Hyun-Ah Lee1, Heon-Jeong Lee2, Joung-Ho Moon3, Taek Lee4, Min-Gwan Kim4, Hoh In4, Chul-Hyun Cho1, Leen Kim1.
Abstract
OBJECTIVE: The purpose of this study was to evaluate the applicability of data obtained from a wearable activity tracker (Fitbit Charge HR) to medical research. This was performed by comparing the wearable activity tracker (Fitbit Charge HR) with actigraphy (Actiwatch 2) for sleep evaluation and circadian rest-activity rhythm measurement.Entities:
Keywords: Actigraphy; Circadian rest-activity rhythm; Comparison; Sleep; Wearable activity tracker
Year: 2017 PMID: 28326116 PMCID: PMC5355016 DOI: 10.4306/pi.2017.14.2.179
Source DB: PubMed Journal: Psychiatry Investig ISSN: 1738-3684 Impact factor: 2.505
Comparison of sleep start time between Actiwatch 2 and Fitbit Charge HR (N=16)
*p<0.05, **p<0.01, †time had passed from 18:00 until the sleep start time, ‡correlation coefficient (r=0.00–0.30: negligible correlation, r=0.30–0.50: low correlation, r=0.50–0.70: moderate correlation, r=0.70–0.90: high correlation, r=0.90–1.00: very high correlation). N: number of participants
Figure 1Scatter plot of sleep start time between Actiwatch 2 and Fitbit Charge HR. Sleep start time was compared between the two devices for all participants. Each of sleep start time was assessed from two different devices. For analysis, sleep start time was calculated by time had passed from 18:00, and converted to minutes. Each number of axis x and y indicate the sleep start time, that is total time (minutes) has passed from 18:00 to sleep start time. The black spots are all values of sleep start time between Actiwatch 2 and Fitbit Charge HR. The solid line indicates correlation coefficient of Spearman's correlation (r=0.869, p<0.01).
Comparison of sleep duration between Actiwatch 2 and Fitbit Charge HR (N=16)
*p<0.05, **p<0.01, †correlation coefficient (r=0.00–0.30: negligible correlation, r=0.30–0.50: low correlation, r=0.50–0.70: moderate correlation, r=0.70–0.90: high correlation, r=0.90–1.00: very high correlation). N: number of participants
Figure 2Scatter plot of sleep duration between Actiwatch 2 and Fitbit Charge HR. Sleep duration was compared between the two devices for all participants. Each of sleep duration was assessed from two different devices. Each number of axis x and y indicate the sleep duration. The black spots are all values of sleep duration between Actiwatch 2 and Fitbit Charge HR. The solid line indicates correlation coefficient of Spearman's correlation (r=0.918, p<0.01).
Comparison of circadian rest-activity rhythm using a Cosinor periodogram between Actiwatch 2 and Fitbit Charge HR (N=15)
**p<0.01, †correlation coefficient (r=0.00–0.30: negligible correlation, r=0.30–0.50: low correlation, r=0.50–0.70: moderate correlation, r=0.70–0.90: high correlation, r=0.90–1.00: very high correlation). N: number of participants