Simultaneous detection of severe acute respiratory syndrome, Middle East respiratory syndrome, and related bat coronaviruses by real-time reverse transcription PCR.

Ji Yeong Noh1,2, Sun-Woo Yoon1,3, Doo-Jin Kim1, Moo-Seung Lee1,3, Ji-Hyung Kim1, Woonsung Na4, Daesub Song4, Dae Gwin Jeong5,6, Hye Kwon Kim7.   

Abstract

Since severe acute respiratory syndrome (SARS) and Middle East respiratory syndrome (MERS) coronaviruses (CoVs) share similar characteristics with respect to clinical signs, etiology, and transmission, methods for a rapid and accurate differential diagnosis are important. Therefore, the aim of this study was to develop a duplex real-time reverse transcription (RT)-PCR method for the simultaneous detection of these viruses. Primers and probes that target the conserved spike S2 region of human SARS-CoV, MERS-CoV, and their related bat CoVs were designed. The results of real-time RT-PCR showed specific reactions for each virus with adequate detection limits of 50-100 copies/mL and 5-100 copies/mL using pUC57-SARS-pS2 (a template for SARS-CoV) and pGEM-MERS-S2 (a template for MERS-CoV), respectively. In addition, this real-time RT-PCR system was able to detect the target viruses SARS-like bat CoV and MERS-CoV in bat fecal samples and sputum of MERS patients, respectively. Therefore, this newly developed real-time RT-PCR method is expected to detect not only SARS-CoV and MERS-CoV in humans but also several bat CoVs that are closely related to these viruses in bats.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28220326      PMCID: PMC7086956          DOI: 10.1007/s00705-017-3281-9

Source DB:  PubMed          Journal:  Arch Virol        ISSN: 0304-8608            Impact factor:   2.574


Introduction

The recent emergence of human coronaviruses (CoVs) causing severe acute respiratory syndrome (SARS) and Middle East respiratory syndrome (MERS) with global spread represents a significant threat to public health. Both SARS-CoV and MERS-CoV cause severe respiratory diseases and belong to the genus Betacoronavirus [1]. Phylogenetic analysis has shown that SARS-CoV belongs to lineage B, which includes SARS-like bat CoVs and some other bat-derived CoVs, whereas MERS-CoV belongs to lineage C, which also includes bat-derived CoVs [1]. SARS-CoV and MERS-CoV share similar transmission characteristics. Both viruses are thought to have originated from bats, which are the primary reservoir of diverse CoVs [2]. Cross-species transmission of these viruses to palm civets and dromedary camels has increased the chance of their zoonotic transmission to humans [3]. Nosocomial transmission is thought to be the main cause of human-to-human transmission of SARS-CoV and MERS-CoV [4]. Although there have been no SARS cases reported since 2004, MERS cases have been reported continuously worldwide [3]. In addition, a recent analysis of the SARS-like bat CoVs that use the ACE2 receptor highlighted the need for preparedness for their potential zoonotic transmission [5, 6]. Bat CoVs related to SARS-CoV and MERS-CoV have been reported continuously worldwide [7-10]. Given the high similarity between SARS-CoV and MERS-CoV with respect to clinical signs, etiology, and transmission, establishment of a simultaneous detection method for these viruses would be useful for their accurate diagnosis and monitoring in humans as well as their reservoir, bats. Therefore, in this study, a duplex real-time reverse transcription (RT)-PCR method was developed based on primers and probes that target the conserved spike S2 region of SARS-CoV, SARS-like bat CoVs, MERS-CoV, and MERS-related bat CoVs.

