| Literature DB >> 28191220 |
Jessie T Childs1, Adrian J Esterman2, Kerry A Thoirs3.
Abstract
Introduction: Ultrasound is an easy and inexpensive method to rapidly assess the size of the adult liver. The literature addressing reliability of liver measurements using ultrasound is poorly reported and inadequate. In this study, intra and inter-rater reliability of multiple measurements of the right lobe, left lobe and entire adult liver were assessed.Entities:
Keywords: liver; measurement; reliability; ultrasound
Year: 2015 PMID: 28191220 PMCID: PMC5024948 DOI: 10.1002/j.2205-0140.2014.tb00026.x
Source DB: PubMed Journal: Australas J Ultrasound Med ISSN: 1836-6864
Ultrasound liver Measurements.
| Measurement Name | Image Measurement Source | Description of Measurement | Example of Measurement |
|---|---|---|---|
|
| |||
| MaxLong | Image 1 | Maximum longitudinal Diameter seen (right or left lobe whichever greater) |
|
| MaxAP | Image 1 | Maximum anteroposterior diameter seen (right or left lobe whichever greater) |
|
| MaxTrans | Image 2 | Maximum transverse diameter |
|
|
| |||
| MCL Dome to Tip | Image 3 | The diameter of the liver from the dome to the tip |
|
| MCL Area | Image 3 | The area of the liver seen on the screen in the mid‐clavicular line |
|
| MCL Perimeter | Image 3 | The distance around the edge of the liver seen on the screen in the mid‐clavicular line |
|
| MCL Max Long | Image 3 | The maximum longitudinal diameter of liver on the screen, measured left to right along the most superior aspect of the screen in the mid‐clavicular line |
|
| MCL Mid AP | Image 3 | The midline of measurement 7 was obtained using the machine's inbuilt callipers and the anteroposterior diameter of the liver measured at this point in the mid‐clavicular line |
|
| MCLMaxAP | Image 3 | The maximum anteroposterior diameter of the liver seen on this image |
|
|
| |||
| Midline area | Image 4 | The area of the liver seen on the screen in the midline |
|
| Midline Perimeter | Image 4 | The distance around the edge of the liver seen on this screen in the midline |
|
| Midline Max Long | Image 4 | The maximum longitudinal diameter of the liver on the screen, measured left to right along the most superior aspect of the screen |
|
| Midline Mid AP | Image 4 | The midpoint of measurement 12 was obtained using the machine's inbuilt callipers and the anteroposterior diameter of the liver was measured at this point in the midline |
|
| Midline Max AP | Image 4 | The maximum anteroposterior diameter of the liver seen on this image |
|
| MHV AP | Image 5 | The anteroposterior diameter of the liver through the middle hepatic vein |
|
Key: Image 1: With the transducer in a longitudinal orientation. An image of the liver was taken at its perceived largest longitudinal diameter. Image 2: With the transducer in a transverse orientation. A panoramic transverse image of the liver was taken to encompass the maximum transverse diameter. Image 3: The right clavicle of each patient was measured with a ruler and the midpoint determined. This was deemed the mid‐clavicular line. An image was taken with the transducer orientated longitudinally into the mid‐clavicular line. This anatomical position best represented the right lobe of the liver. Image 4: The midline of each patient was determined using the xiphisternum as an anatomic marker. An image was taken with the transducer orientated longitudinally along the midline. This anatomical position best represented the left lobe of the liver. Image 5: A transverse image of the liver was taken at the confluence of the three hepatic veins with the middle hepatic vein horizontal in the centre of the screen.
Mean measurements.
| Measurement | Mean | SD | |
|---|---|---|---|
|
| 1. Max Long | 13.7 cm | 1.42 cm |
| 2. Max AP | 11.4 cm | 1.94 cm | |
| 3. Max Trans | 20.6 cm | 1.91 cm | |
|
| 4. MCL Dome to Tip | 14.1 cm | 1.63 cm |
| 5.MCL Area | 98.4 cm2 | 30.84 cm | |
| 6. MCL Perimeter | 40.72 cm | 4.9 cm | |
| 7. MCL Max Long | 11.84 cm | 1.32 cm | |
| 8. MCL MidAP | 11.27 cm | 1.11 cm | |
| 9. MCL Max AP | 11.33 cm | 1.42 cm | |
|
| 10. Midline Area | 43.35 cm2 | 9.89 cm |
| 11. Midline Perimeter | 27.47 cm | 2.95 cm | |
| 12. Midline Max Long | 9.03 cm | 1.07 cm | |
| 13. Midline Mid AP | 6.96 cm | 1.47 cm | |
| 14. Midline Max AP | 7.04 cm | 1.36 cm | |
| 15. MHV AP | 7.21 cm | 0.96 cm |
* Means and Standard Deviations are based on the first examiners first image series measurements
Results of ICC analysis.
