Abstract
A new method is described to calibrate tristimulus colorimeters for high accuracy color measurements. Instead of traditional lamp standards, modern, high accuracy detector standards are suggested for calibration. After high accuracy absolute spectral response determination of the tristimulus receivers, color (spectral) correction and peak (amplitude) normalization can minimize uncertainties caused by imperfect realizations of the Commission Internationale de l'Eclairage (CIE) color matching functions. As a result of the corrections, stable light sources of different spectral power distributions can be measured with an accuracy dominated by the sub tenths of a percent uncertainty of novel spectral response determinations.Entities:
Keywords: chromaticity coordinates; color calibration; color temperature; colorimetry; detector standard; photometry; spectral response; tristimulus values
Year: 1998 PMID: 28009362 PMCID: PMC4890952 DOI: 10.6028/jres.103.040
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X