Materials and methods

Primers and probes

For the universal detection of SARS-CoV and SARS-like bat CoVs, consensus primers and probes (Fig. 1a) were designed based on the conserved sequences of the spike S2 region by aligning the following reference sequences: human SARS-CoVs Sino1 (GenBank no. AY485277), Tor2 (AY274119), and Urbani (AY278741); SARS-like bat CoVs B15-21 (KU528591), RP3 (DQ071615), RsSHC014 (KC881005), Rf1 (DQ412042), HKU3 (DQ084199), and 273 (DQ648856).
Fig. 1

Multiple sequence alignment of the spike S2 region of SARS-CoV and SARS-like bat CoVs (a) and MERS-CoV and MERS-like bat CoVs (b). Primer-binding sites are indicated by unshaded boxes, and probe-binding sites are indicated by shaded boxes

Multiple sequence alignment of the spike S2 region of SARS-CoV and SARS-like bat CoVs (a) and MERS-CoV and MERS-like bat CoVs (b). Primer-binding sites are indicated by unshaded boxes, and probe-binding sites are indicated by shaded boxes For the detection of MERS-CoV and possible related bat CoVs, consensus primers and probes (Fig. 1b) were designed based on the conserved sequences of the spike S2 region by aligning the following reference sequences: human MERS-CoVs THA-CU (KT225476), Jeddah (KF958702), and KOR-NIH (KT029139); MERS-related bat CoVs A434 (DQ648790), PML-PHE1 (KC869678), GX2012 (KJ473822), HKU5 (EF065512), SC2013 (KJ473821), and HKU4 (EF065508). The designed primers and probes are shown in Table 1.
Table 1

Primers and probes

TargetOligo nameSequenceAmplicon size
Spike S2 region of SARS-CoVs and SARS-like bat CoVsSCOV-F5′-TCT GCA ACC TGA GCT TGA CT-3′105 bp
SCOV-R5′-ACA GAA GCR TTA ATG CCT GA-3′
SCOV-Probe5′ 6-FAM-CAC CWG ATG TTG ATC TTG GCG ACA-3′ BHQ-1
Spike S2 region of MERS-CoVs and MERS-related bat CoVsMCOV-F5′-CAG ACA ACC ATT CAG AAR GTT A-3′108 bp
MCOV-R5′-TTT AGA ACA AAA CTG GCC ATA-3′
MCOV-Probe5′ 6-HEX-GTT GAT TGT AAA CAG TAC GTT TGC AAT GG-3′ BHQ-1
Primers and probes

Preparation of standard plasmids

The partial sequences of the S2 region (nucleotides 3221–3620) of the spike gene of the SARS-CoV Sino1 strain (GenBank no. AY485277), including the primer- and probe-binding regions, were synthesized and inserted in the pUC57 vector at CosmoGenetech Co., Ltd. (Seoul, Korea). The nucleotide sequence of the MERS-CoV S2 domain (amino acids 765–1288, EMC strain, GenBank no. JX869059) was cloned in the pGEM vector by Bioneer (Daejeon, Korea). Competent Escherichia coli cells (DH5α) were transformed with the recombinant plasmid vectors (pUC57-SARS-pS2 and pGEM-MERS-S2), and the amplified plasmid was extracted. These plasmids were then used as templates for the positive control as well as in the detection limit test after calculating their copy numbers.

In vitro transcription

A portion of the S2 region of pUC57-SARS-pS2 was amplified using primers containing a T7 promoter sequence: SARS-Tem-F (5’-tcg gta ccT AAT ACG ACT CAC TAT AGG Gaa gaG CCA CCA TGA AGG TGT TTT TGT GTT TAA TGG-3’) and SARS-Tem-R (5’-TTT TTT TTT TTT TTT TTT TTT TTT TTT TTT tta ATT AGT CCA GCA ATG AAG CC-3’). The amplified PCR product was used as template for in vitro transcription using an mMESSAGE mMACHINE T7 kit (Ambion, USA) following the manufacturer’s manual. The in vitro-transcribed mRNA was treated with Turbo DNase (2 U/µL) and purified with lithium chloride precipitation solution (7.5 M lithium chloride, 50 mM EDTA). The final concentration of SARS-pS2-based mRNA was measured using a NanoDrop 2000 spectrophotometer (Thermo Scientific, USA).