| Measurement | Inter‐rater ICC | Inter‐rater 95% Confidence Interval | Intra‐rater ICC | Intra‐rater 95% confidence interval | |
|---|---|---|---|---|---|
|
| 1. Max Long | 0.795 | 0.444 to 0.936 | 0.875 | 0.626 to 0.962 |
| 2. Max AP | 0.853 | 0.528 to 0.957 | 0.962 | 0.880 to 0.989 | |
| 3. Max Trans | 0.453 | −0.062 to 0.797 | 0.786 | 0.404 to 0.934 | |
|
| 4. MCL Dome to Tip | 0.930 | 0.785 to 0.979 | 0.963 | 0.883 to 0.989 |
| 5.MCL Area | 0.954 | 0.853 to 0.987 | 0.968 | 0.825 to 0.992 | |
| 6. MCL Perimeter | 0.943 | 0.818 to 0.984 | 0.973 | 0.879 to 0.993 | |
| 7. MCL Max Long | 0.808 | 0.427 to 0.942 | 0.901 | 0.691 to 0.970 | |
| 8. MCL MidAP | 0.716 | 0.291 to 0.908 | 0.903 | 0.700 to 0.971 | |
| 9. MCL Max AP | 0.834 | 0.520 to 0.949 | 0.974 | 0.912 to 0.992 | |
|
| 10. Midline Area | 0.611 | 0.119 to 0.868 | 0.833 | 0.503 to 0.949 |
| 11. Midline Perimeter | 0.574 | 0.0326 to 0.856 | 0.942 | 0.812 to 0.983 | |
| 12. Midline Max Long | 0.419 | −0.107 to 0.781 | 0.911 | 0.726 to 0.973 | |
| 13. Midline Mid AP | 0.625 | 0.084 to 0.877 | 0.678 | 0.208 to 0.895 | |
| 14. Midline Max AP | 0.865 | 0.593 to 0.959 | 0.909 | 0.724 to 0.973 | |
| 15. MHV AP | 0.412 | −0.154 to 0.783 | 0.721 | 0.303 to 0.909 |
Results of Bland Altman plots and t‐tests.
| Measurement | Mean Bias Intra‐rater | Mean Bias inter‐rater | Significance t‐test intrarater | Significance t‐test interrater | Limits of agreement intra‐rater | Limits of agreement inter‐rater | Pattern of bias intrarater | Pattern of bias inter‐rater | |
|---|---|---|---|---|---|---|---|---|---|
|
| 1. Max Long | −0.10 cm | 0.40 cm | 0.63 | 0.14 | 3.0 cm | 3.5 cm | None | Increases with decreasing liver size |
| 2. Max AP | −0.17 cm | −0.60 cm | 0.28 | 0.06 | 2.0 cm | 3.5 cm | Error increases with increasing liver size | None | |
|
| 4. MCL Dome to Tip | −0.13 cm | −0.21 cm | 0.32 | 0.31 | 1.6 cm | 2.7 cm | Error increase with increasing liver size | None |
| 5. MCL Area | −5.10 cm2 | 3.50 cm2 | 0.02 | 0.20 | 25.5 cm2 | 34.6 cm2 | None | None | |
| 6. MCL Perimeter | −0.60 cm | −0.10 cm | 0.04 | 0.86 | 3.7 cm | 6.6 cm | Error increase with increasing liver size | None | |
| 7. MCL Max Long | −0.04 cm | −0.50 cm | 0.80 | 0.06 | 2.1 cm | 3.3 cm | None | None | |
| 8. MCL Midline AP | −0.07 cm | −0.50 | 0.69 | 0.13 | 2.2 cm | 3.7 cm | None | None | |
| 9. MCL Max AP | 0.14 cm | 0.10 | 0.32 | 0.74 | 1.9 cm | 3.3 cm | None | None | |
|
| 14. Midline Max AP | 0.21 cm | 0.01 | 0.21 | 0.89 | 2.1 cm | 3.1 cm | None | None |