Literature DB >> 27956077

Modular cementless prosthetic reconstruction after resection of lower extremity malignant tumor.

Senol Bekmez1, Mehmet Ayvaz2, Altug Yucekul2, Mazhar Tokgozoglu2.   

Abstract

INTRODUCTION: The aim of the study is to report the preliminary clinical and functional outcomes of a modular cementless tumor resection prosthesis system (Megasystem-C®, Waldemar Link GmbH&Co. KG, Hamburg, Germany) in patients undergoing limb salvage surgery with wide resection in a lower extremity primary or metastatic malignant bone tumor.
MATERIAL AND METHODS: Fifty-two consecutive patients (33 male and 19 female; mean age 37.1 years (range, 16 to 79) with primary or metastatic lower extremity malignant bone tumor who underwent wide resection and reconstruction with cementless Megasystem-C® system were included in the study. Patients were analyzed for age at diagnosis, gender, type and localization of the tumor, time of follow-up, patient and prosthesis survival, complications, oncological and functional outcomes.
RESULTS: Mean follow-up time was 43.2 months (range, 8 to 66). Cumulative patient survival rate was 92.3 percent and cumulative prosthetic survival rate was 65.4 percent. 18 complications were recorded and 9 of them required revision (17.3 percent). Mean overall Musculoskeletal Tumor Society score score was 72.7 percent (range, 52 to 86). Subgroup analyzes demonstrated no difference in complication rates, overall patient or prosthetic survivals. Functional scores according to age, diagnosis and location of the reconstruction also were not significantly different.
CONCLUSION: The preliminary short-term follow-up results revealed that, the new generation modular cementless endoprosthetic system offers promising clinical and functional outcomes with reasonable complication rates. LEVEL OF EVIDENCE: Level IV, Therapeutic study.
Copyright © 2016 Turkish Association of Orthopaedics and Traumatology. Production and hosting by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Bone tumor; Cementless; Limb salvage; Modular; Tumor prosthesis

Mesh:

Year:  2016        PMID: 27956077      PMCID: PMC6197450          DOI: 10.1016/j.aott.2016.05.004

Source DB:  PubMed          Journal:  Acta Orthop Traumatol Turc        ISSN: 1017-995X            Impact factor:   1.511


Introduction

The life expectancy of patients with malignant bone tumors increased after improvements in the knowledge for tumor biology, advances in diagnostic abilities, adjuvant treatment modalities and surgical techniques.1, 2 Because of this, functional status of the patient becomes a major issue in the treatment. Limb-salvage surgery offers better functional outcomes and quality of life without a reduction in survival or an increase in morbidity when compared to amputation.3, 4, 5, 6 Selection of the type of limb-salvage procedure is based on the tumor location and the patient characteristics with aiming a durable reconstruction and favourable functional outcomes.7, 8, 9 Endoprosthetic reconstruction is a reliable option in periarticular tumor resections. It provides component modularity, improved fixation, near anatomic appearance and good-to-excellent functional results.10, 11, 12, 13 Modular endoprosthetic systems have either cemented and cementless stem fixation options. Early reports stated that, cemented modular systems were associated with intermediate to long-term problems of aseptic loosening, mechanical breakage and infection with high failure rates.14, 15, 16 Thereof, cementless stems have gained acceptance in limb sparing surgery to minimize the risk of failure. Recent studies demonstrated that, cementless prosthetic systems have favourable outcomes in terms of infection and aseptic loosening.17, 18, 19 The aim of this prospective case series is to report the preliminary clinical and functional outcomes of a modular cementless tumor resection prosthesis system with titanium tapered stem that has splines for cementless fixation.