Real-time RT-PCR

A SensiFAST Probe No-ROX One-Step Kit (BIOLINE, Taunton, MA, USA) was used for the real-time RT-PCRs. Each 20-μL reaction consisted of 4 μL of template, 0.8 μL of each primer (SCOV-F/R and MCOV-F/R, 10 μM each), 10 μL of 2X SensiFAST Probe No-ROX One-Step Mix, 0.2 μL of each probe (SCOV-probe and MCOV-probe, 10 μM each), 0.2 μL of reverse transcriptase, 0.4 μL of RiboSafe RNase Inhibitor, and 1.8 μL of distilled water. The real-time RT-PCR was performed with reverse transcription at 45 °C for 10 min followed by 95 °C for 2 min and cycling 35 times at 95 °C for 10 s and 60 °C for 20 s. Thermocycling was performed using a LightCycler 96 System (Roche, USA), and positive results were estimated according to analysis of the fluorescent curves originating from each probe within 40 cycles.

Estimation of the detection limits of the real-time RT-PCR

The purified recombinant plasmids were prepared at a concentration of 1 × 108 copies/mL for the templates of SARS-CoV and MERS-CoV. Each template was diluted tenfold in nuclease-free distilled water (Ambion, USA). The diluted plasmids were then used as templates for the real-time RT-PCR to evaluate the detection limits of the primers and probes. The in vitro-transcribed mRNA of SARS-CoV pS2 and RNA extracted from MERS-CoV isolate (KOR/KNIH/002_05_2015) were prepared at a concentration of 1 × 104 pg/µL. They were then diluted tenfold with nuclease-free distilled water and used as template for the real-time RT-PCR to evaluate the detection limits of the primers and probes.

Specificity test

To evaluate the specificity of the primers and probes, several RNA viruses, including dengue virus types 2, 3, and 4 (KBPV-VR-29, -30, and -31, respectively; Korea Bank for Pathogenic Viruses, Seoul, Korea), porcine reproductive and respiratory syndrome virus (PRRSV) strain CP401-9, porcine epidemic diarrhea virus (PEDV) strain DR13 (Green Cross Veterinary Products, Yongin, Korea), and influenza A virus (H1N1) strain sk14 from our laboratory, were tested using the newly developed real-time RT-PCR protocol. In addition, RNA extracted from a MERS-CoV isolate (KOR/KNIH/002_05_2015), which was kindly provided by the Korea Center for Disease Control, was tested using the real-time RT-PCR system. A recombinant plasmid including partial sequences of the S2 region (nucleotides 2231–2630 of the spike gene, MERS-like bat CoV strain HKU4 [GenBank no. EF065508]) was synthesized (pUC57-MERS-like-CoV-pS2) and used as a template for the real-time RT-PCR. Agarose gel electrophoresis was performed to observe the target bands as well as for fluorescent detection.

Bat samples

RNA was extracted from 47 bat samples (feces, urine, and oral swabs) collected between July 2015 and April 2016 in Korea. Among them, samples B15-8, B15-40, B15-41, B16-6, and B16-40, and B15-21 were determined to be positive for CoVs according to consensus primers-based RT-PCR, which can detect diverse CoVs by targeting conserved sequences of the RNA-dependent RNA polymerase gene [11]. The positive samples were found to have RNA sequences that are closely related to those of alphacoronaviruses or SARS-like bat CoV [7]. Real-time PCR was performed with the extracted RNAs and compared to the previous RT-PCR results.

Human samples

MERS-CoV-positive RNA samples extracted from the sputum of MERS-CoV-infected patients were kindly provided by the Department of Laboratory Medicine of Seoul National University Hospital, Korea. According to the regulations of Category B UN 3373, the samples were packaged with triple-packaging shells following the P620 UN packaging guideline and transported to the animal biosafety-level 3 facility of the National Primate Research Center of the Korea Research Institute of Bioscience and Biotechnology by the Green Cross Labcell company, a specialist in the transport of clinical samples. The single (MERS-CoV-specific primers and probe only) and duplex real-time RT-PCR were performed with these RNAs and compared with the results of UpE-based real-time RT-PCR as a reference [12].

