| Literature DB >> 28243387 |
Abstract
PURPOSE: The aim of this study was to compare the retention of mini implant overdenture by the number, the type of magnetic attachment, and the directions of applied dislodging force.Entities:
Keywords: Dental mini implants; Magnets; Overdenture; Precision attachment
Year: 2017 PMID: 28243387 PMCID: PMC5321584 DOI: 10.4047/jap.2017.9.1.14
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1Two experimental models with two and four implants placed in the interforamen area of mandible. (A) group 1, (B) group 2.
Fig. 2Magnetic attachment overdenture sample seated on a mandibular model with nuts and bolts securely fastened.
Means and standard deviations of retentive forces of experimental groups (unit: N)
| Group 1 | Group 2 | |||||||
|---|---|---|---|---|---|---|---|---|
| Flat type | Cushion type | Flat type | Cushion type | |||||
| Mean | SD | Mean | SD | Mean | SD | Mean | SD | |
| Vertical | 1.06 | 0.21 | 1.01 | 0.13 | 5.10 | 1.01 | 3.99 | 0.43 |
| Oblique | 1.65 | 0.21 | 1.37 | 0.18 | 4.66 | 0.46 | 3.71 | 0.43 |
| Ant-Post | 0.81 | 0.12 | 0.75 | 0.12 | 1.77 | 0.42 | 1.52 | 0.51 |
Comparison of retentive force according to the number of implant by Mann-Whitney U test
| Flat type | Cushion type | |
|---|---|---|
| Group 1 – Group 2 | Group 1 – Group 2 | |
| Vertical | .008* | .008* |
| Oblique | .008* | .008* |
| Anterior-Posterior | .008* | .008* |
*denotes the significance at the 0.05 level
Fig. 3Comparison of retentive force according to the number of implant.
Comparison of retentive force according to the type of magnetic attachment by Mann-Whitney U test
| Flat type | Cushion type | |
|---|---|---|
| Group 1 – Group 2 | Group 1 – Group 2 | |
| Vertical | .548* | .056* |
| Oblique | .008* | .008* |
| Anterior-Posterior | .841* | .421* |
*denotes the significance at the 0.05 level
Fig. 4Comparison of retentive force according to the type of magnetic attachment.
Comparison of retentive force according to vertical (V), oblique (O), anterior-posterior (AP) directions by Wilcoxon signed-rank test
| Group 1 | Group 2 | |||||
|---|---|---|---|---|---|---|
| V-O | O-AP | AP-V | V-O | O-AP | AP-V | |
| Flat type | .043* | .043* | .043* | .686 | .043* | .043* |
| Cushion type | .043* | .043* | .042* | .345 | .043* | .043* |
*denotes the significance at the 0.05 level
Fig. 5Comparison of retentive force according to vertical (V), oblique (O), anterior-posterior (AP) directions.