| Literature DB >> 28314162 |
Toshioh Fujibuchi1, Takatoshi Toyoda2, Shingo Baba3, Yoshiyuki Umezu4, Isao Komiya4, Masayuki Sasaki5, Hiroshi Honda3.
Abstract
The distribution of activation inside a compact medical cyclotron was evaluated by measuring 1cm dose equivalent rates and γ-ray spectra. Analysis of the distribution of activation showed high activation at the deflector and the magnetic channel. Radionuclides 60Co, 57Co, 65Zn, and 54Mn were detected. Different radionuclides were generated from different components of the cyclotron, and low-activity radionuclides could be detected under low-background-radiation conditions.Keywords: Activation; Compact medical cyclotron; Neutron; Radioactive concentration
Year: 2017 PMID: 28314162 DOI: 10.1016/j.apradiso.2017.02.045