Results

Detection limits of the real-time RT-PCR

The tenfold-diluted plasmids (pUC57-SARS-pS2 and pGEM-MERS-S2) were tested by real-time RT-PCR as a single template or as mixed templates. The intensity of the fluorescence (FAM and HEX) emitted from the hydrolyzed probes was measured every cycle. Based on the measured amplification curve, the results were expressed as the quantification cycle value (Cq value). The Cq values obtained from the diluted plasmids were found to be significantly correlated with the copy numbers of each plasmid, with an R2 value of 0.9847 and 0.9954 for pUC57-SARS-pS2 and pGEM-MERS-S2, respectively (Fig. 2). The negative control (distilled water) did not yield an amplification curve. As shown in Table 2, the detection limits of the real-time RT-PCR for pUC57-SARS-pS2 and pGEM-MERS-S2 were both 1 × 102 copies/mL in the single-template condition and were 5 × 101 and 5 × 100 copies/mL, respectively, in the mixed-template condition. The detection limits of the real-time RT-PCR for the in vitro-transcribed mRNA of pUC57-SARS-pS2 and viral RNA of MERS CoV (KOR/KNIH/002_05_2015) were 1 × 10−3 and 1 × 100 ng/mL, respectively, in the single-template condition and were 5 × 10−3 and 5 × 101 ng/mL, respectively, in the mixed-template condition (Table 2).
Fig. 2

Correlation curve for plasmids copy number vs. and Cq value. (a) pUC57-SARS-pS2, (b) pGEM-MERS-S2

Table 2

Detection limits of real-time RT-PCR

ConditionPlasmid concentration (copies/mL)RNA concentration (ng/mL)
pUC57-SARS-pS2pGEM-MERS-S2SARS-pS2a MERS-CoVb
Single template1 × 102 1 × 102 1 × 10−3 1 × 100
Mixed template5 × 101 5 × 100 5 × 10−3 5 × 101

amRNA obtained by in vitro transcription

bRNA from MERS-CoV isolate KOR/KNIH/002_05_2015

Correlation curve for plasmids copy number vs. and Cq value. (a) pUC57-SARS-pS2, (b) pGEM-MERS-S2 Detection limits of real-time RT-PCR amRNA obtained by in vitro transcription bRNA from MERS-CoV isolate KOR/KNIH/002_05_2015

Specificity of the real-time RT-PCR

The specificity of the real-time RT-PCR method developed in this study was evaluated using RNAs from several RNA viruses, including MERS-CoV (KOR/KNIH/002_05_2015), a recombinant plasmid for the bat CoV HKU4 strain, and RNA from a bat fecal sample containing SARS-like bat CoV. The designed primers and probes were found to be specific for MERS-CoV and SARS-like bat CoV (B15-21) (Table 3). No positive reactions were found for the bat coronavirus strain HKU4, dengue virus, PRRSV, PEDV, or influenza A virus. In the case of the HKU4 strain, although the HEX fluorescence signal was not detected, a target band of primers for MERS-CoV was detected by agarose gel electrophoresis.
Table 3

Specificity of the real-time RT-PCR with various templates

VirusVirus familyTemplate typeAgarose gel electrophoresisCq
FAMHEX
Dengue virus 2 (KBPV-VR-29) Flaviviridae RNA---
Dengue virus 3 (KBPV-VR-30) Flaviviridae RNA---
Dengue virus 4 (KBPV-VR-31) Flaviviridae RNA---
PRRSV (CP401-9) Arteriviridae RNA---
PEDV (DR13) Coronaviridae RNA---
Influenza A virus (sk14) Orthomyxoviridae RNA---
MERS-CoV (KOR/KNIH/002_05_2015) Coronaviridae RNA+ - 15.64
MERS-like bat CoV (HKU4) Coronaviridae Plasmid+--
SARS-like bat CoV (B15-21) Coronaviridae RNA+12.55-
Specificity of the real-time RT-PCR with various templates

