Improved electromagnetic tracking for catheter path reconstruction with application in high-dose-rate brachytherapy.

Elodie Lugez1, Hossein Sadjadi2, Chandra P Joshi3, Selim G Akl4, Gabor Fichtinger2.   

Abstract

PURPOSE: Electromagnetic (EM) catheter tracking has recently been introduced in order to enable prompt and uncomplicated reconstruction of catheter paths in various clinical interventions. However, EM tracking is prone to measurement errors which can compromise the outcome of the procedure. Minimizing catheter tracking errors is therefore paramount to improve the path reconstruction accuracy.
METHODS: An extended Kalman filter (EKF) was employed to combine the nonlinear kinematic model of an EM sensor inside the catheter, with both its position and orientation measurements. The formulation of the kinematic model was based on the nonholonomic motion constraints of the EM sensor inside the catheter. Experimental verification was carried out in a clinical HDR suite. Ten catheters were inserted with mean curvatures varying from 0 to [Formula: see text] in a phantom. A miniaturized Ascension (Burlington, Vermont, USA) trakSTAR EM sensor (model 55) was threaded within each catheter at various speeds ranging from 7.4 to [Formula: see text]. The nonholonomic EKF was applied on the tracking data in order to statistically improve the EM tracking accuracy. A sample reconstruction error was defined at each point as the Euclidean distance between the estimated EM measurement and its corresponding ground truth. A path reconstruction accuracy was defined as the root mean square of the sample reconstruction errors, while the path reconstruction precision was defined as the standard deviation of these sample reconstruction errors. The impacts of sensor velocity and path curvature on the nonholonomic EKF method were determined. Finally, the nonholonomic EKF catheter path reconstructions were compared with the reconstructions provided by the manufacturer's filters under default settings, namely the AC wide notch and the DC adaptive filter.
RESULTS: With a path reconstruction accuracy of 1.9 mm, the nonholonomic EKF surpassed the performance of the manufacturer's filters (2.4 mm) by 21% and the raw EM measurements (3.5 mm) by 46%. Similarly, with a path reconstruction precision of 0.8 mm, the nonholonomic EKF surpassed the performance of the manufacturer's filters (1.0 mm) by 20% and the raw EM measurements (1.7 mm) by 53%. Path reconstruction accuracies did not follow an apparent trend when varying the path curvature and sensor velocity; instead, reconstruction accuracies were predominantly impacted by the position of the EM field transmitter ([Formula: see text]).
CONCLUSION: The advanced nonholonomic EKF is effective in reducing EM measurement errors when reconstructing catheter paths, is robust to path curvature and sensor speed, and runs in real time. Our approach is promising for a plurality of clinical procedures requiring catheter reconstructions, such as cardiovascular interventions, pulmonary applications (Bender et al. in medical image computing and computer-assisted intervention-MICCAI 99. Springer, Berlin, pp 981-989, 1999), and brachytherapy.

Keywords:  Catheter path reconstruction; Electromagnetic tracking; High-dose-rate brachytherapy; Nonholonomic

Mesh:

Year:  2017        PMID: 28215003     DOI: 10.1007/s11548-017-1534-4

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  20 in total

1.  Accuracy of electromagnetic tracking with a prototype field generator in an interventional OR setting.

Authors:  Lars Eirik Bø; Håkon Olav Leira; Geir Arne Tangen; Erlend Fagertun Hofstad; Tore Amundsen; Thomas Langø
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Simultaneous Electromagnetic Tracking and Calibration for Dynamic Field Distortion Compensation.

Authors:  Hossein Sadjadi; Keyvan Hashtrudi-Zaad; Gabor Fichtinger
Journal:  IEEE Trans Biomed Eng       Date:  2015-11-19       Impact factor: 4.538

3.  Magneto-optical tracking of flexible laparoscopic ultrasound: model-based online detection and correction of magnetic tracking errors.

Authors:  Marco Feuerstein; Tobias Reichl; Jakob Vogel; Joerg Traub; Nassir Navab
Journal:  IEEE Trans Med Imaging       Date:  2009-02-10       Impact factor: 10.048

4.  A small tolerance for catheter displacement in high-dose rate prostate brachytherapy is necessary and feasible.

Authors:  Albert Tiong; Sean Bydder; Martin Ebert; Nikki Caswell; David Waterhouse; Nigel Spry; Peter Camille; David Joseph
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-07-18       Impact factor: 7.038

5.  Prostate HDR brachytherapy catheter displacement between planning and treatment delivery.

Authors:  May Whitaker; George Hruby; Aimee Lovett; Nitya Patanjali
Journal:  Radiother Oncol       Date:  2011-08-31       Impact factor: 6.280

6.  Least-squares fitting of two 3-d point sets.

Authors:  K S Arun; T S Huang; S D Blostein
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1987-05       Impact factor: 6.226

7.  Electromagnetic tracking in surgical and interventional environments: usability study.

Authors:  Elodie Lugez; Hossein Sadjadi; David R Pichora; Randy E Ellis; Selim G Akl; Gabor Fichtinger
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-09-06       Impact factor: 2.924

8.  A system to use electromagnetic tracking for the quality assurance of brachytherapy catheter digitization.

Authors:  Antonio L Damato; Akila N Viswanathan; Sarah M Don; Jorgen L Hansen; Robert A Cormack
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

9.  Electromagnetic tracking for catheter reconstruction in ultrasound-guided high-dose-rate brachytherapy of the prostate.

Authors:  Shyam Bharat; Cynthia Kung; Ehsan Dehghan; Ananth Ravi; Niranjan Venugopal; Antonio Bonillas; Doug Stanton; Jochen Kruecker
Journal:  Brachytherapy       Date:  2014-06-11       Impact factor: 2.362

10.  Simultaneous tracking of catheters and guidewires: comparison to standard fluoroscopic guidance for arterial cannulation.

Authors:  S Condino; E M Calabrò; A Alberti; S Parrini; R Cioni; R N Berchiolli; M Gesi; V Ferrari; M Ferrari
Journal:  Eur J Vasc Endovasc Surg       Date:  2013-10-07       Impact factor: 7.069

View more
  3 in total

Review 1.  Review of strategies for MRI based reconstruction of endocavitary and interstitial applicators in brachytherapy of cervical cancer.

Authors:  José Richart; Vicente Carmona-Meseguer; Teresa García-Martínez; Antonio Herreros; Antonio Otal; Santiago Pellejero; Ana Tornero-López; José Pérez-Calatayud
Journal:  Rep Pract Oncol Radiother       Date:  2018-07-23

2.  Real-time microrobot posture recognition via biplane X-ray imaging system for external electromagnetic actuation.

Authors:  Phu Bao Nguyen; Byungjeon Kang; D M Bappy; Eunpyo Choi; Sukho Park; Seong Young Ko; Jong-Oh Park; Chang-Sei Kim
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-08-20       Impact factor: 2.924

3.  3D localization from 2D X-ray projection.

Authors:  Dagmar Bertsche; Volker Rasche; Wolfgang Rottbauer; Ina Vernikouskaya
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-07-11       Impact factor: 3.421

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.