Failure of Polyethelene Insert Locking Mechanism after a Posterior Stabilised Total Knee Arthroplasty- A Case Report.

Gurava Reddy Ay1, Soundar Rajan D1, Chiranjeevi T1, Karthik C1, Krishna Kiran E2.   

Abstract

INTRODUCTION: Disengagement of polyethylene insert used in total knee arthroplasty is a rare but serious complication. Still rarer is disengagement because of failure of tibial insert locking mechanism. We report a previously unpublished complication of polyethylene insert locking mechanism failure in a 10-months-old posterior stabilized total knee arthroplasty in a 70-year-old woman with osteoarthritis for whom Attune (Depuy) knee implant was used. CASE
PRESENTATION: A 70-year-old female underwent (Attune, Depuy) primary bilateral posterior stabilised total knee arthroplasty in a private hospital. The patient did not have any complaints and had had been functioning well post her arthroplasty. After five months of surgery she had a fall and sustained injury over right hip which was treated with Cemented Bipolar Hemiarthroplasty. Ten months after index surgery, she sustained trivial fall and presented to the same hospital with knee pain and swelling, where the right knee prosthesis was found to be dislocated. An attempted closed reduction under anaesthesia failed, after which she was referred to our centre with an unstable, painful, swollen right knee in a long knee brace. The physical examination at the time of admission showed posterior sag of the tibia, fullness in the postero-lateral corner, quadriceps muscle atrophy without any neurovascular deficit oflower leg. Postero-lateral dislocation was confirmed with radiographs. Surgical error as a possible causative factor was excluded because patient had been functioning well after surgery. Her comorbidities included hypertension and hyponatremia. ESR and CRP were within normal limits. An open reduction surgery was planned. On exposure, polyethylene was found in the postero-lateral corner of the knee. We were not sure that revising the polyethylene alone would suffice as the poly and locking mechanism was of a relatively new design and hence it was decided to proceed with revision of the components. Revision was done with stemmed components, distal femoral augments and a constrained prosthesis (Total Condylar 3, Depuy). Intraoperative cultures were negative. The patient had an uncomplicated post-operative course.
CONCLUSION: Disengagement of polyethylene should be considered as a differential diagnosis in patients who present with acute swelling and instability of the knee. Though it's a rare complication, there is a possibility that design of the implant or its locking mechanism could contribute to dislocations in future.

Entities:  

Keywords:  Disengagement of polyethylene insert; Fixed-bearing; Total knee arthroplasty; locking mechanism failure

Year:  2016        PMID: 28116264      PMCID: PMC5245933          DOI: 10.13107/jocr.2250-0685.492

Source DB:  PubMed          Journal:  J Orthop Case Rep        ISSN: 2250-0685


Disengagement of polyethylene should be a differential diagnosis in patients who present with acute swelling and instability of the knee. Though it’s a rare complication it is quite possible that design of the implant or its locking mechanism could contribute to dislocation.

Introduction

The disadvantage of modular total knee arthroplasty is ‘backside wear’ leading to periarticular osteolysis [1,12]. Tibial insert locking mechanisms are intended to limit interface motion and prevent backside wear [1, 2]. Disengagement of the locking mechanism is a rare cause of early revision and needs to be recognized [2]. The purpose of this case report is to present an unusual case of polyethylene insert locking mechanism failure after a posterior stabilised total knee arthroplasty using an Attune knee and a Vitamin E polyethylene insert (AOX), and to review literature. To our knowledge, this is the first reported case of polyethylene insert dislocation of an Attune knee. An open reduction surgery was planned. On exposure, polyethylene was found in the postero-lateral corner of the knee (Fig.4). We were not sure that revising the polyethylene alone would suffice as the poly and locking mechanism was of a relatively new design and hence, it was decided to proceed with revision of the components. Revision was done with stemmed components, distal femoral augments and a constrained prosthesis(Fig.5) (Total Condylar 3, Depuy). Intraoperative cultures were negative. The patient had an uncomplicated postoperative course
Figure 4

Dislocated polyethelene in postero-lateral corner.

Figure 5

Retrieved components showing fatigue wear of locking mechanism in polyethylene.

Immediate post op radiographs after index surgery. Dislocated Right Knee prosthesis with disengagement of polyethelene Clinical picture Right knee with posterior sag. Dislocated polyethelene in postero-lateral corner. Retrieved components showing fatigue wear of locking mechanism in polyethylene. Post-operative radiograph with stemmed, constrained component(TC 3, Depuy). Locking mechanism in Attune system(DepuySynthes)

Case Presentation

A 70-year-old female underwent (Attune, Depuy) primary bilateral posterior stabilised total knee arthroplasty (May ‘2014) for osteoarthritis in a private hospital. The femoral and tibial components were size 4 and polyethylene size was 10mm. Immediate post-operative X-ray shown in (Fig.1). She had been functioning well post her arthroplasty for five months, when she had a fall and sustained injury over right hip and was treated with Cemented Bipolar Hemiarthroplasty (Sep’2014). Ten months after index surgery, she sustained trivial fall and presented to the same hospital with knee pain and swelling, where the right knee prosthesis was found to be dislocated (Fig.2). An attempted closed reduction under anaesthesia failed, after which she was referred to our centre with an unstable, painful, swollen right knee in a long knee brace. The physical examination at the time of admission showed posterior sag of the tibia (Fig.3), fullness in the postero-lateral corner, quadriceps muscle atrophy without any neurovascular deficit of lower leg. Postero-lateral dislocation was confirmed with radiographs. Surgical error as a possible causative factor was excluded because patient had been functioning well after surgery. Her comorbidities included hypertension and hyponatremia. ESR and CRP were within normal limits.
Figure 1

Immediate post op radiographs after index surgery.

