Tonometer validation and intraocular pressure reference values in the normal chinchilla (Chinchilla lanigera).

Kevin C Snyder1,2, Andrew C Lewin1,2, Christoph Mans1, Gillian J McLellan1,2,3.   

Abstract

OBJECTIVES: To determine accuracy and precision of three commonly used tonometers (TonoVet® and TonoLab® (ICare Oy, Finland)-rebound tonometers, and Tono-Pen VET™ (Reichert, NY)-applanation tonometer) in normal chinchillas, and to establish a normal intraocular pressure (IOP) reference range in this species.
METHODS: The anterior chambers of three chinchilla eyes were cannulated ex vivo and readings obtained at manometric IOPs from 5 to 80 mmHg, using each of the three tonometers in random order. Data were analyzed by linear regression, ANOVA, and Bland-Altman plots. Tonometry was performed in both eyes of 60 chinchillas (age 8 weeks-16.2 years) using the TonoVet® and relationship between age and IOP analyzed using linear regression. For all statistical tests, P < 0.05 was significant.
RESULTS: Intraocular pressure values obtained using the Tono-Pen VET™ and TonoVet® (in dog calibration mode;'d') showed strong linear correlation with manometry within the physiologic and clinically relevant range of IOP (0-50 mmHg). The TonoVet® 'd' setting displayed significantly greater precision over the full range of IOP evaluated than the Tono-Pen VET™, and both TonoVet and Tono-Pen VET™ were significantly more accurate than the TonoLab® tonometer. Mean ± SD IOP (TonoVet® 'd') in chinchillas was 9.7 ± 2.5 mmHg, and the 95% reference interval was 4.7-14.7 mmHg.
CONCLUSIONS: Both the Tono-Pen VET™ and TonoVet® provided clinically acceptable estimates of IOP in chinchillas. The TonoVet® provides accurate and precise IOP values, while Tono-Pen VET™ derived measurements showed greater variability. Values obtained either with the TonoLab® or TonoVet® used in the 'unspecified' calibration setting were inaccurate in this species.
© 2017 American College of Veterinary Ophthalmologists.

Entities:  

Keywords:  Tono-Pen Vet; TonoLab; TonoVet; chinchilla; intraocular pressure; manometry

Mesh:

Year:  2017        PMID: 28303681     DOI: 10.1111/vop.12468

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  4 in total

1.  Intraocular pressure measurements in cattle, sheep, and goats with 2 different types of tonometers.

Authors:  Nina Peche; Johanna Corinna Eule
Journal:  Can J Vet Res       Date:  2018-07       Impact factor: 1.310

2.  A Comparison of Applanation and Rebound Tonometers in Young Chicks.

Authors:  Lisa A Ostrin; Christine F Wildsoet
Journal:  Curr Eye Res       Date:  2020-07-03       Impact factor: 2.424

3.  Validation and comparison of four handheld tonometers in normal ex vivo canine eyes.

Authors:  Andrea L Minella; Julie A Kiland; Shawna Gloe; Gillian J McLellan
Journal:  Vet Ophthalmol       Date:  2020-06-01       Impact factor: 1.644

4.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.