Minwook Yoo1, Sung-Chul Jin1, Seung-Hwan Kim1, Byeong-Sam Choi1, Hae Yu Kim1, SungJun Lee1, Sung Tae Kim2, Hae Woong Jeong3.
Abstract
OBJECTIVE: Excelsior XT-27 (Stryker Neurovascular, Fremont, CA, USA) or Rebar 27 (eV3 Covidien, Irvine, CA, USA) microcatheters have recently been used to overcome the limitations of the Renegade Hi-Flo microcatheter such as interference between two microcatheters (one for stent delivery and the other for cerebral aneurysm coiling) during the jailing technique. We evaluated differences and influential factors related to the interference phenomenon according to these two microcatheters group.Entities:
Keywords: ICA; Microcatheter; Open cell stent; Side wall aneurysm
Year: 2016 PMID: 28184346 PMCID: PMC5298978 DOI: 10.7461/jcen.2016.18.4.363
Source DB: PubMed Journal: J Cerebrovasc Endovasc Neurosurg ISSN: 2234-8565
Comparison of the demographic and aneurysmal characteristics of the Renegade Hi-Flo microcatheter (n = 22) and Rebar 27 or XT 27 microcatheter (n = 72) groups
| Renegade Hi-Flo microcatheter | Rebar 27 or XT 27 microcatheter | ||
|---|---|---|---|
| Gender | 0.537 | ||
| Male | 5 | 18 | |
| Female | 17 | 54 | |
| Age (years) | 56.7 ± 8.8 | 56.5 ± 11.1 | 0.949 |
| Aneurysm location | 0.547 | ||
| Paraclinoid segment | 18 | 60 | |
| p-com segment | 4 | 12 | |
| Carotid siphon angle | 75.8 ± 30.0 | 84.3 ± 30.0 | 0.223 |
| Aneurysm height (mm) | 3.6 ± 1.3 | 3.7 ± 1.7 | 0.667 |
| Aneurysm neck size (mm) | 3.2 ± 0.8 | 3.7 ± 1.6 | 0.179 |
| Aneurysm width (mm) | 3.6 ± 1.1 | 4.5 ± 2.3 | 0.015 |
Values are presented as number or mean ± standard deviation
Univariate analysis of the factors influencing the interference between the two microcatheters during the jailing technique using the Neuroform EZ stent system
| Presence of interference | Absence of interference | ||
|---|---|---|---|
| Carotid siphon angle | 51.2 ± 27.6 | 85.4 ± 27.0 | 0.002 |
| Delivery catheter | 0.000 | ||
| Renegade Hi-Flo | 6 | 15 | |
| Rebar 27 or XT 27 | 1 | 71 | |
| Age of patients | 0.707 | ||
| > 60 years | 2 | 32 | |
| ≤ 60 years | 5 | 53 |
Values are presented as number or mean ± standard deviation
Multiple logistic regression analysis of the factors influencing the interference between the two microcatheters during the jailing technique using the Neuroform EZ stent system
| Factors | Adjusted OR | Adjusted 95% CI | |
|---|---|---|---|
| Carotid siphon angle | 0.959 | 0.922-0.997 | 0.035 |
| Rebar 27 or XT 27 microcatheter | 32.277 | 3.138-311.729 | 0.003 |
OR = odds ratio; CI = confidence interval