Application by the real-time RT-PCR method from bat fecal samples

The extracted RNAs from a total of 47 bat samples that were previously tested using consensus primer-based RT-PCR [11] were re-evaluated by the real-time RT-PCR method developed in this study. As shown in Table 4, six fecal samples were positive in the consensus-primer-based RT-PCR, and these were further sequenced and analyzed using BLASTn (http://www.ncbi.nlm.nih.gov). Among the six positive samples, five were found to be closely related to alphacoronaviruses, while one sample (B15-21) was closely related to SARS-like bat CoV [7]. In the real-time PCR developed in this study, five of the alphacoronavirus-positive samples were negative, but only one SARS-like bat CoV-positive sample was positive.
Table 4

Real-time RT-PCR results from bat samples

Positivea Negative
AlphacoronavirusSARS-like CoV
Real-time RT-PCRPositive010
Negative5041

Samples that were positive in the RT-PCR (11) were sequenced and analyzed using BLASTn (http://www.ncbi.nlm.nih.gov)

Real-time RT-PCR results from bat samples Samples that were positive in the RT-PCR (11) were sequenced and analyzed using BLASTn (http://www.ncbi.nlm.nih.gov)

Application of the real-time RT-PCR method for MERS patient samples

The RNA samples from the sputum of MERS patients were tested by the newly developed real-time RT-PCR method, and the results were compared to those obtained with UpE-based real-time RT-PCR as a reference gold-standard test (Table 5). Among the 20 samples tested, 17 were found to be positive by UpE-based real-time RT-PCR. When the single set of primers and probe for MERS-CoV was used, the newly developed real-time RT-PCR results showed 100% consistency with those of the reference test. However, when mixed sets of primers and probes for SARS-CoV and MERS-CoV were used, 85% of the results were consistent with those of the reference test. Three false-negative samples had Cq values of 33.9, 33.2, and 36.46, respectively, in the reference test.
Table 5

Results of duplex real-time RT-PCR with RNA extracted from MERS patients

Sample no.UpE-based real-time RT-PCRReal-time RT-PCR with single primers and probe set for MERS-CoVReal-time RT-PCR with mixed primers and probe sets for SARS-CoV and MERS-CoV
133.9a 35.1-
232.3933.2831.04
330.2730.8830.01
4-b --
531.6431.429.23
630.3930.5229.48
7---
834.6335.2433.67
9---
1033.234.99-
1124.223.9724.15
1229.4429.3629.39
1329.1430.4429.08
1421.221.4321.74
1531.5431.3731.25
1636.4635.63-
1730.4530.0130.64
1828.929.1529.83
1927.9328.1729.88
2033.4534.5135.19

Cq value with HEX dye

No HEX signal

Results of duplex real-time RT-PCR with RNA extracted from MERS patients Cq value with HEX dye No HEX signal

