Retrieval Analysis of Neck-Stem Coupling in Modular Hip Prostheses.

Sherwin L Su1, Chelsea N Koch2, Thu M Nguyen2, Jayme C Burket3, Timothy M Wright2, Geoffrey H Westrich1.   

Abstract

BACKGROUND: Dual-taper modular stems have suffered from high revision rates caused by adverse local tissue reactions secondary to fretting and corrosion. We compared the fretting and corrosion behavior of a group of modular neck designs to that of a design that had been recalled for risks associated with fretting and corrosion at the modular neck junction.
METHODS: We previously analyzed fretting and corrosion on 60 retrieved Rejuvenate modular neck-stem implants. Here we compare those results to results from 26 retrieved implants from 7 other modular neck designs. For the 26 additional cases, histology slides of tissue collected at revision were reviewed and graded for aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Multivariate analyses were performed to assess differences in fretting and corrosion, adjusting for confounding factors (eg, length of implantation).
RESULTS: The Rejuvenate design had higher damage and corrosion scores than the other 7 designs (P < .01). Histologic samples from the recalled design were 20 times more likely to show ALVAL than samples from the other designs (P < .01). Mixed metal couples had higher fretting (P < .01) and corrosion (P = .02) scores than non-mixed metal couples.
CONCLUSION: Fretting and corrosion occurred on all modular neck-stem retrievals regardless of design. However, mixed metal couples suffered more corrosion than homogenous couples. This may be due to the lower modulus of the titanium alloy used for the stem, allowing for increased metal transfer and surface damage when loaded against a cobalt alloy modular neck, which in turn could account for the higher ALVAL and corrosion scores. Due to increased corrosion risk with mixed metals and increased neck fracture risk with non-mixed metal stem and necks, we suggest that clinicians avoid implantation of modular neck-stem systems.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse local tissue reaction; corrosion; fretting; modular neck; total hip arthroplasty

Mesh:

Substances:

Year:  2017        PMID: 28262451     DOI: 10.1016/j.arth.2017.02.016

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  MRI of THA Correlates With Implant Wear and Tissue Reactions: A Cross-sectional Study.

Authors:  Matthew F Koff; Christina Esposito; Parina Shah; Mauro Miranda; Elexis Baral; Kara Fields; Thomas Bauer; Douglas E Padgett; Timothy Wright; Hollis G Potter
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

2.  Mid-term outcomes of titanium modular neck femoral stems in revision total hip arthroplasty.

Authors:  Hervé Ouanezar; Thomas Jalaguier; Florent Franck; Vincent Pibarot; Hugo Bothorel; Mo Saffarini; Jean-Pierre Piton
Journal:  Ann Transl Med       Date:  2019-03

3.  Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years' follow-up.

Authors:  Sverrir Kiernan; Bart Kaptein; Carl Flivik; Martin Sundberg; Gunnar Flivik
Journal:  Acta Orthop       Date:  2020-12-10       Impact factor: 3.717

4.  Midterm outcomes of titanium modular femoral necks in total hip arthroplasty.

Authors:  Thurial Collet; Jean-Pierre Atanasiu; Jean-Baptiste de Cussac; Kamal Oufroukhi; Hugo Bothorel; Mo Saffarini; François Badatcheff
Journal:  Ann Transl Med       Date:  2017-10
  4 in total