Patients and methods

After obtaining institutional review board approval, we prospectively followed our patients who underwent limb reconstruction with the Megasystem-C® modular prosthesis system (Waldemar Link GmbH&Co. KG, Hamburg, Germany) after wide resection of a lower extremity malignant tumor, starting in 2008 up until 2012, in our institution. Inclusion criteria were to have malignant primary or metastatic tumor in femur and tibia which was histologically proven by biopsy, reconstruction with cementless stem fixation and prosthetic replacement as a primary reconstructive procedure. Patients with pelvic tumor, reconstruction with cemented stem fixation, reconstruction with a different modular prosthetic system and prosthetic replacement after a previously failed reconstructive surgery were excluded. Fifty-two consecutive patients with primary or metastatic lower extremity malignant bone tumor who underwent wide resection and reconstruction with cementless Megasystem-C® system were included in the study. Mean age at diagnosis was 37.1 years (range, 16 to 79). There were 33 male and 19 female patients. Twenty nine of the patients had a right-sided and twenty-three of them had a left-sided tumor. Diagnosis was osteosarcoma in 17, metastatic disease in 11, Ewing sarcoma in 7 patients, giant-cell tumor in 5 patients, chondrosarcoma in 5 patients, multiple myeloma in 3 and liposarcoma, leiomyosarcoma, synovial sarcoma and neurofibrosarcoma in one patient each. Average follow-up time was 43.2 months (range, 8 to 66). The reconstruction was in proximal femur in 18 patients, distal femur in 18 patients, proximal tibia in 10 patients and total femoral replacement in 6 patients. Detailed description of patient information including demographics is summarized in Table 2.
Table 2

Patient demographics and tumor characteristics.

DiagnosisNumberMean ageGenderF/MR/LTumor localization
Proximal femurDiaphysis femurDistal femurProximal tibia
Osteosarcoma1720.15/1210/73275
Ewing sarcoma717.13/43/4232
Metastases1160.84/77/4623
Giant cell tumor532.63/22/332
Chondrosarcoma555.42/34/1311
Othersa751.42/53/452
Overall5237.119/3329/23179179

Including multiple myeloma, liposarcoma, leiomyosarcoma, synovial sarcoma, neurofibrosarcoma.

Patients were analyzed for age at diagnosis, gender, type and location of tumor, follow-up time, patient and prosthesis survivals, complications such as infection, dislocation, implant failure, aseptic loosening and soft-tissue related problems, oncological and functional outcomes. Functional outcomes were determined with the revised Musculoskeletal Tumor Society (MSTS) rating scale. Failure of reconstruction was classified as described by Henderson et al (Table 1). According to this, failed reconstruction was defined as a reconstruction that required revision of the complete or failed portion of prosthesis, fixation of a periprosthetic fracture, soft-tissue reconstruction to restore joint stability such as instability, tendon rupture or aseptic wound dehiscence or endoprosthetic removal without revision and amputation.
Table 1

Classification of the mode of failure in tumor resection prosthesis reconstruction as described by Henderson et al.

Type of failureDescription
Mechanical1Instability, tendon rupture, aseptic wound dehiscence
2Clinical and radiographic evidence of aseptic loosening
3Periprosthetic or prosthetic fracture
Non-mechanical4Infection requiring removal of prosthesis
5Recurrence or progression of tumor
Classification of the mode of failure in tumor resection prosthesis reconstruction as described by Henderson et al. Patient demographics and tumor characteristics. Including multiple myeloma, liposarcoma, leiomyosarcoma, synovial sarcoma, neurofibrosarcoma. Statistical analyzes were performed with the SPSS statistical software package (version 20.0; SPSS, Chicago, Illinois). Kaplan–Meier analyzes were used to determine the patient and prosthesis survivals. Chi-square tests were used to determine a significant difference between overall, region, age or diagnosis-specific complication rates. Kruskal–Wallis test was also used to analyze overall and region-specific functional scores. For all analyzes, a p score less than 0.05 was seeked for a statistical significance.

Results

Survival analysis

At the final follow-up, 48 patients were alive (92.3%) and implants were free of problems in 43 patients (82.3%). Kaplan–Meier survival analyzes revealed that, 5-year cumulative patient survival rate was 91.5% (Fig. 1). With allowing prosthetic removal as an end-point, 5-year cumulative implant survival rate was 65.4% (Fig. 2). There was no significant difference in the region specific prosthetic survivals (p = 0.332).
Fig. 1

Kaplan–Meier analysis diagram demonstrating the overall patient survival.

Fig. 2

Kaplan–Meier analysis diagram demonstrating the implant survival with using prosthesis removal as an end-point.

