Literature DB >> 18253276

Characterization of a polarization-independent transmission trap detector.

T Kübarsepp, P Kärhä, E Ikonen.   

Abstract

A six-element polarization-independent transmission trap detector with coaxial input and output beams has been constructed and full characterized. The measured optical parameters are compared with their values, predicted by Fresnel equations. Measured transmittances are in agreement with the predicted values within 2 x 10(-5) in the wavelength region from 450 to 650nm. The spectral responsivity of the transmission trap detector is in agreement with the predicted values within 0.035% at 543.5- and 633.0-nm vacuum wavelengths. The spatial uniformity of the responsivity is +/-0.03% across the active area of approximately 5 x 6 mm(2), measured with a laser beam of 1-mm diameter. The angular uniformity of the transmission trap detector is better than +/-0.01% for +/-3 degrees rotation around two perpendicular axes.

Year:  1997        PMID: 18253276     DOI: 10.1364/ao.36.002807

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  2 in total

1.  Opto-Mechanical and Electronic Design of a Tunnel-Trap Si Radiometer.

Authors:  G P Eppeldauer; D C Lynch
Journal:  J Res Natl Inst Stand Technol       Date:  2000-12-01

2.  Spectral Response Based Calibration Method of Tristimulus Colorimeters.

Authors:  George Eppeldauer
Journal:  J Res Natl Inst Stand Technol       Date:  1998-12-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.