Literature DB >> 28273733

Real-time SPARSE-SENSE cine MR imaging in atrial fibrillation: a feasibility study.

Juliane Goebel1, Felix Nensa1, Haemi P Schemuth1, Stefan Maderwald2, Harald H Quick2,3, Thomas Schlosser1, Kai Nassenstein1.   

Abstract

Background Cardiac magnetic resonance imaging (MRI) relies on correct ECG-gating, which is hindered in arrhythmia. Purpose To examine whether a prototype free-breathing real-time cine sequence using SPARSE-SENSE (SPARSE) improves left ventricular quantification in atrial fibrillation. Material and Methods On a 1.5T MR system left ventricular short-axis stacks were acquired of the SPARSE sequence and of a "reference" steady-state free precession (SSFP) sequence with arrhythmia rejection in 20 patients with atrial fibrillation. Two radiologists independently rated arrhythmia-caused artifact severity in both sequences using a 4-point scale. Coefficients of variation of myocardial signal intensity for both sequences were acquired. Volumetry was performed twice by one reader and once by another reader. Correlation between artifact severity and employed sequence was analyzed by modified Fisher's exact test. Coefficients of variation and volumetric data were compared by paired t-test and intraclass correlation. Results Median arrhythmia-caused artifact severity was 2 in both readers for SSFP and 0 (reader 1)/1 (reader 2) for SPARSE, being significantly lower in SPARSE ( P < 0.001). Mean coefficient of variance was significantly smaller in SPARSE (0.11 ± 0.04) compared to SSFP (0.22 ± 0.13, P = 0.003), which was interpreted as a hint for fewer artifacts in SPARSE. Only a small difference of 9 ± 15 mL was seen for end-systolic volume ( P = 0.019) between sequences, otherwise no significant difference was detected (end-diastolic volume, P = 0.200; stroke volume, P = 0.554; ejection fraction, P = 0.136; myocardial mass, P = 0.353). Intraclass correlation between sequences was good to excellent (range, 0.80-0.97). Conclusion Real-time MRI with SPARSE data sampling is promising in atrial fibrillation because it reduces arrhythmia-caused artifacts.

Entities:  

Keywords:  Real-time imaging; SPARSE sampling; SPARSE-SENSE cine imaging; absolute arrhythmia; left ventricular volumetry

Mesh:

Year:  2016        PMID: 28273733     DOI: 10.1177/0284185116681037

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  5 in total

1.  Validation of highly accelerated real-time cardiac cine MRI with radial k-space sampling and compressed sensing in patients at 1.5T and 3T.

Authors:  Hassan Haji-Valizadeh; Amir A Rahsepar; Jeremy D Collins; Elwin Bassett; Tamara Isakova; Tobias Block; Ganesh Adluru; Edward V R DiBella; Daniel C Lee; James C Carr; Daniel Kim
Journal:  Magn Reson Med       Date:  2017-09-17       Impact factor: 4.668

2.  Compressed Sensing Cardiac Cine Imaging Compared with Standard Balanced Steady-State Free Precession Cine Imaging in a Pediatric Population.

Authors:  Francesco Secchi; Aurelio Secinaro; Davide Curione; Paolo Ciliberti; Caterina Beatrice Monti; Davide Capra; Veronica Bordonaro; Paolo Ciancarella; Teresa Pia Santangelo; Carmela Napolitano; Dolores Ferrara; Marco Alfonso Perrone
Journal:  Radiol Cardiothorac Imaging       Date:  2022-04-21

3.  Assessment of Left Ventricular Systolic Function by Cardiovascular Magnetic Resonance Compressed Sensing Real-Time Cine Imaging Combined With Area-Length Method in Normal Sinus Rhythm and Atrial Fibrillation.

Authors:  Gang Yin; Chen Cui; Jing An; Kankan Zhao; Kai Yang; Shuang Li; Xinling Yang; Jiaxin Wang; Zhixiang Dong; Shiqin Yu; Jian He; Xiuyu Chen; Minjie Lu; Shihua Zhao
Journal:  Front Cardiovasc Med       Date:  2022-05-27

4.  Compressed sensing real-time cine imaging for assessment of ventricular function, volumes and mass in clinical practice.

Authors:  Mathilde Vermersch; Benjamin Longère; Augustin Coisne; Michaela Schmidt; Christoph Forman; Aurélien Monnet; Julien Pagniez; Valentina Silvestri; Arianna Simeone; Emma Cheasty; David Montaigne; François Pontana
Journal:  Eur Radiol       Date:  2019-08-01       Impact factor: 5.315

5.  60-S Retrogated Compressed Sensing 2D Cine of the Heart: Sharper Borders and Accurate Quantification.

Authors:  Benjamin Longère; Christos V Gkizas; Augustin Coisne; Lucas Grenier; Valentina Silvestri; Julien Pagniez; Arianna Simeone; Justin Hennicaux; Michaela Schmidt; Christoph Forman; Solenn Toupin; David Montaigne; François Pontana
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.