| Literature DB >> 28241847 |
Frank A Rasulo1,2, Rita Bertuetti3, Chiara Robba4, Francesco Lusenti5, Alfredo Cantoni6, Marta Bernini7, Alan Girardini8, Stefano Calza9, Simone Piva3, Nazzareno Fagoni3, Nicola Latronico3,10.
Abstract
BACKGROUND: Untimely diagnosis of intracranial hypertension may lead to delays in therapy and worsening of outcome. Transcranial Doppler (TCD) detects variations in cerebral blood flow velocity which may correlate with intracranial pressure (ICP). We investigated if intracranial hypertension can be accurately excluded through use of TCD.Entities:
Keywords: Brain injury; Intracranial hypertension; Intracranial pressure; Transcranial Doppler
Mesh:
Year: 2017 PMID: 28241847 PMCID: PMC5329967 DOI: 10.1186/s13054-017-1632-2
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
General characteristics and admission diagnoses
| Type of brain injury | Number | Age (years) | EVD | IP bolt-screw | ICHT |
|---|---|---|---|---|---|
| TBI | 20 | 45.1 (12.8) | 4/20 | 16/20 | 8/20 |
| aSAH | 11 | 61.4 (9) | 6/11 | 5/11 | 2/11 |
| ICH | 7 | 72.6 (5.7) | 0/7 | 7/7 | 2/7 |
| TOT cohort | 38 | 57.8 (15.7) | 10/38 | 28/38 | 12/38 |
aSAH aneurysmal subarachnoid hemorrhage, EVD external ventricular drain, ICH intracerebral hemorrhage, ICHT intracranial hypertension, TBI traumatic brain injury
Invasive (ICPi) and transcranial Doppler (ICPtcd) intracranial pressure (ICP) measurements at study times
| ICP values | Number of ICP measurements | |||
|---|---|---|---|---|
| ICPtcd (mmHg) | ICPi (mmHg) | ICPtcd >20 mmHg | ICPi >20 mmHg | |
| TIME 1 | 20.5 (9.1) | 13.5 (8.0) | 18 (47.3%) | 10 (26.3%) |
| TIME 2 | 16.7 (7.7) | 11.1 (7.8) | 10 (26.3%) | 4 (10.5%) |
| TIME 3 | 17.5 (8.7) | 11.5 (8.0) | 13 (34.2%) | 6 (15.8%) |
| Overall | 18.2 (8.6) | 12.0 (7.9) | 21 (55.3%) | 12 (31.6%) |
TIME1 ICPtcd immediately before ICPi insertion, TIME 2 ICPtcd immediately after ICPi insertion, TIME 3 from 2 to 3 h following ICPi insertion
Fig. 1Dichotomized ICP readings, ≤20 mmHg vs >20 mmHg, ≤24.8 mmHg vs >24.8 mmHg. ICPi invasive intracranial pressure, ICPtcd transcranial Doppler estimation of intracranial pressure
Fig. 2Estimated conversion between ICPtcd and ICPi; slope value = 1.02 (95% CI 0.85–1.36). ICPi invasive intracranial pressure, ICPtcd transcranial Doppler estimation of intracranial pressure
Fig. 3Bland-Altman plot showing mean bias of + 6.2 mmHg for ICPtcd compared to ICPi, with an agreement range from -5.6 to 18 mmHg. ICPi invasive intracranial pressure, ICPtcd transcranial Doppler estimation of intracranial pressure
Fig. 4ROC curve analysis for ICPtcd averaged over times (TIME 1, TIME 2, and TIME 3) showing an AUC of 96.0% (95% CI 89.8–100%). The estimated best threshold value was at an ICPi of 24.8 mmHg corresponding to a sensitivity of 100% and a specificity of 91.2%