Kaplan–Meier analysis diagram demonstrating the overall patient survival. Kaplan–Meier analysis diagram demonstrating the implant survival with using prosthesis removal as an end-point.

Oncological outcomes

Histopathologic evaluation have revealed tumor free surgical margins in all of the patients. Despite this, local recurrence or distant metastasis occurred in four patients. There were three patients with local recurrence; an 18 years-old male with osteosarcoma in the distal femur, a 60 years-old male with a lung carcinoma metastasis in the proximal femur and a 61 years-old male with a rectum carcinoma metastases in the distal femur. These patients were dead at the 22nd, 13th and 8th month, respectively, due to heavy metastasis. The fourth patient was a 16 years-old male with Ewing sarcoma in the proximal tibia with lung metastasis. The patient was dead 11 months after surgery due to pulmonary failure.

Complications

There was a total number of 18 complications (34.6 percent). The most common complications were infection (13 percent) and soft-tissue related problems (11 percent). Six patients had soft-tissue related problems such as simple skin necrosis and aseptic wound dehiscence at the early postoperative period, which were all treated with debridement and skin grafting, so did not require revision of the prosthesis. Nine patients (17.3 percent) had failure of reconstruction requiring revision. There was no soft-tissue problem (Type 1 failure) requiring revision. There was no aseptic loosening (Type 2 failure) in any patients. The segmental failure of the prosthesis due to design which was classified as structural (Type 3) failure occured in two patients. One of them had proximal femoral and the other had total femoral reconstruction. These two patients were revised at the first and third months, respectively. No periprosthetic fracture or dislocation occurred in any patients. Seven patients (3 distal femur, 3 proximal tibia and 1 proximal femur) had periprosthetic infection (Type 4 failure) which was successfully managed with two-staged revision surgery. First, prosthesis removed, antibiotic-loaded spacers implemented and proper parenteral antibiotic administrated. After eradication of infection. second stage was performed with modular endoprosthesis. All of these seven patients were infection free at the latest follow up. Local tumor recurrence was observed in three patients (Type 5 failure). Distribution of complications according to the type of reconstruction and diagnosis are summarized in Table 3, Table 4.
Table 3

Distribution of complications according to type of reconstruction.

NumberComplications (%)
Soft-tissue complicationAseptic looseningStructural failureInfectionTumor progressionTotal
ReconstructionProximal femur181113
Total femur611
Distal femur1831329
Proximal tibia10235
Overall52627318
Table 4

Distribution of the complications according to diagnosis.

NumberComplications (%)
Soft-tissue complicationAseptic looseningStructural failureInfectionTumor progressionTotal
DiagnosisOsteosarcoma1754110
Ewing sarcoma711
Metastasis11123
Giant cell tumor5112
Chondrosarcoma5112
Othersa7
Overall52627318

Including multiple myeloma, liposarcoma, leiomyosarcoma, synovial sarcoma, neurofibrosarcoma.

Distribution of complications according to type of reconstruction. Distribution of the complications according to diagnosis. Including multiple myeloma, liposarcoma, leiomyosarcoma, synovial sarcoma, neurofibrosarcoma.

Functional outcomes

Mean overall MSTS score of the study cohort was 72.7 percent (range, 52 to 86). Functional scores of patients with different anatomical regions are summarized in Fig. 3.
Fig. 3

MSTS scores according to different anatomical regions. Please note that, average MSTS score in proximal tibial reconstructions is lower than the other anatomical regions.

MSTS scores according to different anatomical regions. Please note that, average MSTS score in proximal tibial reconstructions is lower than the other anatomical regions.

