Literature DB >> 28064103

Why do some titanium-alloy total hip arthroplasty modular necks fail?

Samo K Fokter1, Andrej Moličnik2, Rajko Kavalar3, Primož Pelicon4, Rebeka Rudolf5, Nenad Gubeljak6.   

Abstract

BACKGROUND: Increased modularity in total hip arthroplasty (THA) through extra junction between the neck and the femoral stem is gaining popularity among orthopaedic community. However, the advantage of the additional junction is shadowed by an increased risk of mechanical failure. The aim of this study was to describe the exact mechanism of fracture of the modular femoral neck in an uncemented stem.
METHODS: Clinical, metallurgical, and mechanical analysis including finite-element modelling and elemental-sensitive tissue analysis with the micro-PIXE method was performed on two patients treated with fully modular primary THA made from Titanium alloy of the same oval taper-cone design. In patient A revision was performed 7.8 years after the unilateral primary procedure because of modular femoral neck fracture, while patient B was left-side revised 15 years after the bilateral primary procedure because of aseptic loosening of the femoral stem.
RESULTS: Body weight was 30% higher and the arm of implanted modular femoral neck was 51% longer in patient A compared to patient B. Therefore, the stress ratio on the modular femoral neck of patient A was calculated to be 2.45 times higher than in patient B, preventing cold welding and producing taper damage and degradation at the neck-stem junction. Large clusters of metallic debris containing Titanium and Vanadium from the alloy were present in the periprosthetic soft tissues of patient A.
CONCLUSIONS: Patients with higher body mass index treated with fully modular Ti-alloy THA may be at increased risk to experience catastrophic failure of the device. Orthopaedic surgeons should avoid using long necks whenever possible, as these are especially prone to develop a vicious circle starting with the fretting process and crevice corrosion at the taper-cone connection, leading to crack initiation and crack propagation, accelerated by the increased vulnerability of the Ti-alloy in biologic media, ultimately ending as fracture at the typical site. Serum Ti concentration may represent a rough estimation of taper degradation and patients with elevated levels should be warned and followed accordingly.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Biomaterials; Fatigue fracture; Modular femoral necks; Revision; Titanium alloys; Total hip arthroplasty

Mesh:

Substances:

Year:  2016        PMID: 28064103     DOI: 10.1016/j.jmbbm.2016.12.012

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  5 in total

1.  Cementless modular neck stems: are they a safe option in primary total hip arthroplasty?

Authors:  José M Pelayo-de-Tomás; José L Rodrigo-Pérez; Carlos D Novoa-Parra; Alejandro Lizaur-Utrilla; María Morales-Suárez-Varela; José Antonio Blas-Dobón
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-29

2.  The mechanical effects of cup inclination and anteversion angle on the bearing surface.

Authors:  Nobuhiro Kaku; Tomonori Tabata; Hiroaki Tagomori; Tetsutaro Abe; Hiroshi Tsumura
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-08-05

3.  Long-term survivorship of an exchangeable-neck hip prosthesis with a Ti-alloy/Ti-alloy neck-stem junction.

Authors:  Massimiliano Baleani; Aldo Toni; Cristina Ancarani; Susanna Stea; Barbara Bordini
Journal:  Arch Orthop Trauma Surg       Date:  2022-09-30       Impact factor: 2.928

4.  Total Knee Replacement with an Uncemented Porous Tantalum Tibia Component: A Failure Analysis.

Authors:  Samo K Fokter; Nenad Gubeljak; Esther Punzón-Quijorna; Primož Pelicon; Mitja Kelemen; Primož Vavpetič; Jožef Predan; Luka Ferlič; Igor Novak
Journal:  Materials (Basel)       Date:  2022-03-31       Impact factor: 3.623

5.  Femoral Stem Fracture in Hip Revision Arthroplasty: A Systematic Literature Review of the Real-World Evidence.

Authors:  Matthias Sukopp; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Z Orthop Unfall       Date:  2021-04-13       Impact factor: 0.923

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.