kurosawa; Keisuke Arai; Satoshi Abe; Yoshiaki Ohashi; Hirofumi Shimada; Ken Yusa; Tatsuaki Kanai; Satoru Yamada; Hidemasa Kawamura; Takeshi Ebara; Tatsuya Ohno; Takashi Nakano
Journal:  Radiol Phys Technol       Date:  2013-04-09

10.  Carbon Ion Radiotherapy at the Gunma University Heavy Ion Medical Center: New Facility Set-up.

Authors:  Tatsuya Ohno; Tatsuaki Kanai; Satoru Yamada; Ken Yusa; Mutsumi Tashiro; Hirofumi Shimada; Kota Torikai; Yukari Yoshida; Yoko Kitada; Hiroyuki Katoh; Takayoshi Ishii; Takashi Nakano
Journal:  Cancers (Basel)       Date:  2011-10-26       Impact factor: 6.639

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.