Discussion

There are several reconstruction options for limb salvage after a tumor resection such as arthrodesis, allografts, reimplanting the autoclaved tumor and endoprosthetic reconstruction. Metallic endoprosthesis is one of the most preferred methods, since it does not require healing of the reconstructed segment as in biological reconstructions. Thus, prosthetic replacement promotes a faster return to function.23, 24 Furthermore, component modularity allows reconstruction of massive defects such as total femoral resection. Previous studies demonstrated favourable outcomes with cementless stem fixation in terms of aseptic loosening, infection and overall prosthesis survival.17, 19, 25 In our series, we also observed the success of cementless stem fixation. At the last follow-up, in 43 of 52 patients, implants were problem-free. Again we did not observe any reconstruction failure associated with aseptic loosening. The new titanium tapered stem with splines may be the cause for success, since this stem has a proven track record in femoral revision of total hip arthroplasty. In the early applications of this system, modular segment dislodgement problem occurred in two patients at postoperative first and third months, respectively. In revision surgery, we observed that the problem was in the locking mechanism of the modular stem segments. With ours and others feedback, the locking mechanism and its surgical application technique was modified by the manufacturer. In the final version, which is in clinical use now, body segments or the stem are impacted together with a specially designed impaction instrumentation and then using two screws opposite, the joined two segments are locked rigidly (Fig. 4). Since then no another dislodgement in any patients was seen after this modification. This will perhaps improve the outcome in future.
Fig. 4

A 54 years-old male admitted to our department with a pathological fracture in the right proximal femur due to lung carcinoma metastasis (a). Wide resection and reconstruction with modular cementless tumor resection prosthesis (b). Five months later, he admitted to the emergency department with significant thigh pain. Radiograph demonstrates the rotational instability between prosthetic modules caused by module dislodgement (c). The patient underwent revision surgery (d).

A 54 years-old male admitted to our department with a pathological fracture in the right proximal femur due to lung carcinoma metastasis (a). Wide resection and reconstruction with modular cementless tumor resection prosthesis (b). Five months later, he admitted to the emergency department with significant thigh pain. Radiograph demonstrates the rotational instability between prosthetic modules caused by module dislodgement (c). The patient underwent revision surgery (d). In our series, infection was the most common cause of failure. Infection is a major concern in prosthetic reconstruction after tumoral resections.28, 29, 30 Our infection rate of 13% was comparable with the literature. In a comprehensive study of Henderson et al, an overall infection rate of 8.4 percent in 2174 patients was reported for various anatomical regions. The two-staged revision approach was successful in the treatment of infection in all six patients (Fig. 5).
Fig. 5

In the final version which is in clinical use now, body segments or the stem are impacted together with a specially designed impaction instrumentation and then using two screws opposite, the joined two segments are locked rigidly. After the modification, no segment dislodgement occurred (Courtesy of Waldemar Link GmbH&Co. KG, Hamburg, Germany).

In the final version which is in clinical use now, body segments or the stem are impacted together with a specially designed impaction instrumentation and then using two screws opposite, the joined two segments are locked rigidly. After the modification, no segment dislodgement occurred (Courtesy of Waldemar Link GmbH&Co. KG, Hamburg, Germany). With modern treatment modalities, life expectancy has increased in patients with malignant musculoskeletal tumors. Thus, functional outcomes became more important to provide a better quality of life for the patients in their remaining lifetime. Modular prosthetic replacement after tumor resection offers good functional outcomes.31, 32 In our study, an average MSTS score of 72.7% indicates that, the new generation cementless modular prosthetic system also has favourable functional outcomes. Average MSTS score in proximal tibial reconstructions was lower than the other anatomical regions. In proximal tibial resections, reattachment of patellar tendon probably has negative impacts on the extensor mechanism function. New generation prosthesis systems as the one used in our study have the advantage of better attachment options for the patellar tendon. The short term follow-up is the major limitation in this study. In addition, lack of homogenity in the study population, especially in terms of diagnosis, chemotherapy or radiotherapy exposure and concomitant soft-tissue reconstruction procedures might have an impact on the results. On the other hand, to our knowledge, this is one of the largest study in the literature of this device. In conclusion, our preliminary results revealed that, the new generation modular cementless prosthetic system with titanium tapered stem design offers promising clinical and functional outcomes and reasonable complication rates. In addition, with cementless stem technology, the risk of failure associated with aseptic loosening seems to be reduced.
  31 in total

1.  Functional outcome of limb salvage surgery with mega-endoprosthetic reconstruction for bone tumors.

Authors:  Irfan Qadir; Masood Umer; Naveed Baloch
Journal:  Arch Orthop Trauma Surg       Date:  2012-05-17       Impact factor: 3.067

