Literature DB >> 28249868

Cortrak-Assisted Feeding Tube Insertion: A Comprehensive Review of Adverse Events in the MAUDE Database.

Annette M Bourgault1, Lillian Aguirre2, Joseph Ibrahim2.   

Abstract

BACKGROUND: Electromagnetic devices to guide feeding tube placement such as the CORTRAK Enteral Access System have shown promising results; however, researchers in recent studies have expressed concern that a higher level of user expertise may be required for safe use.
OBJECTIVES: To review adverse events related to CORTRAK-assisted feeding tube insertion reported in the Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database.
METHODS: A retrospective, secondary analysis of the MAUDE database was performed to evaluate adverse events (ie, injury or death of patient) related to CORTRAK.
RESULTS: Fifty-four adverse events between January 1, 2006 and February 29, 2016 were identified and reviewed. Most events (98%) involved feeding tube placement in the lungs (37%, left lung; 46%, right lung; 15%, not specified). Lung complications included pneumothorax (77%) and pneumonitis (21%). Death occurred in 17% of lung placements. Clinicians failed to recognize placement in 89% of CORTRAK insertion tracings reviewed.
CONCLUSIONS: Lung placement is not unique to CORTRAK and is an inherent risk of all feeding tube insertions. In known or suspected lung placement, feeding tubes should be removed and radiography performed to assess for pneumothorax. Clinicians must observe closely for lung placement and discriminate lung from gastric placement on insertion tracings. Clinicians require specialized training and experience to develop competency in using the CORTRAK device, although the exact amount of experience needed is unknown. ©2017 American Association of Critical-Care Nurses.

Entities:  

Mesh:

Year:  2017        PMID: 28249868     DOI: 10.4037/ajcc2017369

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  3 in total

1.  Varying levels of fidelity on psychomotor skill attainment: a CORTRAK product assessment.

Authors:  Laura Gonzalez; Annette M Bourgault; Lillian Aguirre
Journal:  BMJ Simul Technol Enhanc Learn       Date:  2018-07-09

2.  Tracheopulmonary Complications of a Malpositioned Nasogastric Tube.

Authors:  David B Guthrie; James P Pezzollo; David K Lam; Ralph H Epstein
Journal:  Anesth Prog       Date:  2020-09-01

3.  Feasibility and safety of a novel electromagnetic device for small-bore feeding tube placement.

Authors:  Lewis E Jacobson; May Olayan; Jamie M Williams; Jacqueline F Schultz; Hannah M Wise; Amandeep Singh; Jonathan M Saxe; Richard Benjamin; Marie Emery; Hilary Vilem; Donald F Kirby
Journal:  Trauma Surg Acute Care Open       Date:  2019-11-13
  3 in total