Repeatability of Central Corneal Thickness Measurement Using Rotating Scheimpflug Camera in Dry and Normal Eyes.

Jong-Hyuck Lee1, Jae Hyuck Kim, Sun Woong Kim.   

Abstract

PURPOSE: To compare the repeatability of central corneal thickness (CCT) measurement using the Pentacam between dry eyes and healthy eyes, as well as to investigate the effect of artificial tears on CCT measurement.
METHODS: The corneal thicknesses of 34 patients with dry eye and 28 healthy subjects were measured using the Pentacam. One eye from each subject was assigned randomly to a repeatability test, wherein a single operator performed three successive CCT measurements time points-before and 5 min after instillation of one artificial teardrop. The repeatability of measurements was assessed using the coefficient of repeatability and the intraclass correlation coefficient.
RESULTS: The coefficient of repeatability values of the CCT measurements in dry and healthy eyes were 24.36 and 10.69 μm before instillation, and 16.85 and 9.72 μm after instillation, respectively. The intraclass correlation coefficient was higher in healthy eyes than that of in dry eyes (0.987 vs. 0.891), and it had improved significantly in dry eyes (0.948) after instillation of one artificial teardrop. The CCT measurement fluctuated in dry eyes (repeated-measures analysis of variance, P<0.001), whereas no significant changes were detected in healthy eyes, either before or after artificial tear instillation.
CONCLUSIONS: Central corneal thickness measurement is less repeatable in dry eyes than in healthy eyes. Artificial tears improve the repeatability of CCT measurements obtained using the Pentacam in dry eyes.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 28244934     DOI: 10.1097/ICL.0000000000000373

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  4 in total

1.  Monitoring of central corneal thickness after phacoemulsification-comparison of statical and rotating Scheimpflug pachymetry, and spectral-domain OCT.

Authors:  Daniel M Handzel; Carsten H Meyer; Alfred Wegener
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

2.  Reference Values of the Central Corneal Thickness with Different Refractive Errors for the Adult Egyptian Population.

Authors:  Amr A Gab-Alla
Journal:  Clin Ophthalmol       Date:  2020-10-22

3.  The repeatability of corneal topography measurements in severe Dry Eye disease.

Authors:  Soner Guven
Journal:  BMC Ophthalmol       Date:  2022-07-15       Impact factor: 2.086

4.  Does Dry Eye Affect Repeatability of Corneal Topography Measurements?

Authors:  Aysun Şanal Doğan; Canan Gürdal; Mehmet Talay Köylü
Journal:  Turk J Ophthalmol       Date:  2018-04-25
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.