2.  Loosening of massive proximal femoral cemented endoprostheses. Radiographic evidence of loosening mechanism.

Authors:  W G Ward; K S Johnston; F J Dorey; J J Eckardt
Journal:  J Arthroplasty       Date:  1997-10       Impact factor: 4.757

3.  The effect of adjuvant chemotherapy on relapse-free survival in patients with osteosarcoma of the extremity.

Authors:  M P Link; A M Goorin; A W Miser; A A Green; C B Pratt; J B Belasco; J Pritchard; J S Malpas; A R Baker; J A Kirkpatrick
Journal:  N Engl J Med       Date:  1986-06-19       Impact factor: 91.245

4.  Autoclaved tumor bone for reconstruction: an alternative in developing countries.

Authors:  Mujahid Jamil Khattak; Masood Umer; Mohammad Umar
Journal:  Clin Orthop Relat Res       Date:  2006-06       Impact factor: 4.176

Review 5.  Failure mode classification for tumor endoprostheses: retrospective review of five institutions and a literature review.

Authors:  Eric R Henderson; John S Groundland; Elisa Pala; Jeremy A Dennis; Rebecca Wooten; David Cheong; Reinhard Windhager; Rainer I Kotz; Mario Mercuri; Philipp T Funovics; Francis J Hornicek; H Thomas Temple; Pietro Ruggieri; G Douglas Letson
Journal:  J Bone Joint Surg Am       Date:  2011-03-02       Impact factor: 5.284

6.  Prosthetic survival and clinical results with use of large-segment replacements in the treatment of high-grade bone sarcomas.

Authors:  M M Malawer; L B Chou
Journal:  J Bone Joint Surg Am       Date:  1995-08       Impact factor: 5.284

Review 7.  Limb salvage for osteosarcoma in the 1980s.

Authors:  M A Simon
Journal:  Clin Orthop Relat Res       Date:  1991-09       Impact factor: 4.176

8.  Prosthetic and extremity survivorship after limb salvage for sarcoma. How long do the reconstructions last?

Authors:  S M Horowitz; D B Glasser; J M Lane; J H Healey
Journal:  Clin Orthop Relat Res       Date:  1993-08       Impact factor: 4.176

9.  Survival of modern knee tumor megaprostheses: failures, functional results, and a comparative statistical analysis.

Authors:  Elisa Pala; Giulia Trovarelli; Teresa Calabrò; Andrea Angelini; Caterina N Abati; Pietro Ruggieri
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

10.  Cemented distal femoral endoprostheses for musculoskeletal tumor: improved survival of modular versus custom implants.

Authors:  Adam J Schwartz; J Michael Kabo; Fritz C Eilber; Frederick R Eilber; Jeffrey J Eckardt
Journal:  Clin Orthop Relat Res       Date:  2009-12-22       Impact factor: 4.176

View more
  3 in total

1.  Short-term clinical outcomes of Kyocera Modular Limb Salvage System designed cementless stems for the endoprosthetic reconstruction of lower extremities: a Japanese Musculoskeletal Oncology Group multi-institutional study.

Authors:  Satoshi Tsukushi; Yoshihiro Nishida; Takeshi Hirose; Eiji Nakata; Rumi Nakagawa; Tomoki Nakamura; Jungo Imanishi; Akihito Nagano; Hironari Tamiya; Takafumi Ueda
Journal:  BMC Cancer       Date:  2022-07-16       Impact factor: 4.638

2.  [Non-cemented modular prosthetic reconstruction for bone defect after tumor resection in lower extremities].

Authors:  Yuan Li; Hairong Xu; Huachao Shan; Yang Sun; Zhen Huang; Xiaohui Niu
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2019-09-15

Review 3.  Recent Advances in the Treatment of Bone Metastases and Primary Bone Tumors: An Up-to-Date Review.

Authors:  Adrian Emilian Bădilă; Dragoș Mihai Rădulescu; Adelina-Gabriela Niculescu; Alexandru Mihai Grumezescu; Marius Rădulescu; Adrian Radu Rădulescu
Journal:  Cancers (Basel)       Date:  2021-08-23       Impact factor: 6.639

  3 in total