Eliezer Be'eri1, Simon Owen1, Mark Shachar2, Yaron Barlavie3, Arik Eisenkraft4,5,6.
Abstract
BACKGROUND: Standard-issue Chemical-Biological-Radio-Nuclear (CBRN) gasmasks, as used for protection from non-conventional warfare agents or toxic industrial compounds, cannot be used by ventilated patients, leaving them exposed to toxic agents inhaled via their ventilators. This study was conducted to determine the safety of a CBRN filter added to the patient circuit of a ventilator, as a method for affording inhalational protection to ventilated patients.Entities:
Keywords: CBRN filter; Protection; Toxic inhalational compounds; Ventilation
Year: 2016 PMID: 28265437 PMCID: PMC5330114 DOI: 10.1186/s40696-016-0015-6
Source DB: PubMed Journal: Disaster Mil Med ISSN: 2054-314X
Fig. 1CBRN filter attached to the air-outflow port of a home ventilator
Ventilation protocols used
| Protocol number | Make of ventilator | Mode | Cycling mechanism | Parameters | Notes |
|---|---|---|---|---|---|
| 1 | Pulmonetic LTV-1000 | Assist-control | Volume | TV = 275, PEEP = 4, rate = 20, Ti = 1 | a |
| 2 | Pulmonetic LTV-1000 | Assist-control | Volume | TV = 320, PEEP = 4, rate = 20, Ti = 1 | a |
| 3 | Pulmonetic LTV-1000 | Assist-control | Volume | TV = 400, PEEP = 4, rate = 20, Ti = 1 | a |
| 4 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 15, PEEP = 5, rate = 20, Ti = 1 | a |
| 5 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 20, PEEP = 5, rate = 20, Ti = 1 | a |
| 6 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 25, PEEP = 5, rate = 20, Ti = 1 | a |
| 7 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 10, PEEP = 5, Ti = 1.4, rate = 20 | a |
| 8 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 15, PEEP = 5, Ti = 1.4, rate = 20 | a |
| 9 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 20, PEEP = 5, Ti = 1.4, rate = 20 | a |
| 10 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 15, PEEP = 4, Ti = 1.4 | a |
| 11 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 15, PEEP = 6, Ti = 1.4 | a |
| 12 | Respironics BiPAP Synchrony | Spontaneous-timed | Flow (pressure support) | PIP = 15, PEEP = 8, Ti = 1.4 | a |
| 13 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 10, PEEP = 5, rate = 20, Ti = 1 | b |
| 14 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 15, PEEP = 5, rate = 20, Ti = 1 | b |
| 15 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 20, PEEP = 5, rate = 20, Ti = 1 | b |
| 16 | Versamed iVent | Assist-control | Time (pressure limited) | PIP = 25, PEEP = 5, rate = 20, Ti = 1 | b |
| 17 | Versamed iVent | Assist-control | Volume | TV = 260, PEEP = 4, rate = 20, Ti = 1 | c |
a CBRN filter mounted on air outflow port of ventilator
b CBRN filter mounted on endotracheal tube
c CBRN filter mounted on air outflow port of ventilator, ventilation for 30 min
Ti Inspiratory time
Fig. 2CBRN Filter positioned between an endotracheal tube and the respiratory circuit of a home ventilator
Change in air-flow parameters caused by addition of CBRN filter to circuit
| Protocol number | Loss of PIP cmH2O (%) | Loss of PEEP cmH2O (%) | Loss of TV ml (%) | Notes |
|---|---|---|---|---|
| 1 | 3 (25) | 0 | a | |
| 2 | 1 (7) | −1 (−20) | a | |
| 3 | 3 (15) | 0 | a | |
| 4 | 0 | 0 | 10 (3) | b |
| 5 | 1 (5) | 0 | 40 (10) | b |
| 6 | 3 (12) | 0 | 50 (11) | b |
| 7 | 1 (10) | 0 | 10 (7) | c |
| 8 | 1 (7) | 0 | 30 (10) | c |
| 9 | 2 (10) | 1 (20) | 20 (5) | c |
| 10 | 1 (8) | 1 (25) | 30 (9) | c |
| 11 | 1 (7) | 0 | 20 (7) | c |
| 12 | 1 (7) | 0 | 10 (5) | c |
| 13 | 1 (8) | 0 | 10 (5) | d |
| 14 | 2 (11) | 0 | 0 | d |
| 15 | 3 (12) | 0 | −10 (−2) | d |
| 16 | 3 (10) | 1 (20) | −10 (−2) | d |
|
|
|
|
a Volume cycled ventilation with CBRN filter on air outflow port of ventilator
b Time cycled pressure limited ventilation with CBRN filter on air outflow port of ventilator
c Flow cycled ventilation with CBRN filter on air outflow port of ventilator
d Time cycled pressure limited ventilation with CBRN filter on endotracheal tube
Fig. 3Resistance vs flow of a standard-issue CBRN filter
Trigger gradient across the CBRN filter (cm H2O)
| Pressure generated proximal to filter | Pressure measured distal to filter | Pressure gradient across filter |
|---|---|---|
| −0.72 | −0.50 | 0.22 |
| −0.78 | −0.43 | 0.35 |
| −0.86 | −0.63 | 0.23 |
| −1.27 | −0.78 | 0.49 |
| −1.30 | −1.11 | 0.19 |
| −1.53 | −1.32 | 0.21 |
| −1.65 | −1.21 | 0.44 |
| −1.90 | −1.36 | 0.54 |
| −2.37 | −1.66 | 0.71 |
| −2.54 | −1.57 | 0.97 |
| −3.07 | −1.68 | 1.39 |
| −4.04 | −2.65 | 1.39 |
|
|
|
|