On the feasibility to conduct gradient liquid chromatography separations in narrow-bore columns at pressures up to 2000bar.

Ruben De Pauw1, Tim Swier1, Bart Degreef1, Gert Desmet1, Ken Broeckhoven2.   

Abstract

The limits in operating pressures are extended for narrow-bore columns in gradient elution up to 2000bar. As the required pumps for these pressures are incompatible with common chromatographic solvents and are not suitable to apply a mobile phase composition gradient, a mobile phase delivery and injection system is described and experimentally validated which allows to use any possible chromatographic solvent in isocratic and gradient elution. The mobile phase delivery and injection system also allows to perform multiple separations without the need to depressurize the column. This system consists out of 5 dual on/off valves and two large volume loops in which the gradient and equilibration volume of initial mobile phase are loaded by a commercial liquid chromatography pump. The loops are then flushed toward the column at extreme pressures. The mobile phase delivery and injection system is first evaluated in isocratic elution and shows a comparable performance to a state-of-the-art commercial flow-through-needle injector but with twice the pressure rating. Distortion of the loaded gradient by dispersion in the gradient storage loop is studied. The effect of the most important parameters (such as flow rate, pressure and gradient steepness) is experimentally investigated. Different gradient steepnesses and volumes can be applied at different flow rates and operating pressures with a good repeatability. Due to the isobaric operation of the pumps, the gradient is monitored in real-time by a mass flow meter installed at the detector outlet. The chromatograms are then converted from time to volume-base. A separation of a 19-compound sample is performed on a 300×2.1mm column at 1000bar and on a 600×2.1mm column at 2000bar. The peak capacity was found to increase from 141 to 199 and thus scales with L as is predicted by theory. This allows to conclude that the inlet pressure for narrow-bore columns in gradient elution can be increased up to 2000bar without fundamental pressure-induced limitations.
Copyright © 2016 Elsevier B.V. All rights reserved.

Keywords:  2.1mm ID; Gradient elution; Narrow-bore columns; Preloaded gradient; Ultra-high pressure; Viscous heating

Mesh:

Substances:

Year:  2016        PMID: 28029367     DOI: 10.1016/j.chroma.2016.10.008

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Liquid chromatography above 20,000 PSI.

Authors:  Matthew J Sorensen; Brady G Anderson; Robert T Kennedy
Journal:  Trends Analyt Chem       Date:  2020-01-21       Impact factor: 12.296

2.  Ultrahigh-Performance capillary liquid chromatography-mass spectrometry at 35 kpsi for separation of lipids.

Authors:  Matthew J Sorensen; Kelsey E Miller; James W Jorgenson; Robert T Kennedy
Journal:  J Chromatogr A       Date:  2019-09-26       Impact factor: 4.759

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.