Metal Artifact Reduction Magnetic Resonance Imaging Around Arthroplasty Implants: The Negative Effect of Long Echo Trains on the Implant-Related Artifact.

Neil M Kumar1, Cesar de Cesar Netto, Lew C Schon, Jan Fritz.   

Abstract

OBJECTIVES: Long echo train length (ETL) is an often recommended but unproven technique to decrease metal artifacts on magnetic resonance imaging (MRI) scans. Therefore, we quantitatively and qualitatively assessed the effects of ETL on metal artifact on MRI scans using a cobalt-chromium-containing arthroplasty implant system.
MATERIALS AND METHODS: Using a total ankle arthroplasty system implanted into a human cadaver ankle and a clinical 1.5 T MRI system, turbo spin echo (TSE) pulse sequences were acquired with ETL ranging from 3 to 23 and receiver bandwidth (BW) from 100 to 750 Hz/pixel, whereas effective echo time and spatial resolution were controlled. A compressed sensing slice encoding for metal artifact correction TSE prototype pulse sequence was used as reference standard. End points included the total implant-related artifact area and implant-related signal void areas. Two raters evaluated the overall image quality and preference across varying BW and ETL. Two-factor analysis of variance, Friedman test, Kruskal-Wallis test, and Pearson correlation were used. P values of less than 0.05 were considered statistically significant.
RESULTS: The total implant-related artifact area ranged from 0.119 for compressed sensing slice encoding for metal artifact correction (BW, 600 Hz/pixel; ETL, 3) to 0.265 for TSE (BW, 100 Hz/pixel; ETL, 23). Longer ETL significantly increases the total implant-related artifact area (P = 0.0004), whereas it decreased with increasing BW (P < 0.0001). Implant-related signal void areas were not significantly affected by larger echo train length, but reduced with higher BW (P < 0.0001). Readers had a significant preference for images with high BW and short ETL (P < 0.0001).
CONCLUSIONS: High receiver BW is the most effective parameter for reduction of arthroplasty implant-induced metal artifact on MRI scans, whereas in contradiction to prevalent notions, long echo trains fail to reduce implant-related metal artifacts, but in fact cause degradation of image quality around the implant with resultant larger appearing total metal artifacts.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28079703     DOI: 10.1097/RLI.0000000000000350

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  7 in total

1.  Ultrashort Time-to-Echo Magnetic Resonance Imaging at 3 T for the Detection of Spondylolysis in Cadaveric Spines: Comparison With CT.

Authors:  Tim Finkenstaedt; Palanan Siriwanarangsun; Suraj Achar; Michael Carl; Sina Finkenstaedt; Nirusha Abeydeera; Christine B Chung; Won C Bae
Journal:  Invest Radiol       Date:  2019-01       Impact factor: 6.016

2.  New-Generation Low-Field Magnetic Resonance Imaging of Hip Arthroplasty Implants Using Slice Encoding for Metal Artifact Correction: First In Vitro Experience at 0.55 T and Comparison With 1.5 T.

Authors:  Iman Khodarahmi; Inge M Brinkmann; Dana J Lin; Mary Bruno; Patricia M Johnson; Florian Knoll; Mahesh B Keerthivasan; Hersh Chandarana; Jan Fritz
Journal:  Invest Radiol       Date:  2022-03-04       Impact factor: 10.065

3.  The mysterious needle in the heart: a case report.

Authors:  Ciara Mahon; Peter Gatehouse; John Baksi; Raad H Mohiaddin
Journal:  Eur Heart J Case Rep       Date:  2020-05-03

Review 4.  State-of-the-Art Imaging Techniques in Metastatic Spinal Cord Compression.

Authors:  Tricia Kuah; Balamurugan A Vellayappan; Andrew Makmur; Shalini Nair; Junda Song; Jiong Hao Tan; Naresh Kumar; Swee Tian Quek; James Thomas Patrick Decourcy Hallinan
Journal:  Cancers (Basel)       Date:  2022-07-05       Impact factor: 6.575

5.  Reduction in Acquisition Time and Improvement in Image Quality in T2-Weighted MR Imaging of Musculoskeletal Tumors of the Extremities Using a Novel Deep Learning-Based Reconstruction Technique in a Turbo Spin Echo (TSE) Sequence.

Authors:  Daniel Wessling; Judith Herrmann; Saif Afat; Dominik Nickel; Ahmed E Othman; Haidara Almansour; Sebastian Gassenmaier
Journal:  Tomography       Date:  2022-07-06

Review 6.  [Metallic Artifacts on MR Imaging and Methods for Their Reduction].

Authors:  Hye Jung Choo; Sun Joo Lee; Young Han Lee
Journal:  Taehan Yongsang Uihakhoe Chi       Date:  2020-01-31

7.  Detailed bone assessment of the sacroiliac joint in a prospective imaging study: comparison between computed tomography, zero echo time, and black bone magnetic resonance imaging.

Authors:  Lucas Wolharn; Roman Guggenberger; Kai Higashigaito; Thomas Sartoretti; Sebastian Winklhofer; Christine B Chung; Tim Finkenstaedt
Journal:  Skeletal Radiol       Date:  2022-06-30       Impact factor: 2.128

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.