Daniela Serio1, Andreas Zierer2, Mirko Doss1, Anton Moritz2.
Abstract
Transcatheter aortic valve implantation (TAVI) has been recently established as a less invasive alternative to conventional aortic valve replacement (CAVR) in patients presenting with expected high procedural risk. The rapid technologic advances and the recent improvement of clinical outcomes with TAVI have made it possible to treat degenerated bioprosthesis using the valve-in-valve implantation concept (Walther T, Simon P, Dewey T, et al. Transapical minimally invasive aortic valve implantation: multicenter experience. Circulation 2007;116(11, Suppl):I240-I245; Webb JG, Pasupati S, Humphries K, et al. Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis. Circulation 2007;116(7):755-763). Recently, concerns were raised regarding the appropriate sizing of transcatheter valve prosthesis and its effect on residual stenosis (Klaaborg KE, Egeblad H, Jakobsen CJ, et al. Transapical transcatheter treatment of a stenosed aortic valve bioprosthesis using the Edwards SAPIEN Transcatheter Heart Valve. Ann Thorac Surg 2009;87(6):1943-1946; Zedig R, Achouh P, Berrebi A, et al. Valve- in- a- valve implantation: a word of caution. Ann Thorac Surg 2009;87:1943-1946), eventually resulting in a patient-prosthesis mismatch (PPM). We report a case with severe PPM after inserting a 23-mm Sapien (Edwards Lifesciences, Irvine, United States) valve into degenerated 23-mm Medtronic Mosaic prosthesis (Medtronic, Minneapolis, Minnesota, United States).Entities:
Keywords: TAVI; aortic valve and root; heart valve; surgery; transapical
Year: 2016 PMID: 28018817 PMCID: PMC5177444 DOI: 10.1055/s-0036-1583298
Source DB: PubMed Journal: Thorac Cardiovasc Surg Rep ISSN: 2194-7635
Fig. 1The 23-mm Sapien valve was implanted by transapical procedure in the degenerated 23-mm Medtronic Mosaic valve.
Fig. 2View on the aortic valve: The right leaflet shows a thrombus and a rotation of one of the commissure of the Sapien valve in relation to the stents of the Mosaic prosthesis.
Fig. 3The explanted Sapien valve shows an asymmetric expansion within the Mosaic stent.