Figure 2

Dislocated Right Knee prosthesis with disengagement of polyethelene

Figure 3

Clinical picture Right knee with posterior sag.

Discussion

The advantages of modularity of the components are to increase intraoperative flexibility, facilitate revision surgery with poly exchange and to simplify inventory management [3, 4]. Modularity in total knee arthroplasty mandated the development of locking mechanisms to capture the polyethylene into the metal base plate [5, 6]. Locking mechanism is intended to reduce interface motion and to prevent disengagement of the insert. However, rarely it may lead to failure or disengagement of locking mechanism [4, 12] resulting in dislocation of the polyethylene insert. Disengagement of locking mechanism of the insert has been described in revision total knee arthroplsty [8, 9]. To our knowledge, disengagement of polyethylene of Attune Knee system (Attune, Depuy), which has been recently introduced, has not been reported in the literature. The design features include gradually reducing femoral radius, S-curve design of cam and spine, AOX(anti-oxidant) polyethylene and LOGICLOCKTMtibial base that has a central locking mechanism[7]. In our case, there was complete disengagement of the polyethylene. Retrieval analysis could not be done so it is difficult to say if the failure of locking mechanism is related to design flaw. There are many case reports regarding screw loosening due to micro motion [2,8,10], but to the best of our knowledge disengagement of polyethylene with central locking mechanism in a newly introduced system (Attune, Depuy) as in our case has not been described. Wright et al., postulated that repetitive flexion produced an anterior superior force leading to failure of the locking mechanism [11]. Locking mechanism failure can lead to subluxation or dissociation of the polyethylene. The mechanism are anterior lift off of the bearing in flexion resulting in an anterior displacement or posterior lift off in extension with posterior dislocation of the polyethylene [2]. In our case, there was complete dislocation of the polyethylene into the posterolateral compartment. In our case, immediate postoperative radiographs after index surgery showed posterior slope with flexion of femoral component. Patient functioned well for 10 months after surgery, hence, surgical error at the time of index surgery can be ruled out. There are many limitations in our case report.. We could not exactly point out the exact mechanism of locking mechanism failure and the retrieval analysis of the explanted polyethylene was not done.

Conclusion

Tibial insert locking mechanism failure is a rare but serious complication. Though it is described in revision scenario, it can occur in primary with acute knee swelling and instability. Prompt diagnosis of this issue and addressing the underlying cause is mandatory to prevent any further dislocation. Disengagement of polyethylene should be a differential diagnosis in patients who present with acute swelling and instability of the knee. Though it’s a rare complication it is quite possible that design of the implant or its locking mechanism could contribute to dislocations in future.
  11 in total

1.  Backside wear of modular ultra-high molecular weight polyethylene tibial inserts.

Authors:  Michael A Conditt; Sabir K Ismaily; Jerry W Alexander; Philip C Noble
Journal:  J Bone Joint Surg Am       Date:  2004-05       Impact factor: 5.284

2.  Factors affecting the severity of backside wear of modular tibial inserts.

Authors:  Michael A Conditt; Jason A Stein; Philip C Noble
Journal:  J Bone Joint Surg Am       Date:  2004-02       Impact factor: 5.284

3.  Nontraumatic, spontaneous dislocation of polyethylene tibial insert 1 year after TKA.

Authors:  Russell C Wright; Andrew Crouch; Stephan V Yacoubian; Raymond B Raven; Raymond B Ravan; Yuri Falkinstein; Shahan V Yacoubian
Journal:  Orthopedics       Date:  2011-12-06       Impact factor: 1.390

4.  Failure of tibial polyethylene insert locking mechanism in posterior stabilized arthroplasty.

Authors:  Emmanuel Thienpont
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-01       Impact factor: 4.342

5.  Five cases of failure of the tibial polyethylene insert locking mechanism in one design of constrained knee arthroplasty.

Authors:  Venkata R Rapuri; Henry D Clarke; Mark J Spangehl; Christopher P Beauchamp
Journal:  J Arthroplasty       Date:  2010-08-30       Impact factor: 4.757

6.  Dissociation of polyethylene insert from the tibial baseplate following revision total knee arthroplasty.

Authors:  Chin-En Chen; Rei-Jahn Juhn; Jih-Yang Ko
Journal:  J Arthroplasty       Date:  2010-05-31       Impact factor: 4.757

7.  Acute locking in revision total knee arthroplasty due to disengagement of the locking screw.

Authors:  J B Nadkarni; D G Carden
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-10-09       Impact factor: 4.342

8.  How prevalent are implant wear and osteolysis, and how has the scope of osteolysis changed since 2000?

Authors:  Amanda Marshall; Michael D Ries; Wayne Paprosky
Journal:  J Am Acad Orthop Surg       Date:  2008       Impact factor: 3.020

9.  Tibial insert micromotion of various total knee arthroplasty devices.

Authors:  Safia Bhimji; Aiguo Wang; Thomas Schmalzried
Journal:  J Knee Surg       Date:  2010-09       Impact factor: 2.757

10.  Knee biomechanics and total knee replacement.

Authors:  T P Andriacchi; T S Stanwyck; J O Galante
Journal:  J Arthroplasty       Date:  1986       Impact factor: 4.757

View more
  1 in total

1.  Locking mechanism failure between tibial baseplate and polyethylene insert in cruciate retaining total knee arthroplasty.

Authors:  Anirudh Sharma; Vijay Killampalli; Arpit Patel
Journal:  J Clin Orthop Trauma       Date:  2022-09-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.