Comparing pre-therapeutic 124I and 131I uptake tests with intra-therapeutic 131I uptake in benign thyroid disorders.

Falk Gühne1, Christian Kühnel1, Martin Freesmeyer2.   

Abstract

PURPOSE: 124I-PET/CT can be used for pre-therapeutic assessment of radioactive iodine uptake in benign thyroid disorders, however systematic comparisons with intra-therapeutic uptake are still lacking for these disorders. The goals of this study were to compare 124I RAIU and conventional 131I RAIU tests with each other; to compare both tests with intra-therapeutic uptake (reference); and to verify the time course of radioactive iodine uptake at three time points (30, 102, and 336 h [14 days] post administration; p.a.).
METHODS: Thirteen patients with benign thyroid diseases underwent 131I RAIU test and 124I RAIU test one after another before the intra-therapeutic 131I uptake (reference) was measured via short-range and long-range measurements. After correction for decay, relative uptake differences were calculated and subjected to the Bland-Altman method for the evaluation of levels of agreement.
RESULTS: Radioactive iodine uptake tests with 124I-PET/CT and 131I probe did not show systematic deviations at any time point. Likewise, at 30 and 102 h p.a. there was no systematic discrepancy between pre-therapeutic and intra-therapeutic uptake levels. At 14 days p.a., however, both pre-therapeutic tests tended to overestimate the uptake compared to reference. Findings showed, for the first time with 124I, that radioiodine therapy has some early radiobiological effects possibly limiting the accuracy of pre-therapeutic dosimetry.
CONCLUSIONS: 124I RAIU tests represent a feasible alternative to standard 131I RAIU tests. The additional benefits of 124I-PET/CT (e.g., functional topography, inclusion of retrosternal areas, possibility to enable fusion imaging) may thus increase the scope of this technology in benign thyroid disorders.

Entities:  

Keywords:  124Iodine; Benign thyroid disorders; PET/CT; RAIU; Uptake

Mesh:

Substances:

Year:  2017        PMID: 28271395     DOI: 10.1007/s12020-017-1267-8

Source DB:  PubMed          Journal:  Endocrine        ISSN: 1355-008X            Impact factor:   3.633


  40 in total

Review 1.  Clinical outcomes after estimated versus calculated activity of radioiodine for the treatment of hyperthyroidism: systematic review and meta-analysis.

Authors:  A de Rooij; J P Vandenbroucke; J W A Smit; M P M Stokkel; O M Dekkers
Journal:  Eur J Endocrinol       Date:  2009-08-11       Impact factor: 6.664

Review 2.  Clinical applications of 124I-PET/CT in patients with differentiated thyroid cancer.

Authors:  Lutz S Freudenberg; Walter Jentzen; Alexander Stahl; Andreas Bockisch; Sandra J Rosenbaum-Krumme
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-04-12       Impact factor: 9.236

3.  Procedure guideline for thyroid uptake measurement: 1.0. Society of Nuclear Medicine.

Authors:  D Becker; N D Charkes; H Dworkin; J Hurley; I R McDougall; D Price; H Royal; S Sarkar
Journal:  J Nucl Med       Date:  1996-07       Impact factor: 10.057

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Assessment of Minimum 124I Activity Required in Uptake Measurements Before Radioiodine Therapy for Benign Thyroid Diseases.

Authors:  Anja S Gabler; Christian Kühnel; Thomas Winkens; Martin Freesmeyer
Journal:  J Nucl Med       Date:  2016-03-16       Impact factor: 10.057

6.  Reduced iodide transport (stunning) and DNA synthesis in thyrocytes exposed to low absorbed doses from 131I in vitro.

Authors:  Charlotta Lundh; Madeleine M Nordén; Mikael Nilsson; Eva Forssell-Aronsson
Journal:  J Nucl Med       Date:  2007-03       Impact factor: 10.057

7.  Evaluation of dosimetry of radioiodine therapy in benign and malignant thyroid disorders by means of iodine-124 and PET.

Authors:  Susanne M Eschmann; Gerald Reischl; Kilian Bilger; Juergen Kupferschläger; Marcel H Thelen; Bernhard M Dohmen; Hariolf Besenfelder; Roland Bares
Journal:  Eur J Nucl Med Mol Imaging       Date:  2002-03-27       Impact factor: 9.236

8.  Discrepancies in thyroid uptake values. Use of commercial thyroid probe systems versus scintillation cameras.

Authors:  K H Lee; M E Siegel; O A Fernandez
Journal:  Clin Nucl Med       Date:  1995-03       Impact factor: 7.794

9.  Comparative evaluation of 24-hour thyroid 131I uptake between γ camera-based method using medium-energy collimator and standard uptake probe-based method.

Authors:  Biju K Menon; Ramesh D Rao; Amit Abhyankar; M G Ramakrishna Rajan; Sandip Basu
Journal:  J Nucl Med Technol       Date:  2014-06-26

10.  Statistical and radiobiological analysis of the so-called thyroid stunning.

Authors:  Stephan Walrand; Michel Hesse; François Jamar
Journal:  EJNMMI Res       Date:  2015-11-19       Impact factor: 3.138

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.