| Literature DB >> 2828395 |
Abstract
A rapid, sensitive and selective determination of delta 9-tetrahydrocannabinol (THC) in human plasma, serum and saliva was developed with high-performance liquid chromatography with electrochemical detection. Initially, samples were deproteinized, followed by a one step liquid-liquid extraction. Samples were measured by high-performance liquid chromatography with electrochemical detection with 4-dodecylresorcinol as the internal standard. The minimal detectable limit for THC in biological samples was ca. 1 ng/ml with a signal-to-noise ratio greater than 3, corresponding to an on-column sensitivity for THC of ca. 0.5 ng. The detector was operated at + 0.90 V vs. Ag/AgCl and exhibited linearity over a concentration range of 1-150 ng/ml with correlation coefficients of the standard curves greater than 0.99.Entities:
Mesh:
Substances:
Year: 1987 PMID: 2828395 DOI: 10.1016/0378-4347(87)80382-1
Source DB: PubMed Journal: J Chromatogr