Discussion

The aim of this study was to develop a duplex real-time RT-PCR method for the simultaneous detection of SARS-CoV and MERS-CoV, which are recently emerged pathogens infecting humans. As these viruses are also closely related to several bat CoVs (SARS-like bat CoVs and MERS-related bat CoVs), the primers and probes were designed based on the conserved region of the spike S2 domains of SARS-CoV, MERS-CoV, and their related bat CoVs. In a previous study, two regions within the spike protein of SARS-CoV were identified that showed a high degree of sequence conservation with those of other CoVs [13]. By applying our new approach, we expected to detect not only SARS-CoV and MERS-CoV in humans but also several bat CoVs that are closely related to these viruses. Indeed, the newly developed real-time RT-PCR method could detect SARS-CoV- and MERS-CoV-specific sequences when the plasmids pUC57-SARS-pS2 and pGEM-MERS-S2 were used as templates. In addition, positive results were obtained when RNA extracted from MERS-CoV (KOR/KNIH/002_05_2015), which was isolated from a patient in Korea in 2015 [14], was used as a template. The new real-time RT-PCR method also showed positive results for RNA extracted from a fecal sample containing SARS-like bat CoV (B15-21) [7]. In addition, no nonspecific amplification was found when using RNAs obtained from several RNA viruses belonging to the families Flaviviridae, Arteriviridae, and Orthomyxoviridae, and the genus Alphacoronavirus. The specific binding of the primers and probes used in the real-time RT-PCR was evaluated to determine their detection limits using serially diluted plasmids (pUC57-SARS-pS2 and pGEM-MERS-S2). The detection limits of the real time RT-PCR were 50–100 copies/mL and 5–100 copies/mL for pUC57-SARS-pS2 and pGEM-MERS-S2, respectively. As the previous real-time RT-PCRs were able to detect as few as 1000 copies/mL and 291 copies/mL for SARS-CoV and their related bat CoVs and MERS-CoV, respectively [12, 15], the duplex real-time RT-PCR method developed in this study can be applied with an adequate limit of detection. When the real-time RT-PCR was applied to the bat samples, it could specifically detect a SARS-like bat CoV, and no positive reactions were obtained with the other samples, including alphacoronavirus-positive samples. The real-time RT-PCR method developed in this study could not detect the bat CoV HKU4 strain, which belongs to the same group as human MERS-CoV. Although these viruses are in the same group, the nucleotide sequences were found to be more variable than those of SARS-CoV and its related bat CoVs (Fig. 1). However, as a target band was found in the agarose gel electrophoresis of a PCR product from the HKU4 template, the negative result obtained with HKU4 in the real-time RT-PCR method developed in this study might have been due to inefficient binding of the probe. The new real-time RT-PCR method was also tested with RNAs extracted from the sputum of MERS patients in Korea. The single set of primers and probe for MERS-CoV could detect the same positive samples detected in the reference test, showing 100% agreement. However, when the mixed sets of primers and probes for SARS-CoV and MERS-CoV were used, the agreement was reduced to 85%. Three samples that tested positive in the reference test were not detected with the newly developed method. This may be due to the low amount of viral RNA in the sputum samples and the reduced sensitivity due to the duplex approach. In addition to using a standard test for the diagnosis of MERS-CoV infection in humans, consideration of additional tests would be helpful to avoid missing true positives. Collectively, the newly developed real-time RT-PCR method was demonstrated to be applicable for the simultaneous detection of SARS-CoV and MERS-CoV, showing an adequate detection limit and specificity. By testing the new method with bat samples as well as human samples, it could be applicable to survey SARS-CoV, MERS-CoV, and potentially their related bat CoVs in bats and human samples. It could successfully detect SARS-like CoV in bat samples but showed limited detection ability for the bat CoV HKU4 strain, which is related to MERS-CoV. However, according to a recent finding of EMC-like MERS-CoV, which was detected in bats of Saudi Arabia [18], we assume that the new method can be helpful for screening for MERS-CoV in bat samples. As novel CoVs continue to emerge around the world, real-time-PCR-based detection methods have been developed by pioneering researchers [12, 15–17]. These methods are applicable with good limits of detection and have been validated using field samples. However, simultaneous detection of SARS-CoV, MERS-CoV, and their related bat CoVs would be particularly useful for the detection of these viruses in humans as well as in bats, which are a known reservoir of the viruses. Therefore, the duplex real-time RT-PCR approach for the simultaneous detection of SARS-CoV and MERS-CoV developed in this study might allow more-convenient and rapid detection of these viruses in clinical samples. In addition, this new method can be used to closely monitor related bat CoVs in bat populations, which are considered to be a primary reservoir of these viruses.
  18 in total

