An external dosimetry audit programme to credential static and rotational IMRT delivery for clinical trials quality assurance.

David J Eaton1, Justine Tyler2, Alex Backshall2, David Bernstein2, Antony Carver3, Anne Gasnier2, Julia Henderson2, Jonathan Lee4, Rushil Patel4, Yatman Tsang4, Huiqi Yang4, Rada Zotova4, Emma Wells2.   

Abstract

PURPOSE: External dosimetry audits give confidence in the safe and accurate delivery of radiotherapy. The RTTQA group have performed an on-site audit programme for trial recruiting centres, who have recently implemented static or rotational IMRT, and those with major changes to planning or delivery systems.
METHODS: Measurements of reference beam output were performed by the host centre, and by the auditor using independent equipment. Verification of clinical plans was performed using the ArcCheck helical diode array.
RESULTS: A total of 54 measurement sessions were performed between May 2014 and June 2016 at 28 UK institutions, reflecting the different combinations of planning and delivery systems used at each institution. Average ratio of measured output between auditor and host was 1.002±0.006. Average point dose agreement for clinical plans was -0.3±1.8%. Average (and 95% lower confidence intervals) of gamma pass rates at 2%/2mm, 3%/2mm and 3%/3mm respectively were: 92% (80%), 96% (90%) and 98% (94%). Moderately significant differences were seen between fixed gantry angle and rotational IMRT, and between combination of planning systems and linac manufacturer, but not between anatomical treatment site or beam energy.
CONCLUSION: An external audit programme has been implemented for universal and efficient credentialing of IMRT treatments in clinical trials. Good agreement was found between measured and expected doses, with few outliers, leading to a simple table of optimal and mandatory tolerances for approval of dosimetry audit results. Feedback was given to some centres leading to improved clinical practice.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Dosimetry audit; IMRT; Quality assurance; Radiotherapy trials

Mesh:

Year:  2017        PMID: 28236559     DOI: 10.1016/j.ejmp.2017.02.012

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  An overview of the medical-physics-related verification system for radiotherapy multicenter clinical trials by the Medical Physics Working Group in the Japan Clinical Oncology Group-Radiation Therapy Study Group.

Authors:  Teiji Nishio; Mitsuhiro Nakamura; Hiroyuki Okamoto; Satoshi Kito; Toshiyuki Minemura; Shuichi Ozawa; Yu Kumazaki; Masayori Ishikawa; Naoki Tohyama; Masahiko Kurooka; Takeo Nakashima; Hidetoshi Shimizu; Ryusuke Suzuki; Satoshi Ishikura; Yasumasa Nishimura
Journal:  J Radiat Res       Date:  2020-11-16       Impact factor: 2.724

2.  Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand.

Authors:  Lalida Tuntipumiamorn; Puangpen Tangboonduangjit; Taweap Sanghangthum; Rattapol Rangseevijitprapa; Chirasak Khamfongkhruea; Thida Niyomthai; Boonrut Vuttiprasertpong; Supranee Supanant; Nongnuch Chatchaipaiboon; Porntip Iampongpaiboon; Pitchayut Nakkrasae; Tanwiwat Jaikuna
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28

3.  Driving developments in UK oesophageal radiotherapy through the SCOPE trials.

Authors:  S Gwynne; E Higgins; A Poon King; G Radhakrishna; L Wills; S Mukherjee; Maria Hawkins; G Jones; J Staffurth; T Crosby
Journal:  Radiat Oncol       Date:  2019-02-04       Impact factor: 3.481

4.  Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans.

Authors:  Enrica Seravalli; Antonetta C Houweling; Leo Van Battum; Thom A Raaben; Marc Kuik; Jacco A de Pooter; Marion P R Van Gellekom; Jochem Kaas; Wilfred de Vries; Erik A Loeff; Jeroen B Van de Kamer
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-02

5.  A remote EPID-based dosimetric TPS-planned audit of centers for clinical trials: outcomes and analysis of contributing factors.

Authors:  Narges Miri; Kimberley Legge; Kim Colyvas; Joerg Lehmann; Philip Vial; Alisha Moore; Monica Harris; Peter B Greer
Journal:  Radiat Oncol       Date:  2018-09-17       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.