Nandhinee Radha Shanmugam1, Anjan Panneer Selvam1, Thomas W Barrett2, Steven C Kazmierczak3, Milin Nilesh Rana4, Shalini Prasad1.
Abstract
AIM: To demonstrate the design, fabrication and testing of a portable, label-free biosensor for ultrasensitive detection of the cardiac Troponin-T (cTnT) from patient blood. MATERIALS &Entities:
Keywords: Troponin-T; electrical double layer; impedance spectroscopy; nanoconfinement; nanoporous
Year: 2015 PMID: 28031897 PMCID: PMC5137999 DOI: 10.4155/fso.15.22
Source DB: PubMed Journal: Future Sci OA ISSN: 2056-5623
Components of the developed sensor platform.
(A) Assembled sensor platform. (B) SEM micrograph of nanoporous nylon membrane with average pore diameter of 200 nm. (C) Characteristic representation of immunoassay inside nanowells.
cTnT:Cardiac Troponin-T; DSP:Dithiobis succinimidyl propionate; EDL:Electrical double layer; IHP:Inner Helmholtz plane; OHP:Outer Helmholtz plane.
Table representing the comparison of cardiac Troponin-T concentrations between electrochemical and chemiluminescent immunoassay detection method.
| P1 | 0.61 | <10 | N/A |
| P2 | 70.85 | 68 | 2.9 |
| P3 | 68 | 73 | 5.01 |
| P4 | 79.8 | 81 | 1.06 |
| P5 | 85.5 | 91 | 4.41 |
| P6 | 92.65 | 94 | 1.02 |
| P7 | 106.05 | 101 | 3.45 |
| P8 | 90 | 102 | 8.84 |
| P9 | 110.35 | 111 | 0.42 |
| P10 | 112 | 112 | 0 |
| P11 | 118.5 | 126 | 4.34 |
| P12 | 116.5 | 127 | 6.1 |
| P13 | 133.5 | 137 | 1.83 |
| P14 | 154.65 | 153 | 0.76 |
| P15 | 154.5 | 155 | 0.23 |
| P16 | 156 | 167 | 4.82 |
| P17 | 163 | 169 | 2.56 |
| P18 | 195.9 | 199 | 1.11 |
| P19 | 201.1 | 210 | 3.06 |
| P20 | 255.9 | 259 | 0.85 |
| P21 | 263 | 276 | 3.41 |
| P22 | 284.4 | 284 | 0.1 |
| P23 | 288.55 | 289 | 0.11 |
| P24 | 312.65 | 315 | 0.53 |
| P25 | 509.7 | 507 | 0.38 |
| P26 | 508 | 517 | 1.24 |
| P27 | 609.5 | 635 | 2.9 |
| P28 | 671.5 | 686 | 1.51 |
| P29 | 763.75 | 767 | 0.3 |
| P30 | 771.5 | 782 | 0.96 |
| P31 | 802.8 | 803 | 0.02 |
| P32 | 837.25 | 821 | 1.39 |
| P33 | 842.5 | 878 | 2.92 |
| P34 | 981.5 | 911 | 5.27 |
| P35 | 1075 | 1100 | 1.63 |
| P36 | 1261.5 | 1120 | 8.4 |
| P37 | 1222 | 1160 | 3.68 |
| P38 | 1255 | 1310 | 3.03 |
| P39 | 1765 | 1820 | 2.17 |
| P40 | 2775 | 2840 | 1.64 |
Sensor performance was compared for n = 40 clinical samples.
Calibration dose response analysis for cardiac Troponin-T antigen diluted in 7% bovine serum albumin.
Cardiac Troponin-T concentrations in the range of 0.001 pg/ml to 10,000 pg/ml were analyzed for n = 3 replicates. Error bars indicate two standard deviation for n = 3 replicates.
Regression plot for linearity experiments.
Interference experiments with triglyceride concentrations in the range of 0 mg/dl–2000 mg/dl.
Triglyceride aliquots with cTnT concentration of (A) 1000 pg/ml; (B) 4000 pg/ml; (C) 7000 pg/ml and (D) 10,000 pg/ml.
cTnT:Cardiac Troponin-T.