1.  Detection of a novel human coronavirus by real-time reverse-transcription polymerase chain reaction.

Authors:  V M Corman; I Eckerle; T Bleicker; A Zaki; O Landt; M Eschbach-Bludau; S van Boheemen; R Gopal; M Ballhause; T M Bestebroer; D Muth; M A Müller; J F Drexler; M Zambon; A D Osterhaus; R M Fouchier; C Drosten
Journal:  Euro Surveill       Date:  2012-09-27

2.  Structural characterization of the fusion-active complex of severe acute respiratory syndrome (SARS) coronavirus.

Authors:  Paolo Ingallinella; Elisabetta Bianchi; Marco Finotto; Giovanna Cantoni; Debra M Eckert; Vinit M Supekar; Chiara Bruckmann; Andrea Carfi; Antonello Pessi
Journal:  Proc Natl Acad Sci U S A       Date:  2004-05-25       Impact factor: 11.205

3.  A real-time PCR assay for bat SARS-like coronavirus detection and its application to Italian greater horseshoe bat faecal sample surveys.

Authors:  Andrea Balboni; Laura Gallina; Alessandra Palladini; Santino Prosperi; Mara Battilani
Journal:  ScientificWorldJournal       Date:  2011-11-22

4.  Middle East respiratory syndrome coronavirus in bats, Saudi Arabia.

Authors:  Ziad A Memish; Nischay Mishra; Kevin J Olival; Shamsudeen F Fagbo; Vishal Kapoor; Jonathan H Epstein; Rafat Alhakeem; Abdulkareem Durosinloun; Mushabab Al Asmari; Ariful Islam; Amit Kapoor; Thomas Briese; Peter Daszak; Abdullah A Al Rabeeah; W Ian Lipkin
Journal:  Emerg Infect Dis       Date:  2013-11       Impact factor: 6.883

5.  Close relative of human Middle East respiratory syndrome coronavirus in bat, South Africa.

Authors:  Ndapewa Laudika Ithete; Samantha Stoffberg; Victor Max Corman; Veronika M Cottontail; Leigh Rosanne Richards; M Corrie Schoeman; Christian Drosten; Jan Felix Drexler; Wolfgang Preiser
Journal:  Emerg Infect Dis       Date:  2013-10       Impact factor: 6.883

6.  Isolation and characterization of a bat SARS-like coronavirus that uses the ACE2 receptor.

Authors:  Xing-Yi Ge; Jia-Lu Li; Xing-Lou Yang; Aleksei A Chmura; Guangjian Zhu; Jonathan H Epstein; Jonna K Mazet; Ben Hu; Wei Zhang; Cheng Peng; Yu-Ji Zhang; Chu-Ming Luo; Bing Tan; Ning Wang; Yan Zhu; Gary Crameri; Shu-Yi Zhang; Lin-Fa Wang; Peter Daszak; Zheng-Li Shi
Journal:  Nature       Date:  2013-10-30       Impact factor: 49.962

7.  Development of a molecular-beacon-based multi-allelic real-time RT-PCR assay for the detection of human coronavirus causing severe acute respiratory syndrome (SARS-CoV): a general methodology for detecting rapidly mutating viruses.

Authors:  Andreas V Hadjinicolaou; Gabriella A Farcas; Victoria L Demetriou; Tony Mazzulli; Susan M Poutanen; Barbara M Willey; Donald E Low; Jagdish Butany; Sylvia L Asa; Kevin C Kain; Leondios G Kostrikis
Journal:  Arch Virol       Date:  2011-01-11       Impact factor: 2.574

Review 8.  SARS and MERS: recent insights into emerging coronaviruses.

Authors:  Emmie de Wit; Neeltje van Doremalen; Darryl Falzarano; Vincent J Munster
Journal:  Nat Rev Microbiol       Date:  2016-06-27       Impact factor: 60.633

9.  Detection of Severe Acute Respiratory Syndrome-Like, Middle East Respiratory Syndrome-Like Bat Coronaviruses and Group H Rotavirus in Faeces of Korean Bats.

