A comparison of two ambulatory blood pressure monitors, the Del Mar Avionics Pressurometer IV and the Spacelabs 90202.

S Santucci1, E M Cates, G D James, Y R Schussel, D Steiner, T G Pickering.   

Abstract

The purpose of this study was to evaluate the accuracy of two noninvasive ambulatory blood pressure (BP) monitors, the new Del Mar Avionics Pressurometer IV (PIV) and the second generation Spacelabs 90202. Two sets of comparisons were made between two trained observers and the monitors on 17 normotensive and 28 hypertensive subjects. The result of the first comparison showed that the average difference between observers and the PIV was 1.2/-2.2 mm Hg for systolic and diastolic BP. Comparisons of both monitors to the observers showed differences of 0.9/1.0 mm Hg for the PIV and 0.3/0.8 mm Hg for the 90202 (systolic/diastolic). Correlations between the average observer reading and the monitors over both sets of comparisons ranged from 0.90 (diastolic 90202 vs observer) to 0.98 (systolic PIV vs observer). The percentage of readings within 5 mm Hg between observer and monitor over all comparisons ranged from a low of 68% (90202, systolic and diastolic) to a high of 83% (PIV, diastolic). In general, both devices are satisfactory and reliable machines for ambulatory BP monitoring, but each has a slightly different bias compared to auscultatory readings.

Entities:  

Mesh:

Year:  1989        PMID: 2803673     DOI: 10.1093/ajh/2.10.797

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  8 in total

Review 1.  Principles and techniques of blood pressure measurement.

Authors:  Gbenga Ogedegbe; Thomas Pickering
Journal:  Cardiol Clin       Date:  2010-11       Impact factor: 2.213

2.  Twenty-four hour blood pressure monitoring in healthy children.

Authors:  F Krull; T Buck; G Offner; J Brodehl
Journal:  Eur J Pediatr       Date:  1993-07       Impact factor: 3.183

3.  Relationship between waking-sleep blood pressure and catecholamine changes in African-American and European-American women.

Authors:  Helene M van Berge-Landry; Dana H Bovbjerg; Gary D James
Journal:  Blood Press Monit       Date:  2008-10       Impact factor: 1.444

4.  A comparison study of brachial blood pressure recorded with Spacelabs 90217A and Mobil-O-Graph NG devices under static and ambulatory conditions.

Authors:  P A Sarafidis; A A Lazaridis; K P Imprialos; P I Georgianos; K A Avranas; A D Protogerou; M N Doumas; V G Athyros; A I Karagiannis
Journal:  J Hum Hypertens       Date:  2016-03-03       Impact factor: 3.012

5.  A comparison of two ambulatory blood pressure monitors worn at the same time.

Authors:  Radhakrishna R Kallem; Kevin E C Meyers; Deirdre L Sawinski; Raymond R Townsend
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-04-08       Impact factor: 3.738

6.  What is the truth? Differences in ambulatory blood pressure data between monitors.

Authors:  Michiaki Nagai; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-04-04       Impact factor: 3.738

Review 7.  Recent Research and Developing Trends of Wearable Sensors for Detecting Blood Pressure.

Authors:  Toshiya Arakawa
Journal:  Sensors (Basel)       Date:  2018-08-23       Impact factor: 3.576

8.  Design and Fabrication of a New Wearable Pressure Sensor for Blood Pressure Monitoring.

Authors:  Marian Ion; Silviu Dinulescu; Bogdan Firtat; Mihaela Savin; Octavian N Ionescu; Carmen Moldovan
Journal:  Sensors (Basel)       Date:  2021-03-16       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.