Literature DB >> 28216272

Comparison of Cyclic Fatigue Resistance of WaveOne and WaveOne Gold Small, Primary, and Large Instruments.

Mehmet Adıgüzel1, Ismail Davut Capar2.   

Abstract

INTRODUCTION: To compare the cyclic fatigue of resistance of WaveOne (WO; Dentsply Maillefer, Ballaigues, Switzerland) and WaveOne Gold (WOG, Dentsply Maillefer) small, primary, and large instruments in simulated root canals. Each instrument was rotated until fracture occurred.
METHODS: One hundred twenty nickel-titanium endodontic instruments were tested in two different curved artificial canals with different angles and radii of curvatures. WO and WOG small, primary, and large instruments were evaluated (n = 10 for each curvature). Each instrument was rotated until fracture occurred. The time to failure was recorded, and the length of the fractured tip was measured. An independent Student't test was used to compare the means between the 2 groups. One-way analysis of variance and Tukey' post hoc tests were used for multiple comparisons (P < .05).
RESULTS: At all the sizes tested and both curvatures, the cyclic fatigue resistance of the WOG instruments was higher than that of the WO instruments of corresponding sizes (P < .001). The WOG instruments were approximately twice as resistant to failure as the WO instruments were at a 60° angle of curvature, whereas the WOG instruments were 3 times more resistant at a 90° angle of curvature. The length of the fractured part of the instruments was similar among all the groups (P > .05).
CONCLUSIONS: Within the limitations of this study, WOG instruments were more resistant to cyclic fatigue than WO instruments.
Copyright © 2016 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  M-Wire; WaveOne Gold; endodontics; reciprocating instruments

Mesh:

Substances:

Year:  2017        PMID: 28216272     DOI: 10.1016/j.joen.2016.11.021

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  20 in total

1.  Shaping ability of reciprocating single-file systems in severely curved canals: WaveOne and Reciproc versus WaveOne Gold and Reciproc blue.

Authors:  Sebastian Bürklein; Stefanie Flüch; Edgar Schäfer
Journal:  Odontology       Date:  2018-05-18       Impact factor: 2.634

2.  Canal transportation and centering ratio after preparation in severely curved canals: analysis by micro-computed tomography and double-digital radiography.

Authors:  Ane Poly; Fouad AlMalki; Fernando Marques; Bekir Karabucak
Journal:  Clin Oral Investig       Date:  2019-03-05       Impact factor: 3.573

3.  Analysis of the Importance of the Motion Used in the Resistance of Different Mechanical Instrumentation Systems in Endodontics: A Comparative Study.

Authors:  Jesús Mena-Álvarez; Manuel Almanzor-López; Norberto Quispe-López; Ana De Pedro-Muñoz; Cristina Rico-Romano
Journal:  Materials (Basel)       Date:  2022-06-24       Impact factor: 3.748

4.  Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments.

Authors:  Emmanuel João Nogueira Leal Silva; Victor Talarico Leal Vieira; Fernanda Hecksher; Marja Ribeiro Silva Dos Santos Oliveira; Henrique Dos Santos Antunes; Edson Jorge Lima Moreira
Journal:  Clin Oral Investig       Date:  2018-02-01       Impact factor: 3.573

5.  Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments.

Authors:  Cangül Keskin; Uğur İnan; Murat Demiral; Ali Keleş
Journal:  Clin Oral Investig       Date:  2018-02-17       Impact factor: 3.573

6.  Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal.

Authors:  Taha Özyürek; Mustafa Gündoğar; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-12-13

7.  Cyclic Fatigue Resistance of WaveOne Gold, ProDesign R and ProDesign Logic Files in Curved Canals In Vitro.

Authors:  Sílvio Emanuel Acioly Conrado de Menezes; Shirley Machado Batista; Juliana Ouro Preto Lira; Gabriela Queiroz de Melo Monteiro
Journal:  Iran Endod J       Date:  2017

Review 8.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

9.  Torsional Behavior of WaveOne Gold Endodontic File with the Dedicated Motor of the Original WaveOne File.

Authors:  Jung-Hong Ha; Asgeir Sigurdsson; Gustavo De-Deus; Antheunis Versluis; Sang Won Kwak; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2018-07-06       Impact factor: 3.623

10.  Comparison of the ability of Reciproc and Reciproc Blue instruments to reach the full working length with or without glide path preparation.

Authors:  Mehmet Adıguzel; Pelin Tufenkci
Journal:  Restor Dent Endod       Date:  2018-11-01