Literature DB >> 28152195

A 2D silicon detector array for quality assurance in small field dosimetry: DUO.

Khalsa Al Shukaili1,2, Marco Petasecca1, Matthew Newall1, Anthony Espinoza1, Vladimir L Perevertaylo3, Stéphanie Corde1,4, Michael Lerch1,5, Anatoly B Rosenfeld1,5.   

Abstract

PURPOSE: Nowadays, there are many different applications that use small fields in radiotherapy treatments. The dosimetry of small radiation fields is not trivial due to the problems associated with lateral disequilibrium and source occlusion and requires reliable quality assurance (QA). Ideally such a QA tool should provide high spatial resolution, minimal beam perturbation and real time fast measurements. Many different types of silicon diode arrays are used for QA in radiotherapy; however, their application in small filed dosimetry is limited, in part, due to a lack of spatial resolution. The Center of Medical Radiation Physics (CMRP) has developed a new generation of a monolithic silicon diode array detector that will be useful for small field dosimetry in SRS/SRT. The objective of this study is to characterize a monolithic silicon diode array designed for dosimetry QA in SRS/SRT named DUO that is arranged as two orthogonal 1D arrays with 0.2 mm pitch.
METHODS: DUO is two orthogonal 1D silicon detector arrays in a monolithic crystal. Each orthogonal array contains 253 small pixels with size 0.04 × 0.8 mm2 and three central pixels are with a size of 0.18 × 0.18 mm2 each. The detector pitch is 0.2 mm and total active area is 52 × 52 mm2 . The response of the DUO silicon detector was characterized in terms of dose per pulse, percentage depth dose, and spatial resolution in a radiation field incorporating high gradients. Beam profile of small fields and output factors measured on a Varian 2100EX LINAC in a 6 MV radiation fields of square dimensions and sized from 0.5 × 0.5 cm2 to 5 × 5 cm2 . The DUO response was compared under the same conditions with EBT3 films and an ionization chamber.
RESULTS: The DUO detector shows a dose per pulse dependence of 5% for a range of dose rates from 2.7 × 10-4 to 1.2 × 10-4 Gy/pulse and 23% when the rate is further reduced to 2.8 × 10-5 Gy/pulse. The percentage depth dose measured to 25 cm depth in solid water phantom beyond the surface and for a field size of 10 × 10 cm2 agrees with that measured using a Markus IC within 1.5%. The beam profiles in both X and Y orthogonal directions showed a good match with EBT3 film, where the FWHM agreed within 1% and penumbra widths within 0.5 mm. The effect of an air gap above the DUO detector has also been studied. The output factor for field sizes ranging from 0.5 × 0.5 cm2 to 5 × 5 cm2 measured by the DUO detector with a 0.5 mm air gap above silicon surface agrees with EBT3 film and MOSkin detectors within 1.8%.
CONCLUSIONS: The CMRP's monolithic silicon detector array, DUO, is suitable for SRS/SRT dosimetry and QA because of its very high spatial resolution (0.2 mm) and real time operation.
© 2016 American Association of Physicists in Medicine.

Entities:  

Keywords:  build-up caps; output factors; silicon detector; small field dosimetry; stereotactic radiotherapy

Mesh:

Substances:

Year:  2017        PMID: 28152195     DOI: 10.1002/mp.12060

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Quality assurance of VMAT on flattened and flattening filter-free accelerators using a high spatial resolution detector.

Authors:  F S Matar; D Wilkinson; J Davis; G Biasi; T Causer; I Fuduli; O Brace; N Stansook; M Carolan; A B Rosenfeld; Marco Petasecca
Journal:  J Appl Clin Med Phys       Date:  2020-04-11       Impact factor: 2.102

2.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

3.  Real-time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy.

Authors:  Mitchell Duncan; Matthew K Newall; Vincent Caillet; Jeremy T Booth; Paul J Keall; Michael Lerch; Vladimir Perevertaylo; Anatoly B Rosenfeld; Marco Petasecca
Journal:  J Appl Clin Med Phys       Date:  2018-06-05       Impact factor: 2.102

4.  Consistency of small-field dosimetry, on and off axis, in beam-matched linacs used for stereotactic radiosurgery.

Authors:  Luis Muñoz; Tomas Kron; Marco Petasecca; Joseph Bucci; Michael Jackson; Peter Metcalfe; Anatoly B Rosenfeld; Giordano Biasi
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.