Real-Time Verification of a High-Dose-Rate Iridium 192 Source Position Using a Modified C-Arm Fluoroscope.

Takayuki Nose1, Masashi Chatani2, Yuki Otani3, Teruki Teshima4, Shinichirou Kumita5.   

Abstract

PURPOSE: High-dose-rate (HDR) brachytherapy misdeliveries can occur at any institution, and they can cause disastrous results. Even a patient's death has been reported. Misdeliveries could be avoided with real-time verification methods. In 1996, we developed a modified C-arm fluoroscopic verification of an HDR Iridium 192 source position prevent these misdeliveries. This method provided excellent image quality sufficient to detect errors, and it has been in clinical use at our institutions for 20 years. The purpose of the current study is to introduce the mechanisms and validity of our straightforward C-arm fluoroscopic verification method. METHODS AND MATERIALS: Conventional X-ray fluoroscopic images are degraded by spurious signals and quantum noise from Iridium 192 photons, which make source verification impractical. To improve image quality, we quadrupled the C-arm fluoroscopic X-ray dose per pulse. The pulse rate was reduced by a factor of 4 to keep the average exposure compliant with Japanese medical regulations. The images were then displayed with quarter-frame rates.
RESULTS: Sufficient quality was obtained to enable observation of the source position relative to both the applicators and the anatomy. With this method, 2 errors were detected among 2031 treatment sessions for 370 patients within a 6-year period.
CONCLUSIONS: With the use of a modified C-arm fluoroscopic verification method, treatment errors that were otherwise overlooked were detected in real time. This method should be given consideration for widespread use.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 28244423     DOI: 10.1016/j.ijrobp.2016.12.005

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Radiation protection in radiological imaging: a survey of imaging modalities used in Japanese institutions for verifying applicator placements in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Kito; Naoki Tohyama; Shunsuke Yonai; Ryu Kawamorita; Masaru Nakamura; Takahiro Fujimoto; Syoji Tani; Akihiro Yomoda; Toru Isobe; Hiroshi Furukawa; Kikuo Kotaka; Jun Itami; Hitoshi Ikushima; Takushi Dokiya; Yoshiyuki Shioyama
Journal:  J Radiat Res       Date:  2021-01-01       Impact factor: 2.724

2.  High-dose rate intracavitary brachytherapy pretreatment dwell position verification using a transparent applicator.

Authors:  Yuki Otani; Iori Sumida; Takayuki Nose; Shigetoshi Shimamoto; Hirofumi Okubo; Kazuhiko Ogawa
Journal:  J Appl Clin Med Phys       Date:  2018-06-30       Impact factor: 2.102

3.  Independent assessment of source position for gynecological applicator in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Nakamura; Shie Nishioka; Kotaro Iijima; Akihisa Wakita; Yukinao Abe; Naoki Tohyama; Shinji Kawamura; Toshiyuki Minemura; Jun Itami
Journal:  J Contemp Brachytherapy       Date:  2017-10-19

4.  First experience of 192Ir source stuck event during high-dose-rate brachytherapy in Japan.

Authors:  Shinobu Kumagai; Norikazu Arai; Takeshi Takata; Daisuke Kon; Toshiya Saitoh; Hiroshi Oba; Shigeru Furui; Jun'ichi Kotoku; Kenshiro Shiraishi
Journal:  J Contemp Brachytherapy       Date:  2020-02-28
  4 in total