Authors:  H K Kim; S-W Yoon; D-J Kim; B-S Koo; J Y Noh; J H Kim; Y G Choi; W Na; K-T Chang; D Song; D G Jeong
Journal:  Transbound Emerg Dis       Date:  2016-05-23       Impact factor: 5.005

Review 10.  Ecology, evolution and classification of bat coronaviruses in the aftermath of SARS.

Authors:  Jan Felix Drexler; Victor Max Corman; Christian Drosten
Journal:  Antiviral Res       Date:  2013-10-31       Impact factor: 5.970

View more
  31 in total

1.  Risk factors of COVID-19 vertical transmission among pregnant and non-pregnant Filipinos in Metro Manila: a multicentre cohort study protocol.

Authors:  Erlidia F Llamas-Clark; Rebecca B Singson; Francisco M Heralde; Mayan U Lumandas; Maria Esterlita T Villanueva-Uy; Maria Stephanie Fay S Cagayan; Zypher Jude G Regencia; Emmanuel S Baja
Journal:  BMJ Open       Date:  2022-04-25       Impact factor: 3.006

Review 2.  MERS-CoV: Understanding the Latest Human Coronavirus Threat.

Authors:  Aasiyah Chafekar; Burtram C Fielding
Journal:  Viruses       Date:  2018-02-24       Impact factor: 5.048

3.  A one-step, one-tube real-time RT-PCR based assay with an automated analysis for detection of SARS-CoV-2.

Authors:  Bhasker Dharavath; Neelima Yadav; Sanket Desai; Roma Sunder; Rohit Mishra; Madhura Ketkar; Prasanna Bhanshe; Anurodh Gupta; Archana Kumari Redhu; Nikhil Patkar; Shilpee Dutt; Sudeep Gupta; Amit Dutt
Journal:  Heliyon       Date:  2020-07-07

4.  Commentary: Phyllostomid bat microbiome composition is associated to host phylogeny and feeding strategies.

Authors:  Arinjay Banerjee; Edel Pérez-López; Karen Mossman
Journal:  Front Microbiol       Date:  2018-11-22       Impact factor: 5.640

Review 5.  Sensitivity-Enhancing Strategies in Optical Biosensing.

Authors:  Youngsun Kim; John Gonzales; Yuebing Zheng
Journal:  Small       Date:  2020-12-28       Impact factor: 13.281

Review 6.  Recent advances in detection technologies for COVID-19.

Authors:  Tingting Han; Hailin Cong; Youqing Shen; Bing Yu
Journal:  Talanta       Date:  2021-06-12       Impact factor: 6.057

7.  SARS-CoV-2 exhibits intra-host genomic plasticity and low-frequency polymorphic quasispecies.

Authors:  Timokratis Karamitros; Gethsimani Papadopoulou; Maria Bousali; Anastasios Mexias; Sotirios Tsiodras; Andreas Mentis
Journal:  J Clin Virol       Date:  2020-08-11       Impact factor: 3.168

8.  Real-time RT-PCR in COVID-19 detection: issues affecting the results.

Authors:  Alireza Tahamtan; Abdollah Ardebili
Journal:  Expert Rev Mol Diagn       Date:  2020-04-22       Impact factor: 5.225

Review 9.  Recent advances and perspectives of nucleic acid detection for coronavirus.

Authors:  Minzhe Shen; Ying Zhou; Jiawei Ye; Abdu Ahmed Abdullah Al-Maskri; Yu Kang; Su Zeng; Sheng Cai
Journal:  J Pharm Anal       Date:  2020-03-01

Review 10.  Measures for diagnosing and treating infections by a novel coronavirus responsible for a pneumonia outbreak originating in Wuhan, China.

Authors:  Fei Yu; Lanying Du; David M Ojcius; Chungen Pan; Shibo Jiang
Journal:  Microbes Infect       Date:  2020-02-01       Impact factor: 2.700

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.